A Monte Carlo simulation study on the effectiveness of electron filters designed for telecobalt radiation therapy treatment

被引:1
|
作者
Shukla, R. [1 ]
Patel, N. P. [2 ]
Yadav, H. P. [1 ]
Kausha, V [2 ]
机构
[1] GGS Med Coll, Dept Radiotherapy, Faridkot, India
[2] Pt BD Sharma Post Grad Inst Med Sci, Dept Radiotherapy, Rohtak, Haryana, India
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2019年 / 17卷 / 02期
关键词
Telecobalt; electron contamination; electron filters; surface dose; Monte Carlo simulation; CONTAMINATION; CO-60; LEAD; BEAM;
D O I
10.18869/acadpub.ijrr.17.2.217
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The aim of present study was to analyze the effectiveness of electron filters in the Telecobalt radiotherapy treatment by simulation technique. Materials and Methods: The BEAMnrc Monte Carlo code was used to simulate the electron filters of thickness of 0.5 gm/cm(2) below the trimmer bar for 35 x 35 cm(2) field size in Theratron Equinox-80 telecobalt unit. The electron filters were made of an aluminum, copper, nickel, tin, PMMA, and lead with single or composite materials. The radiation beams at treatment distance were analyzed by generating profiles for photon and electron along the X-axis of radiation field. Results: The electron energy fluence for unfiltered beam was 0.32% of the photon energy. The photon energy fluence intensity reduction due to filter was 3.7%. The filters with low atomic number have shown poor electron contamination removal efficiency. The tin, copper and nickel were found effective filters, removing nearly 38% of contaminant electron energy. The lead filter is equally effective as tin, however the high energy electrons emitted from filter due to "photo peaks" adds significant dose at 3.0 to 4.0 mm depth. Conclusion: The tin filter dominates over to other filters on the subject of surface dose reduction and depth of dose maximum (d(max)). It reduces the surface dose by 9.6% and 13.9 % of unfiltered beam for 15 x 15 and 20 x 20 cm(2) field sizes respectively.
引用
收藏
页码:217 / 227
页数:11
相关论文
共 50 条
  • [11] Analysis on the emission and potential application of Cherenkov radiation in boron neutron capture therapy: A Monte Carlo simulation study
    Shu, Di-Yun
    Geng, Chang-Ran
    Tang, Xiao-Bin
    Gong, Chun-Hui
    Shao, Wen-Cheng
    Ai, Yao
    APPLIED RADIATION AND ISOTOPES, 2018, 137 : 219 - 224
  • [12] DNA damage by radiation as a function of electron energy and interaction at the atomic level with Monte Carlo simulation
    Lamghari, Youssef
    Lu, Huizhong
    Bentourkia, Mhamed
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 489 - 498
  • [13] Monte Carlo Simulation of Backscattered and Secondary Electron Profiles
    Eisenschmidt, Christian
    Werner, Ulf
    MIKROCHIMICA ACTA, 1992, : 125 - 129
  • [14] Monte Carlo simulation of electron dynamics in liquid water
    Huthmacher, Klaus
    Herzwurm, Andre
    Gnewuch, Michael
    Ritter, Klaus
    Rethfeld, Baerbel
    PHYSICA A-STATISTICAL MECHANICS AND ITS APPLICATIONS, 2015, 429 : 242 - 251
  • [15] Monte Carlo simulation of electron slowing down in indium
    Rouabah, Z.
    Hannachi, M.
    Champion, C.
    Bouarissa, N.
    JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 2015, 202 : 7 - 10
  • [16] Monte Carlo simulation of projection electron beam lithography
    Sun, X.
    Xiao, P.
    Ding, Z. J.
    NANOSCIENCE AND TECHNOLOGY, PTS 1 AND 2, 2007, 121-123 : 1097 - 1101
  • [17] Monte Carlo simulation of electron emission from solids
    Kuhr, JC
    Fitting, HJ
    JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 1999, 105 (2-3) : 257 - 273
  • [18] Monte Carlo simulation of electron interaction with a thin film
    Stary, V
    THIN SOLID FILMS, 2003, 433 (1-2) : 326 - 331
  • [19] Monte Carlo based investigations of electron contamination from telecobalt unit head in build up region and its impact on surface dose
    Jagtap, A. S.
    Selvam, T. Palani
    Patil, B. J.
    Chavan, S. T.
    Pethe, S. N.
    Kulkarni, Gauri
    Dahiwale, S. S.
    Bhoraskar, V. N.
    Dhole, S. D.
    APPLIED RADIATION AND ISOTOPES, 2016, 118 : 175 - 181
  • [20] Monte Carlo characterization of photoneutrons in the radiation therapy with high energy photons: a Comparison between simplified and full Monte Carlo models
    Ghiasi, H.
    Mesbahi, A.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2010, 8 (03): : 187 - 193