A Monte Carlo simulation study on the effectiveness of electron filters designed for telecobalt radiation therapy treatment

被引:1
|
作者
Shukla, R. [1 ]
Patel, N. P. [2 ]
Yadav, H. P. [1 ]
Kausha, V [2 ]
机构
[1] GGS Med Coll, Dept Radiotherapy, Faridkot, India
[2] Pt BD Sharma Post Grad Inst Med Sci, Dept Radiotherapy, Rohtak, Haryana, India
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2019年 / 17卷 / 02期
关键词
Telecobalt; electron contamination; electron filters; surface dose; Monte Carlo simulation; CONTAMINATION; CO-60; LEAD; BEAM;
D O I
10.18869/acadpub.ijrr.17.2.217
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The aim of present study was to analyze the effectiveness of electron filters in the Telecobalt radiotherapy treatment by simulation technique. Materials and Methods: The BEAMnrc Monte Carlo code was used to simulate the electron filters of thickness of 0.5 gm/cm(2) below the trimmer bar for 35 x 35 cm(2) field size in Theratron Equinox-80 telecobalt unit. The electron filters were made of an aluminum, copper, nickel, tin, PMMA, and lead with single or composite materials. The radiation beams at treatment distance were analyzed by generating profiles for photon and electron along the X-axis of radiation field. Results: The electron energy fluence for unfiltered beam was 0.32% of the photon energy. The photon energy fluence intensity reduction due to filter was 3.7%. The filters with low atomic number have shown poor electron contamination removal efficiency. The tin, copper and nickel were found effective filters, removing nearly 38% of contaminant electron energy. The lead filter is equally effective as tin, however the high energy electrons emitted from filter due to "photo peaks" adds significant dose at 3.0 to 4.0 mm depth. Conclusion: The tin filter dominates over to other filters on the subject of surface dose reduction and depth of dose maximum (d(max)). It reduces the surface dose by 9.6% and 13.9 % of unfiltered beam for 15 x 15 and 20 x 20 cm(2) field sizes respectively.
引用
收藏
页码:217 / 227
页数:11
相关论文
共 50 条
  • [1] A novel approach in electron beam radiation therapy of lips carcinoma: A Monte Carlo study
    Shokrani, Parvaneh
    Baradaran-Ghahfarokhi, Milad
    Zadeh, Maryam Khorami
    MEDICAL PHYSICS, 2013, 40 (04)
  • [2] Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapy
    Shimozato, Tomohiro
    Okudaira, Kuniyasu
    Fuse, Hiraku
    Tabushi, Katsuyoshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (04): : 388 - 396
  • [3] Automated Monte Carlo Simulation of Proton Therapy Treatment Plans
    Verburg, Joost Mathijs
    Grassberger, Clemens
    Dowdell, Stephen
    Schuemann, Jan
    Seco, Joao
    Paganetti, Harald
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2016, 15 (06) : NP35 - NP46
  • [4] Feasibility assessment of the interactive use of a Monte Carlo algorithm in treatment planning for intraoperative electron radiation therapy
    Guerra, Pedro
    Udias, Jose M.
    Herranz, Elena
    Antonio Santos-Miranda, Juan
    Herraiz, Joaquin L.
    Valdivieso, Manlio F.
    Rodriguez, Raul
    Calama, Juan A.
    Pascau, Javier
    Calvo, Felipe A.
    Illana, Carlos
    Ledesma-Carbayo, Maria J.
    Santos, Andres
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (23) : 7159 - 7179
  • [5] Monte Carlo simulation study of electron interaction with solids and surfaces
    Ding, ZJ
    Salma, K
    Li, HM
    Zhang, ZM
    Tokesi, K
    Varga, D
    Toth, J
    Goto, K
    Shimizu, R
    SURFACE AND INTERFACE ANALYSIS, 2006, 38 (04) : 657 - 663
  • [6] Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study
    Nevelsky, Alexander
    Borzov, Egor
    Daniel, Shahar
    Bar-Deroma, Raquel
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (01): : 196 - 201
  • [7] A Monte Carlo simulation study on the image resolution in scanning electron microscopy
    Mao, S. F.
    Ding, Z. J.
    SURFACE AND INTERFACE ANALYSIS, 2010, 42 (6-7) : 1096 - 1099
  • [8] Hardware acceleration of a Monte Carlo simulation for photodynamic therapy treatment planning
    Lo, William Chun Yip
    Redmond, Keith
    Luu, Jason
    Chow, Paul
    Rose, Jonathan
    Lilge, Lothar
    JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (01)
  • [9] Macroscopic and microscopic investigation of maximum effectiveness of proton-boron capture therapy using Monte Carlo simulation
    Ganjeh, Zahra Ahmadi
    Mosleh-Shirazi, Mohammad Amin
    RADIATION PHYSICS AND CHEMISTRY, 2024, 214
  • [10] Monte Carlo simulation of cost-effectiveness ratios
    Filho, JAF
    Calil, SJ
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 595 - 598