Clinical introduction of Monte Carlo treatment planning for lung stereotactic body radiotherapy

被引:29
|
作者
Miura, Hideharu [1 ]
Masai, Norihisa [1 ]
Oh, Ryoong-Jin [1 ]
Shiomi, Hiroya [1 ]
Yamada, Kouichi [1 ]
Sasaki, Junichi [1 ]
Inoue, Toshihiko [1 ]
机构
[1] Miyakojima IGRT Clin, Osaka 5340021, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2014年 / 15卷 / 01期
关键词
Monte Carlo dose calculation; lung density; GTV size; dose prescription; stereotactic body radiation therapy; RADIATION-THERAPY; DOSIMETRIC EVALUATION; DOSE CALCULATION; CANCER; VERIFICATION; ALGORITHMS; FRACTIONS; OUTCOMES; VOLUME; GY;
D O I
10.1120/jacmp.v15i1.4202
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to investigate the impact of Monte Carlo (MC) calculations and optimized dose definitions in stereotactic body radiotherapy (SBRT) for lung cancer patients. We used a retrospective patient review and basic virtual phantom to determine dose prescriptions. Fifty-three patients underwent SBRT. A basic virtual phantom had a gross tumor volume (GTV) of 10.0 mm with equivalent water density of 1.0 g/cm(3), which was surrounded by equivalent lung surrounding the GTV of 0.25 g/cm(3). D95 of the planning target volume (PTV) and D99 of the GTV were evaluated with different GTV sizes (5.0 to 30.0 mm) and different lung densities (0.05 to 0.45 g/cm(3)). Prescribed dose was defined as 95% of the PTV should receive 100% of the dose (48 Gy/4 fractions) using pencil beam (PB) calculation and recalculated using MC calculation. In the patient study, average doses to the D95 of the PTV and D99 of the GTV using the MC calculation plan were 19.9% and 10.2% lower than those by the PB calculation plan, respectively. In the phantom study, decreased doses to the D95 of the PTV and D99 of the GTV using the MC calculation plan were accompanied with changes GTV size from 30.0 to 5.0 mm, which was decreased from 8.4% to 19.6% for the PTV and from 17.4% to 27.5% for the GTV. Similar results were seen with changes in lung density from 0.45 to 0.05 g/cm(3), with doses to the D95 of the PTV and D99 of the GTV were decreased from 12.8% to 59.0% and from 7.6% to 44.8%, respectively. The decrease in dose to the PTV with MC calculation was strongly dependent on lung density. We suggest that dose definition to the GTV for lung cancer SBRT be optimized using MC calculation. Our current clinical protocol for lung SBRT is based on a prescribed dose of 44 Gy in 4 fractions to the GTV using MC calculation.
引用
收藏
页码:38 / 46
页数:9
相关论文
共 50 条
  • [41] Correlation of Clinical and Dosimetric Parameters With Radiographic Lung Injury Following Stereotactic Body Radiotherapy
    Kishan, Amar U.
    Wang, Pin-Chieh
    Sheng, Ke
    Yu, Victoria
    Ruan, Dan
    Cao, Minsong
    Tenn, Stephen
    Low, Daniel A.
    Lee, Percy
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2015, 14 (04) : 411 - 418
  • [42] Stereotactic Body Radiotherapy for Lung Tumors at the Pulmonary Hilum
    Oshiro, Yoshiko
    Aruga, Takashi
    Tsuboi, Koji
    Marino, Kan
    Hara, Ryusuke
    Sanayama, Yasushi
    Itami, Jun
    STRAHLENTHERAPIE UND ONKOLOGIE, 2010, 186 (05) : 274 - 279
  • [43] Stereotactic Body Radiotherapy for Early Stage Lung Cancer
    Nagata, Yasushi
    CANCER RESEARCH AND TREATMENT, 2013, 45 (03): : 155 - 161
  • [44] LungTech, an EORTC Phase II trial of stereotactic body radiotherapy for centrally located lung tumours: a clinical perspective
    Adebahr, S.
    Collette, S.
    Shash, E.
    Lambrecht, M.
    Le Pechoux, C.
    Faivre-Finn, C.
    De Ruysscher, D.
    Peulen, H.
    Belderbos, J.
    Dziadziuszko, R.
    Fink, C.
    Guckenberger, M.
    Hurkmans, C.
    Nestle, U.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1051):
  • [45] A novel CRT-IMRT-combined (Co-CRIM) planning technique for peripheral lung stereotactic body radiotherapy in pinnacle treatment planning system
    Duan, Yanhua
    Zhou, Lijun
    Wang, Hao
    Chen, Hua
    Gu, Hengle
    Shao, Yan
    Feng, Aihui
    Huang, Ying
    Fu, Xiaolong
    Yue, Ning Jeff
    Ma, Kui
    Kong, Qing
    Xu, Zhiyong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 97 - 107
  • [46] The HILUS-Trial-a Prospective Nordic Multicenter Phase 2 Study of Ultracentral Lung Tumors Treated With Stereotactic Body Radiotherapy
    Lindberg, Karin
    Grozman, Vitali
    Karlsson, Kristin
    Lindberg, Sara
    Lax, Ingmar
    Wersall, Peter
    Persson, Gitte Fredberg
    Josipovic, Mirjana
    Khalil, Azza Ahmed
    Moeller, Ditte Sloth
    Nyman, Jan
    Drugge, Ninni
    Bergstrom, Per
    Olofsson, Jorgen
    Rogg, Lotte Victoria
    Ramberg, Christina
    Kristiansen, Charlotte
    Jeppesen, Stefan Starup
    Nielsen, Tine Bjorn
    Loden, Britta
    Rosenbrand, Hans-Olov
    Engelholm, Silke
    Haraldsson, Andre
    Billiet, Charlotte
    Lewensohn, Rolf
    JOURNAL OF THORACIC ONCOLOGY, 2021, 16 (07) : 1200 - 1210
  • [47] Dose calculation differences between Monte Carlo and pencil beam depend on the tumor locations and volumes for lung stereotactic body radiation therapy
    Zhuang, Tingliang
    Djemil, Toufik
    Qi, Peng
    Magnelli, Anthony
    Stephans, Kevin
    Videtic, Gregory
    Xia, Ping
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 38 - 51
  • [48] A phase I/II study on stereotactic body radiotherapy with real-time tumor tracking using CyberKnife based on the Monte Carlo algorithm for lung tumors
    Iwata, Hiromitsu
    Ishikura, Satoshi
    Murai, Taro
    Iwabuchi, Michio
    Inoue, Mitsuhiro
    Tatewaki, Koshi
    Ohta, Seiji
    Yokota, Naoki
    Shibamoto, Yuta
    INTERNATIONAL JOURNAL OF CLINICAL ONCOLOGY, 2017, 22 (04) : 706 - 714
  • [49] Comparison of 3D and 4D Monte Carlo optimization in robotic tracking stereotactic body radiotherapy of lung cancer
    Chan, Mark K. H.
    Werner, Rene
    Ayadi, Miriam
    Blanck, Oliver
    STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (02) : 161 - 171
  • [50] Dosimetric accuracy of Acuros® XB and AAA algorithms for stereotactic body radiotherapy (SBRT) lung treatments: evaluation with PRIMO Monte Carlo code
    Sarin, B.
    Bindhu, B.
    Kumar, P. Raghu
    Sumeesh, S.
    Saju, B.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 22