Polytetrafluoroethylene-encapsulated stent-grafts: Use in experimental abdominal aortic aneurysm

被引:8
|
作者
Benson, AE [1 ]
Palmaz, JC [1 ]
Tio, FO [1 ]
Sprague, EA [1 ]
Encarnacion, CE [1 ]
Josephs, SC [1 ]
机构
[1] Univ Texas, Hlth Sci Ctr, Dept Radiol, Cardiovasc & Special Intervent Sect, San Antonio, TX 78284 USA
关键词
aneurysm; therapy; aorta; endovascular stent grafts; stents and prostheses;
D O I
10.1016/S1051-0443(99)70091-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate expanded polytetrafluoroethylene (ePTFE) encapsulated stents for the treatment of aortic aneurysms with emphasis on the blood and tissue-material interactions. MATERIALS AND METHODS: Experimental aortic aneurysms were created in dogs by enlarging the aortic lumen with an abdominal fascial patch. Twenty animals underwent endoluminal repair after allowing the surgically created aneurysm to heal for 2 months prior to transluminal aneurysmal exclusion. The device used consisted of an 8-cm-long ePTFE encapsulated stent graft. The animals were killed in groups at 1 week and at 1, 2.25, 6, and 12 months. Specimens were processed for histologic and luminal surface studies. RESULTS: Before the animals were killed, aortography demonstrated two thrombosed aortae in the 6-month group and two endoleaks in the 12-month group. Endothelialized neointima extended into the proximal and distal portions of the prosthetic lumen, with minimal cell coverage in the center of the graft. The overall percent surface area covered by endothelialized neointima was 22% +/- 6% at 6 months and 18% +/- 10% by 1 year (P = .75), Histologic examination demonstrated minimal tisue penetration into the ePTFE, CONCLUSION: Transluminal exclusion of abdominal aortic aneurysms by encapsulated stent-graft is easily acomplished, With this device, tissue coverage and penetration of the stent graft is limited and does not tend to increase with time.
引用
收藏
页码:605 / 612
页数:8
相关论文
共 50 条
  • [31] Abdominal aortic aneurysms: Preliminary technical and clinical results with transfemoral placement of endovascular self-expanding stent-grafts
    Blum, U
    Langer, M
    Spillner, G
    Mialhe, C
    Beyersdorf, F
    BuitragoTellez, C
    Voshage, G
    Duber, C
    Schlosser, V
    Cragg, AH
    RADIOLOGY, 1996, 198 (01) : 25 - 31
  • [32] Suitability of Endovascular Repair with Current Stent Grafts for Abdominal Aortic Aneurysm in Korean Patients
    Park, Kay-Hyun
    Lim, Cheong
    Lee, Jae Hang
    Yoo, Jae Suk
    JOURNAL OF KOREAN MEDICAL SCIENCE, 2011, 26 (08) : 1047 - 1051
  • [33] Branched Stent-Grafts for Endovascular Repair of Aortic and Iliac Aneurysms
    Chuter, Timothy A. M.
    TECHNIQUES IN VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 8 (01) : 56 - 60
  • [34] Elective endovascular treatment of descending thoracic aortic aneurysms and chronic dissections with stent-grafts
    Won, JY
    Lee, DY
    Shim, WH
    Chang, BC
    Park, SI
    Yoon, CS
    Kwon, HM
    Park, BH
    Jung, GS
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (05) : 575 - 582
  • [35] Use of through-and-through guidewire for delivering large stent-grafts into the distal aortic arch
    Muhammad Al Shammari
    Peter Taylor
    John F. Reidy
    CardioVascular and Interventional Radiology, 2000, 23 : 237 - 238
  • [36] Midterm outcomes of thoracic aortic stent-grafts: Complications and imaging techniques
    Chabbert, V
    Otal, P
    Bouchard, L
    Soula, P
    Van, TT
    Kos, X
    Meites, G
    Claude, C
    Joffre, F
    Rousseau, H
    JOURNAL OF ENDOVASCULAR THERAPY, 2003, 10 (03) : 494 - 504
  • [37] Acute rupture of the descending thoracic aorta: Repair with use of endovascular stent-grafts
    Semba, CP
    Kato, N
    Kee, ST
    Lee, GK
    Mitchell, RS
    Miller, DC
    Dake, MD
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (03) : 337 - 342
  • [38] Treatment of aortoiliac aneurysms with use of single-piece tapered stent-grafts
    Kato, N
    Dake, MD
    Semba, CP
    Razavi, MK
    Kee, ST
    Slonim, SM
    Samuels, SLW
    Terasaki, KK
    Zarins, CK
    Mitchell, RS
    Miller, DC
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 9 (01) : 41 - 49
  • [39] Macroporous polyester-covered stent in an experimental abdominal aortic aneurysm model
    Soula, P
    d'Othée, BJ
    Otal, P
    Amin, C
    El Khoury, J
    Delisle, MB
    Cérène, A
    Joffre, F
    Rousseau, H
    JOURNAL OF ENDOVASCULAR THERAPY, 2001, 8 (04) : 390 - 400
  • [40] Endovascular treatment of complicated type-B aortic dissection with stent-grafts:: Midterm results
    Lopera, J
    Patiño, JH
    Urbina, C
    García, G
    Alvarez, LG
    Upegui, L
    Jhanchai, A
    Qian, Z
    Castañeda-Zuñiga, W
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (02) : 195 - 203