New metallic implant used for permanent contraception in women: Evaluation of MR safety

被引:17
作者
Shellock, FG [1 ]
机构
[1] Inst Magnet Resonance Safety Educ & Res, Los Angeles, CA 90045 USA
关键词
D O I
10.2214/ajr.178.6.1781513
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. This ex vivo investigation evaluated the safety of using MR imaging with a new metallic implant designed to provide permanent birth control. CONCLUSION. The findings indicated that it should be safe for patients with this metallic contraceptive implant to undergo MR imaging with systems using static magnetic fields of 1.5 T or less.
引用
收藏
页码:1513 / 1516
页数:4
相关论文
共 13 条
[1]   POTENTIAL HAZARDS IN NMR IMAGING - HEATING EFFECTS OF CHANGING MAGNETIC-FIELDS AND RF FIELDS ON SMALL METALLIC IMPLANTS [J].
DAVIS, PL ;
CROOKS, L ;
ARAKAWA, M ;
MCREE, R ;
KAUFMAN, L ;
MARGULIS, AR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) :857-860
[2]   Safety of intrauterine contraceptive devices during MR imaging [J].
Hess, T ;
Stepanow, B ;
Knopp, MV .
EUROPEAN RADIOLOGY, 1996, 6 (01) :66-68
[3]   INTRAUTERINE CONTRACEPTIVE DEVICES - MR IMAGING [J].
MARK, AS ;
HRICAK, H .
RADIOLOGY, 1987, 162 (02) :311-314
[4]   Heating near implanted medical devices by the MRI RF-magnetic field [J].
Nyenhuis, JA ;
Kildishev, AV ;
Bourland, JD ;
Foster, KS ;
Graber, G ;
Athey, TW .
IEEE TRANSACTIONS ON MAGNETICS, 1999, 35 (05) :4133-4135
[5]   CARDIOVASCULAR ASPECTS OF HUMAN THERMOREGULATION [J].
ROWELL, LB .
CIRCULATION RESEARCH, 1983, 52 (04) :367-379
[6]   The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds [J].
Schenck, JF .
MEDICAL PHYSICS, 1996, 23 (06) :815-850
[7]  
Shellock FG, 2000, J MAGN RESON IMAGING, V12, P30, DOI 10.1002/1522-2586(200007)12:1<30::AID-JMRI4>3.0.CO
[8]  
2-S
[9]  
Shellock FG, 2001, MAGNETIC RESONANCE PROCEDURES: HEALTH EFFECTS AND SAFETY, P271
[10]   Metallic neurosurgical implants: Evaluation of magnetic field interactions, heating, and artifacts at 1.5-tesla [J].
Shellock, FG .