Effect of Electrode to Modiolus Distance on Electrophysiological and Psychophysical Parameters in CI Patients With Perimodiolar and Lateral Electrode Arrays

被引:16
作者
Degen, Chantal Vanessa [1 ]
Buechner, Andreas [1 ]
Kludt, Eugen [1 ]
Lenarz, Thomas [1 ]
机构
[1] Hannover Med Sch, Dept Otorhinolaryngol, Hannover, Germany
关键词
Cochlear implant; Electrode design; Electrode– modiolus distance; Perimodiolar arrays; STAPEDIUS REFLEX; COCHLEAR;
D O I
10.1097/MAO.0000000000002751
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives: To investigate the intracochlear position of the latest Cochlear Nucleus 532 electrode array compared with the straight Nucleus 522 and the precurved 512 arrays and determine the effect of the electrode-modiolus distance on electrically evoked compound action potential, C-levels, electrically evoked stapedius reflex thresholds (ESRTs), and impedances. Methods: Postoperative high-resolution cone beam computational tomography images of 30 patients with Cochlear Nucleus 532, 522, and 512 implants were evaluated using the Comet (Cochlea Measurement Tool) program to determine the distance between the 22 individual electrode contacts and the medial wall. ESRTs were documented intraoperatively and electrophysiological as well as psychophysical parameters were measured at multiple time points including the first fitting after the initial activation. Results: The electrode-modiolus distance in perimodiolar arrays is uniformly small across the array, whereas in a straight electrode carrier it varies significantly along the length of the array. Electrically evoked compound action potential thresholds and C-levels are larger with increased distance to the modiolus. Impedances and stapedius reflex thresholds do not differ significantly between the arrays. Our results show that the electrode position has a significant effect on both electrophysiological and psychophysical parameters, while impedances and ESRTs are not impacted. Conclusion: Novel tools can be used in the evaluation of high resolution cone beam computational tomography images to determine individual electrode-modiolus distances after cochlear implantation. The results of this study suggest that the correlations between electrode-modiolus distance and electrophysiological and psychophysical parameters are not sufficiently strong to adjust CI-fitting based on imaging data.
引用
收藏
页码:E1091 / E1097
页数:7
相关论文
共 16 条
[1]   Clinical investigation of the Nucleus Slim Modiolar Electrode [J].
Aschendorff, Antje ;
Briggs, Robert ;
Brademann, Goetz ;
Helbig, Silke ;
Hornung, Joachim ;
Lenarz, Thomas ;
Marx, Mathieu ;
Ramos, Angel ;
Stoever, Timo ;
Escude, Bernard ;
James, Chris J. .
AUDIOLOGY AND NEURO-OTOLOGY, 2017, 22 (03) :169-179
[2]   Variations in Microanatomy of the Human Cochlea [J].
Avci, Ersin ;
Nauwelaers, Tim ;
Lenarz, Thomas ;
Hamacher, Volkmar ;
Kral, Andrej .
JOURNAL OF COMPARATIVE NEUROLOGY, 2014, 522 (14) :3245-3261
[3]   Audiological Outcome of the Pull-Back Technique in Cochlear Implantees [J].
Basta, Dietmar ;
Todt, Ingo ;
Ernst, Arne .
LARYNGOSCOPE, 2010, 120 (07) :1391-1396
[4]   Dose-Dependent Suppression of the Electrically Elicited Stapedius Reflex by General Anesthetics in Children Undergoing Cochlear Implant Surgery [J].
Crawford, Mark W. ;
White, Michelle C. ;
Propst, Evan J. ;
Zaarour, Christian ;
Cushing, Sharon ;
Pehora, Carolyne ;
James, Adrian L. ;
Gordon, Karen A. ;
Papsin, Blake C. .
ANESTHESIA AND ANALGESIA, 2009, 108 (05) :1480-1487
[5]   Is Electrode-Modiolus Distance a Prognostic Factor for Hearing Performances after Cochlear Implant Surgery? [J].
Esquia Medina, Gonzalo N. ;
Borel, Stephanie ;
Yann Nguyen ;
Ambert-Dahan, Emmanuele ;
Ferrary, Evelyne ;
Sterkers, Olivier ;
Grayeli, Alexis Bozorg .
AUDIOLOGY AND NEURO-OTOLOGY, 2013, 18 (06) :406-413
[6]   Factors Affecting Open-Set Word Recognition in Adults With Cochlear Implants [J].
Holden, Laura K. ;
Finley, Charles C. ;
Firszt, Jill B. ;
Holden, Timothy A. ;
Brenner, Christine ;
Potts, Lisa G. ;
Gotter, Brenda D. ;
Vanderhoof, Sallie S. ;
Mispagel, Karen ;
Heydebrand, Gitry ;
Skinner, Margaret W. .
EAR AND HEARING, 2013, 34 (03) :342-360
[7]   Current focussing in cochlear implants: An analysis of neural recruitment in a computational model [J].
Kalkman, Randy K. ;
Briaire, Jeroen J. ;
Frijns, Johan H. M. .
HEARING RESEARCH, 2015, 322 :89-98
[8]   3D finite element analyses of insertion of the Nucleus standard straight and the Contour electrode arrays into the human cochlea [J].
Kha, H. N. ;
Chen, B. K. ;
Clark, G. M. .
JOURNAL OF BIOMECHANICS, 2007, 40 (12) :2796-2805
[9]   Visualization, measurement and modelling of the cochlea using rotating midmodiolar slice planes [J].
Lexow, G. Jakob ;
Schurzig, Daniel ;
Gellrich, Nils-Claudius ;
Lenarz, Thomas ;
Majdani, Omid ;
Rau, Thomas S. .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2016, 11 (10) :1855-1869
[10]   The Value of Electrically Evoked Stapedius Reflex in Determining the Maximum Comfort Level of a Cochlear Implant [J].
Lira de Andrade, Kelly Cristina ;
Muniz, Lilian Ferreira ;
Menezes, Pedro de Lemos ;
Caldas Neto, Silvio da Silva ;
Lins Carnauba, Aline Tenorio ;
Leal, Mariana de Carvalho .
JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2018, 29 (04) :292-299