Influence of a Variable Overall Diameter Hydrophilic Acrylic Sharp-Edged Single-piece Intra-Ocular Lens on Capsule Opacification One Year after Surgery

被引:4
作者
Mylonas, Georgios [1 ]
Georgopoulos, Michael [1 ]
Prinz, Ana [1 ]
Vock, Lorenz [1 ]
Blum, Robert A. [1 ]
Schmidt-Erfurth, Ursula [1 ]
机构
[1] Med Univ Vienna, Dept Ophthalmol, A-1090 Vienna, Austria
关键词
1-piece IOL; cataract surgery; intraocular lens; IOL diameter; posterior capsule opacification; ND-YAG LASER; SURGICAL PREVENTION; POSTERIOR CAPSULOTOMY; CATARACT-SURGERY; OPTIC SIZE; CAPSULORHEXIS; IMPLANTATION; SILICONE;
D O I
10.3109/02713683.2013.863939
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: To investigate the influence of the overall intraocular lens (IOL) diameter on posterior capsule opacification (PCO) formation. Methods: In this prospective randomized clinical trial, 124 eyes of 62 patients with bilateral age-related cataract were included. Each patient received a Corneal A501D IOL in one eye and a Corneal J501D IOL in the fellow eye. Best corrected visual acuity (BCVA) and digital slitlamp photographs were taken a 1 h, 1 week, 1, 3, 6 and 12 months postoperatively. The amount of PCO was assessed subjectively at the slitlamp and objectively using automated image-analysis software. Results: We found a mean BCVA of 0.81 +/- 0.2 for the Corneal A501D group and 0.79 +/- 0.21 for the Corneal J501D group. There was no significant difference 12 months after surgery between the two IOLs (p>0.05). Objective PCO assessment resulted in a mean PCO score (scale 0-10) of 1.65 +/- 1.71 was found for the Corneal J501D group and a score of 1.54 +/- 1.64 was found for the Corneal A501D group (p>0.05). The subjective PCO assessment at 1 year resulted in a mean PCO score of 2.0 +/- 1.74 in the Corneal J501D group and 2.13 +/- 1.64 in the Corneal A501D group (p>0.05). Conclusion: Both investigated IOLs showed good clinical performance regarding PCO and BCVA. Our study suggests that the use of an IOL with variable total diameter seems not to influence the rate of PCO formation.
引用
收藏
页码:620 / 625
页数:6
相关论文
共 25 条
[1]   Results of hydrophilic acrylic, hydrophobic acrylic, and silicone intraocular lenses in uveitic eyes with cataract - Comparison to a control group [J].
Abela-Formanek, C ;
Amon, M ;
Schauersberger, J ;
Kruger, A ;
Nepp, J ;
Schild, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) :1141-1152
[2]   Surgical prevention of posterior capsule opacification Part 1: Progress in eliminating this complication of cataract surgery [J].
Apple, DJ ;
Peng, Q ;
Visessook, N ;
Werner, L ;
Pandey, SK ;
Escobar-Gomez, M ;
Ram, J ;
Whiteside, SB ;
Schoderbeck, R ;
Ready, EL ;
Guindi, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (02) :180-187
[3]   Reproducibility of standardized retroillumination photography for quantification of posterior capsule opacification [J].
Buehl, W ;
Findl, O ;
Menapace, R ;
Georgopoulos, M ;
Rainer, G ;
Wirtitsch, M ;
Siegl, H ;
Pinz, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :265-270
[4]  
Cimberle AM, 2007, OCULAR SURG NEWS, V25, P8
[5]   Comparison of 4 methods for quantifying posterior capsule opacification [J].
Findl, O ;
Buehl, W ;
Menapace, R ;
Georgopoulos, M ;
Rainer, G ;
Siegl, H ;
Kaider, A ;
Pinz, A .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) :106-111
[6]   Posterior capsule opacification after implantation of a hydrogel intraocular lens [J].
Hayashi, K ;
Hayashi, H .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (02) :182-185
[7]   Comparison of posterior capsule opacification rates between hydrophilic and hydrophobic single-viece acrylic intraocular lenses [J].
Heatley, CJ ;
Spalton, DJ ;
Kumar, A ;
Jose, R ;
Boyce, J ;
Bender, LE .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04) :718-724
[8]   The effect of polymethylmethacrylate, silicone, and polyacrylic intraocular lenses on posterior capsular opacification 3 years after cataract surgery [J].
Hollick, EJ ;
Spalton, DJ ;
Ursell, PG ;
Pande, MV ;
Barman, SA ;
Boyce, JF ;
Tilling, K .
OPHTHALMOLOGY, 1999, 106 (01) :49-54
[9]  
KAPPELHOF JP, 1992, ACTA OPHTHALMOL, V70, P13
[10]   Posterior capsule opacification after implantation of a hydrophilic or a hydrophobic acrylic intraocular lens -: One-year follow-up [J].
Kugelberg, Maria ;
Wejde, Gisela ;
Jayaram, Hari ;
Zetterstrom, Charlotta .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (10) :1627-1631