Comparison of red and green laser Doppler imaging of blood flow

被引:32
|
作者
Murray, AK
Gorodkin, RE
Moore, TL
Gush, RJ
Herrick, AL
King, TA
机构
[1] Univ Manchester, Dept Phys & Astron, Laser Photon Grp, Manchester M13 9PL, Lancs, England
[2] Hope Hosp, Ctr Rheumat Dis, Salford M6 8HD, Lancs, England
[3] Univ Manchester, Arc Epidemil Unit, Manchester M13 9PL, Lancs, England
[4] Moor Instruments, Millwey EX13 5HU, Axminster, England
关键词
laser doppler; vein; perfusion; occlusion;
D O I
10.1002/lsm.20085
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and Objectives: Laser Doppler imaging (LDI) of perfusion has been performed with a novel green wavelength (532 nm) for comparison with a HeNe laser (633 nm), the aim being validation of the green laser wavelength as a research tool. Study Design/Materials and Methods: The effect of wavelength and power on images was investigated and perfusion response following both finger occlusion and local heating of the dorsum were examined as reproducible stimuli for clinical studies. Results: The most striking difference between red and green LDI is the absence of veins on green LDI, which are seen with red LDI. Differences have been quantified using vein LDI profiles. Differences were found between blood flow responses imaged by red and green LDI (3 and 5 mW, respectively) for occlusion and heat stimuli. Results are discussed in the context of light penetration. Conclusions: Red and green wavelengths appear to image different components of the microcirculation. Lasers Surg. Med. 35:191-200, 2004. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:191 / 200
页数:10
相关论文
共 50 条
  • [1] Comparison between laser speckle contrast imaging and laser Doppler imaging to assess skin blood flow in humans
    Millet, C.
    Roustit, M.
    Blaise, S.
    Cracowski, J. L.
    MICROVASCULAR RESEARCH, 2011, 82 (02) : 147 - 151
  • [2] Holographic laser Doppler imaging of microvascular blood flow
    Magnain, C.
    Castel, A.
    Boucneau, T.
    Simonutti, M.
    Ferezou, I.
    Rancillac, A.
    Vitalis, T.
    Sahel, J. A.
    Paques, M.
    Atlan, M.
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2014, 31 (12) : 2723 - 2735
  • [3] Holographic laser Doppler imaging of pulsatile blood flow
    Bencteux, Jeffrey
    Pagnoux, Pierre
    Kostas, Thomas
    Bayat, Sam
    Atlan, Michael
    JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (06)
  • [4] High-speed laser Doppler imaging of blood flow in biological tissue
    Serov, Alexandre
    Lasser, Theo
    SARATOV FALL MEETING 2005: OPTICAL TECHNOLOGIES IN BIOPHYSICS AND MEDICINE VII, 2006, 6163
  • [5] Cortical blood flow autoregulation revisited using laser Doppler perfusion imaging
    Kimme, P
    Ledin, T
    Sjöberg, F
    ACTA PHYSIOLOGICA SCANDINAVICA, 2002, 176 (04): : 255 - 262
  • [6] Evaluation of laser Doppler imaging to measure blood flow in knee ligaments of adult rabbits
    Bray, R
    Forrester, K
    McDougall, JJ
    Damji, A
    Ferrell, WR
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1996, 34 (03) : 227 - 231
  • [7] Laser Doppler Blood Flow Imaging Using a CMOS Imaging Sensor with On-Chip Signal Processing
    He, Diwei
    Nguyen, Hoang C.
    Hayes-Gill, Barrie R.
    Zhu, Yiqun
    Crowe, John A.
    Gill, Cally
    Clough, Geraldine F.
    Morgan, Stephen P.
    SENSORS, 2013, 13 (09) : 12632 - 12647
  • [8] Laser Doppler, speckle and related techniques for blood perfusion mapping and imaging
    Briers, JD
    PHYSIOLOGICAL MEASUREMENT, 2001, 22 (04) : R35 - R66
  • [9] Comparison of laser Doppler and laser speckle contrast imaging using a concurrent processing system
    Sun, Shen
    Hayes-Gill, Barrie R.
    He, Diwei
    Zhu, Yiqun
    Huynh, Nam T.
    Morgan, Stephen P.
    OPTICS AND LASERS IN ENGINEERING, 2016, 83 : 1 - 9
  • [10] Laser speckle contrast imaging for measuring blood flow
    Briers, J. David
    OPTICA APPLICATA, 2007, 37 (1-2) : 139 - 152