In vitro performance of zirconia and titanium implant/abutment systems for anterior application

被引:42
作者
Rosentritt, Martin [1 ]
Hagemann, Anna [1 ]
Hahnel, Sebastian [1 ]
Behr, Michael [1 ]
Preis, Verena [1 ]
机构
[1] Univ Med Ctr Regensburg, Dept Prosthet Dent, D-93042 Regensburg, Germany
关键词
Implant; Abutment; Zirconia; Titanium; Chewing simulation; Fracture resistance; ALL-CERAMIC RESTORATIONS; FRACTURE-RESISTANCE; SINGLE-TOOTH; IMPLANT ABUTMENTS; DENTAL IMPLANTS; STRENGTH; FAILURE; SUCCESS; LOAD;
D O I
10.1016/j.jdent.2014.03.010
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To investigate the type of failure and fracture resistance behaviour of different zirconia and titanium implant/abutment systems for anterior application. Methods: Eight groups of implant-abutment combinations (n = 8/system) were restored with identical full-contour zirconia crowns. The systems represented one-piece and multi-piece zirconia (Z) or titanium (T) implants/abutments with different types of connection (screwed = 5, bonded = B). The following combinations (implant-abutment-connection) were investigated: ZZS, ZZB, ZZZB (three-piece), ZTS, TTS, TTS reference, and Z (one-piece, 2x). To simulate clinical anterior loading situations the specimens were mounted into the chewing simulator at an angle of 135 degrees and subjected to thermal cycling (2 x 3000 x 5 degrees/ 55 degrees C) and mechanical loading (1.2 x 10(6) x 50 N; 1.6 Hz). Fracture resistance and maximum bending stress were determined for all specimens that survived ageing. Data were statistically analyzed with the Kolmogorov-Smimov-test and one-way ANOVA (alpha = 0.05). Survival performance was calculated with the Kaplan-Meier Log-Rank test. Results: Independent of the material combinations screwed systems showed partly failures of the screws during simulation (ZZS: 3x, ZTS: 8x, TTS: 3x). Screw failures were combined with implant/abutment fractures of zirconia systems. Zirconia one-piece implants and the reference system did not show any failures, and only one specimen of the systems with a bonded connection (ZZZB) fractured. Mean (+/- standard deviation) fracture forces and maximum bending stresses differed significantly (p = 0.000) between 187.4 +/- 42.0 N/ 250.0 +/- 56.0 N/mm(2) (ZZZB) and 524.3 +/- 43.1 N/753.0 +/- 61.0 N/mm(2) (Z). Conclusions: Both material (zirconia or titanium) and the type of connection influenced failure resistance during fatigue testing, fracture force, and maximum bending stress. Clinical significance: Different material combinations for implants and abutments as well as different types of connection achieved acceptable or even good failure and fracture resistance that may be satisfactory for anterior clinical application. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1019 / 1026
页数:8
相关论文
共 46 条
  • [1] Clinical trials in zirconia: a systematic review
    Al-Amleh, B.
    Lyons, K.
    Swain, M.
    [J]. JOURNAL OF ORAL REHABILITATION, 2010, 37 (08) : 641 - 652
  • [2] Fracture Strength of Zirconia Implants after Artificial Aging
    Andreiotelli, Marina
    Kohal, Ralf-Joachim
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2009, 11 (02) : 158 - 166
  • [3] Are ceramic implants a viable alternative to titanium implants? A systematic literature review
    Andreiotelli, Marina
    Wenz, Hans J.
    Kohal, Ralf-Joachim
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 : 32 - 47
  • [4] Fracture resistance of single-tooth implant-supported all-ceramic restorations after exposure to the artificial mouth
    Att, W
    Kurun, S
    Gerds, T
    Strub, JR
    [J]. JOURNAL OF ORAL REHABILITATION, 2006, 33 (05) : 380 - 386
  • [5] Fracture resistance of single-tooth implant-supported all-ceramic restorations: An in vitro study
    Att, W
    Kurun, S
    Gerds, T
    Strub, JR
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2006, 95 (02) : 111 - 116
  • [6] Aydin C, 2013, J PROSTHET DENT, V109, P70, DOI 10.1016/S0022-3913(13)00027-9
  • [7] In vitro performance of full-contour zirconia single crowns
    Beuer, Florian
    Stimmelmayr, Michael
    Gueth, Jan-Frederik
    Edelhoff, Daniel
    Naumann, Michael
    [J]. DENTAL MATERIALS, 2012, 28 (04) : 449 - 456
  • [8] Bianchi A E, 2004, J Appl Biomater Biomech, V2, P143
  • [10] Biting and chewing in overdentures, full dentures, and natural dentitions
    Fontijn-Tekamp, FA
    Slagter, AP
    Van Der Bilt, A
    Van 'T Hof, MA
    Witter, DJ
    Kalk, W
    Jansen, JA
    [J]. JOURNAL OF DENTAL RESEARCH, 2000, 79 (07) : 1519 - 1524