Applicability of cloud point extraction for the separation trace amount of lead ion in environmental and biological samples prior to determination by flame atomic absorption spectrometry

被引:28
作者
Mohammadi, Sayed Zia [1 ,2 ]
Shamspur, Tayebeh [3 ]
Afzali, Daryoush [4 ]
Taher, Mohammad Ali [3 ]
Baghelani, Yar Mohammad [2 ]
机构
[1] Payame Noor Univ, Dept Chem, Bam, Iran
[2] Payame Noor Univ, Dept Chem, Kerman, Iran
[3] Shahid Bahonar Univ, Dept Chem, Kerman, Iran
[4] Int Ctr Sci High Technol Environm Sci, Kerman, Iran
关键词
Cloud point extraction; Preconcentration; Lead determination; WATER SAMPLES; SPECTROPHOTOMETRIC DETERMINATION; FLUORESCENCE SPECTROMETRY; CADMIUM; PRECONCENTRATION; EMISSION; CU(II); UPDATE; COPPER;
D O I
10.1016/j.arabjc.2011.07.003
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A sensitive cloud point extraction procedure(CPE) for the preconcentration of trace lead prior to its determination by flame atomic absorption spectrometry (FAAS) has been developed. The CPE method is based on the complex of Pb(II) ion with 1-(2-pyridylazo)-2-naphthol (PAN), and then entrapped in the non-ionic surfactant Triton X-114. The main factors affecting CPE efficiency, such as pH of sample solution, concentration of PAN and Triton X-114, equilibration temperature and time, were investigated in detail. A preconcentration factor of 30 was obtained for the preconcentration of Pb(II) ion with 15.0 mL solution. Under the optimal conditions, the calibration curve was linear in the range of 7.5 ng mL(-1) -3.5 mu g mL(-1) of lead with R-2 = 0.9998 (n = 10). Detection limit based on three times the standard deviation of the blank (3S(b)) was 5.27 ng mL(-1). Eight replicate determinations of 1.0 mu g mL(-1) lead gave a mean absorbance of 0.275 with a relative standard deviation of 1.6%. The high efficiency of cloud point extraction to carry out the determination of analytes in complex matrices was demonstrated.The proposed method has been applied for determination of trace amounts of lead in biological and water samples with satisfactory results. (C) 2011 Production and hosting by Elsevier B.V. on behalf of King Saud University.
引用
收藏
页码:S610 / S615
页数:6
相关论文
共 19 条
[1]   Evaluation of polychlorotrifluoroethylene as sorbent material for on-line solid phase extraction systems: Determination of copper and lead by flame atomic absorption spectrometry in water samples [J].
Anthemidis, Aristidis N. ;
Ioannou, Kallirroy-Ioanna G. .
ANALYTICA CHIMICA ACTA, 2006, 575 (01) :126-132
[2]   Ligandless cloud point extraction of Cr(III), Pb(II), Cu(II), Ni(II), Bi(III), and Cd(II) ions in environmental samples with Tween 80 and flame atomic absorption spectrometric determination [J].
Candir, Secil ;
Narin, Ibrahim ;
Soylak, Mustafa .
TALANTA, 2008, 77 (01) :289-293
[3]   Determination of lead in water samples by graphite furnace atomic absorption spectrometry after cloud point extraction [J].
Chen, JR ;
Xiao, SM ;
Wu, XH ;
Fang, KM ;
Liu, WH .
TALANTA, 2005, 67 (05) :992-996
[4]   Determination of cadmium, copper, lead and zinc in water samples by flame atomic absorption spectrometry after cloud point extraction [J].
Chen, JR ;
Teo, KC .
ANALYTICA CHIMICA ACTA, 2001, 450 (1-2) :215-222
[5]   Automatic flow procedure based on multicommutation exploiting liquid-liquid extraction for spectrophotometric lead determination in plant material [J].
Comitre, ALD ;
Reis, BF .
TALANTA, 2005, 65 (04) :846-852
[6]   A sensitive spectrophotometric method for lead determination by flow injection analysis with on-line preconcentration [J].
Di Nezio, MS ;
Palomeque, ME ;
Band, BSF .
TALANTA, 2004, 63 (02) :405-409
[7]   Atomic spectrometry update. Advances in atomic emission, absorption and fluorescence spectrometry, and related techniques [J].
Evans, EH ;
Day, JA ;
Palmer, CD ;
Price, WJ ;
Smith, CMM ;
Tyson, JF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2005, 20 (06) :562-590
[8]   Atomic spectrometry update. Advances in atomic emission, absorption and fluorescence spectrometry and related techniques [J].
Evans, EH ;
Day, JA ;
Fisher, A ;
Price, WJ ;
Smith, CMM ;
Tyson, JF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (06) :775-812
[9]   Applicability of cloud point extraction coupled with microwave-assisted back-extraction to the determination of organophosphorous pesticides in human urine by gas chromatography with flame photometry detection [J].
Jia, GuiFang ;
Lv, ChunGuang ;
Zhu, WenTao ;
Qiu, Jing ;
Wang, XinQuan ;
Zhou, ZhiQiang .
JOURNAL OF HAZARDOUS MATERIALS, 2008, 159 (2-3) :300-305
[10]   Determination of lead in human saliva by combined cloud point extraction-capillary zone electrophoresis with indirect UV detection [J].
Luconi, MO ;
Olsina, RA ;
Fernández, LP ;
Silva, MF .
JOURNAL OF HAZARDOUS MATERIALS, 2006, 128 (2-3) :240-246