Error Analysis of FBG-Based Shape Sensors for Medical Needle Tracking

被引:85
作者
Henken, Kirsten R. [1 ]
Dankelman, Jenny [1 ]
van den Dobbelsteen, John J. [1 ]
Cheng, Lun K. [2 ]
van der Heiden, Maurits S. [2 ]
机构
[1] Delft Univ Technol, Dept Biomech Engn, NL-2628 CN Delft, Netherlands
[2] TNO, Dept Tech Sci Opto Mechatron, NL-2628 CK Delft, Netherlands
关键词
Fiber Bragg grating (FBG); medical robotics; needle; shape sensing; PROSTATE BIOPSY; ACCURACY;
D O I
10.1109/TMECH.2013.2287764
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Robotic needle steering requires accurate spatial information about the needle tip. Presumably, fiber Bragg gratings (FBGs) can provide this information at an appropriate update rate. We performed an extensive error analysis to quantify the accuracy of needle tip tracking with FBGs and to assess the suitability of this method for robotic needle steering. An FBG-based shape sensing model was determined and simulations were performed to quantify the effect of design parameters on the position accuracy. Inputs that were investigated include accuracy of wavelength measurement and sensor geometry as well as different sensor configurations and interpolation models. For the purpose of validation of the simulations, two needles with two different configurations of FBGs were built and evaluated. The simulations show that the accuracy of FBG-based shape sensing of a needle can be in the order of 10% of the deflection at the tip, depending on the configuration. However, tip deflections that are smaller than approximately 1 mm cannot be detected accurately. Calibration of the needle reduces the bias, but does not improve the accuracy, because of drift in read-out of the FGBs. The analysis shows that the combined sources of errors limit the accuracy of tip estimation to approximately 1 mm, although the accuracy is influenced by the sensor configuration as well. This accuracy is suitable for common medical applications like taking biopsies or performing ablation.
引用
收藏
页码:1523 / 1531
页数:9
相关论文
共 22 条
[1]   Needle insertion into soft tissue: A survey [J].
Abolhassani, Niki ;
Patel, Rajni ;
Moallem, Mehrdad .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (04) :413-431
[2]   Active Microcoil Tracking in the Lungs Using a Semisolid Rubber as Signal Source [J].
Alt, Stefan ;
Homagk, Ann-Kathrin ;
Umathum, Reiner ;
Semmler, Wolfhard ;
Bock, Michael .
MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (01) :271-279
[3]   Integrated active tracking detector for MRI-guided interventions [J].
Anders, Jens ;
SanGiorgio, Paul ;
Deligianni, Xeni ;
Santini, Francesco ;
Scheffler, Klaus ;
Boero, Giovanni .
MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (01) :290-296
[4]   Automatic passive tracking of an endorectal prostate biopsy device using phase-only cross-correlation [J].
de Oliveira, Andre ;
Rauschenberg, Jaane ;
Beyersdorff, Dirk ;
Semmler, Wolfhard ;
Bock, Michael .
MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) :1043-1050
[5]   Needle steering and motion planning in soft tissues [J].
DiMaio, SP ;
Salcudean, SE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (06) :965-974
[6]   Image-guided robotic flexible needle steering [J].
Glozman, Daniel ;
Shoham, Moshe .
IEEE TRANSACTIONS ON ROBOTICS, 2007, 23 (03) :459-467
[7]  
Gorlitz R. A., 2010, P SOC PHOTO-OPT INS, V7625, DOI [10.1117/12/844332, DOI 10.1117/12/844332]
[8]  
He X. C., 2012, OPTICAL FIBERS SENSO, V8218
[9]   Accuracy of needle position measurements using fiber Bragg gratings [J].
Henken, Kirsten ;
Van Gerwen, Dennis ;
Dankelman, Jenny ;
Van Den Dobbelsteen, John .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2012, 21 (06) :408-414
[10]   Fiber Bragg grating technology fundamentals and overview [J].
Hill, KO ;
Meltz, G .
JOURNAL OF LIGHTWAVE TECHNOLOGY, 1997, 15 (08) :1263-1276