Steerable Guidewire for Magnetic Resonance Guided Endovascular Interventions

被引:15
作者
Clogenson, H. C. M. [1 ]
Dankelman, J. [1 ]
van den Dobbelsteen, J. J. [1 ]
机构
[1] Delft Univ Technol, Dept Biomech Engn, NL-2628 CD Delft, Netherlands
来源
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME | 2014年 / 8卷 / 02期
关键词
interventional radiology; navigation; steerability; magnetic resonance imaging; COMPATIBLE GUIDEWIRE; IN-VITRO; MRI; CATHETERIZATION; FEASIBILITY; ANGIOPLASTY; ARTERIES; SWINE; WIRE;
D O I
10.1115/1.4026560
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In endovascular interventions, thin, flexible instruments are inserted through the skin into the blood vessels to diagnose and treat various diseases of the vascular system. One drawback is that the instruments are difficult to maneuver in the desired direction due to limitations in shape and flexibility. Another disadvantage is that the interventions are performed under intermittent fluoroscopy/angiography imaging. Magnetic resonance imaging (MRI) may offer advantages over X-ray guidance. It presents a good soft tissue contrast without the use of nephrotoxic media or ionizing radiation. The aim of this study is to develop a guidewire that is compatible with MRI and includes a steerable segment at the tip. This added degree-of-freedom may improve the maneuverability of the devices thereby the efficiently and safety of the navigation. A 1.6m (5 ft, 3 in.) long and 0.035 in. diameter guidewire that consists of MR compatible materials and has a flexible tip was designed. The only metallic part was a nitinol rod that was implemented at the distal flexible tip. To limit the risk of heating in the MRI, this rod was kept shorter than 30 mm. The tip could be deflected in one direction by pulling on a Dyneema wire that was placed in the lumen of the shaft of the guidewire. To drive the steerable tip, a handle that could be easily attached/detached from the instrument was designed and implemented. Using the handle, the tip of the 1.60m long guidewire prototype could be actuated to reach angles from 30 deg to 250 deg. The handle could easily be placed on and removed from the guidewire, so conventional 0.035 in.-compatible catheters could slide over from the proximal end. However, in order to make the guidewire more efficient to enter a bifurcation, the stiffness of the tip should progressively increase from its proximal to its distal end. The guidewire was imaged in a 1.5T MRI using real-time imaging without producing artifacts that would have shaded the anatomy. It was possible to assemble a guidewire with a steerable segment in the required size, using MR compatible materials. Therefore, the current design is a promising proof of concept and allowed us to clearly identify the features that need to be improved in order to come to a clinically applicable instrument.
引用
收藏
页数:7
相关论文
共 21 条
  • [1] MR-guided balloon angioplasty: In vitro demonstration of the potential of MRI for guiding, monitoring, and evaluating endovascular interventions
    Bakker, CJ
    Smits, HF
    Bos, C
    van der Weide, R
    Zuiderveld, KJ
    van Vaals, JJ
    Hurtak, WF
    Viergever, MA
    Mall, WP
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01): : 245 - 250
  • [2] Evaluation of vascular and interventional procedures with time-action analysis: A pilot study
    Bakker, NH
    Tanase, D
    Reekers, JA
    Grimbergen, CA
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (05) : 483 - 488
  • [3] Use of a nonmetallic guide wire for magnetic resonance-guided coronary artery catheterization
    Buecker, A
    Spuentrup, E
    Schmitz-Rode, T
    Kinzel, S
    Pfeffer, J
    Hohl, C
    van Vaals, JJ
    Günther, RW
    [J]. INVESTIGATIVE RADIOLOGY, 2004, 39 (11) : 656 - 660
  • [4] Safety of MRI-guided vascular interventions
    Buecker, Arno
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (02) : 65 - 70
  • [5] Does virtual reality training make sense in interventional radiology? Training skill-, rule and knowledge-based behavior
    Dankelman, J
    Wentink, M
    Grimbergen, CA
    Stassen, HG
    Reekers, J
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 27 (05) : 417 - 421
  • [6] Steerable catheters in minimally invasive vascular surgery
    Fu, Yili
    Liu, Hao
    Huang, Wentai
    Wang, Shuguo
    Liang, Zhaoguang
    [J]. INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (04) : 381 - 391
  • [7] MR-compatible polyetheretherketone-based guide wire assisting MR-guided stenting of iliac and supraaortic arteries in swine: Feasibility study
    Kos, Sebastian
    Huegli, Rolf
    Hofmann, Eugen
    Quick, Harald H.
    Kuehl, Hilmar
    Aker, Stephanie
    Kaiser, Gernot M.
    Borm, Paul J. A.
    Jacob, Augustinus L.
    Bilecen, Deniz
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2009, 18 (03) : 181 - 188
  • [8] Feasibility of Real-Time Magnetic Resonance-Guided Angioplasty and Stenting of Renal Arteries in Vitro and in Swine, Using a New Polyetheretherketone-Based Magnetic Resonance-Compatible Guidewire
    Kos, Sebastian
    Huegli, Rolf
    Hofmann, Euggen
    Quick, Harald H.
    Kuehl, Hilmar
    Aker, Stephanie
    Kaiser, Gernot M.
    Borm, Paul J.
    Jacob, Augustinus L.
    Bilecen, Deniz
    [J]. INVESTIGATIVE RADIOLOGY, 2009, 44 (04) : 234 - 241
  • [9] Krombach GA, 2010, 8 INT MRI S LEIPZ GE, P128
  • [10] An MR Guidewire Based on Micropultruded Fiber-Reinforced Material
    Krueger, Sascha
    Schmitz, Sebastian
    Weiss, Steffen
    Wirtz, Daniel
    Linssen, Marita
    Schade, Heinz
    Kraemer, Nils
    Spuentrup, Elmar
    Krombach, Gabriele
    Buecker, Arno
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (05) : 1190 - 1196