Reliability of Apnea-Hypopnea Index Measured by a Home Bi-Level Pressure Support Ventilator Versus a Polysomnographic Assessment

被引:39
作者
Georges, Marjolaine [1 ]
Adler, Dan [1 ]
Contal, Olivier [1 ]
Espa, Fabrice [2 ]
Perrig, Stephen [2 ]
Pepin, Jean-Louis [2 ,3 ,4 ]
Janssens, Jean-Paul [1 ]
机构
[1] Univ Hosp Geneva, Dept Med Specialties, Div Pulm Dis, CH-1211 Geneva 14, Switzerland
[2] Univ Hosp Geneva, Sleep Lab, Dept Psychiat, CH-1211 Geneva 14, Switzerland
[3] Univ Hosp, Sleep Lab, Grenoble, France
[4] Univ Grenoble Alpes, Grenoble, France
关键词
apnea-hypopnea index; noninvasive ventilation; monitoring; obesity-hypoventilation syndrome; home ventilator software; OBESITY HYPOVENTILATION SYNDROME; NONINVASIVE VENTILATION; RESPIRATORY EVENTS; SLEEP; PATTERNS; DEVICE; CPAP;
D O I
10.4187/respcare.03633
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Ventilators designed for home care provide clinicians with built-in software that records items such as compliance, leaks, average tidal volume, total ventilation, and indices of residual apnea and hypopnea. Recent studies have showed, however, an important variability between devices regarding reliability of data provided. In this study, we aimed to compare apnea-hypopnea indices (AHI) provided by home ventilators (AHI(NIV)) versus data scored manually during polysomnography (AHI(PSG)) in subjects on noninvasive ventilation (NIV) for obesity-hypoventilation syndrome. METHODS: Stable subjects with obesity-hypoventilation syndrome on NIV, all using the same device, underwent 3 consecutive polysomnographic sleep studies with different backup breathing frequencies (spontaneous mode, low and high backup breathing frequencies). During each recording, AHI(NIV) was compared with AHI(PSG). RESULTS: Ten subjects (30 polysomnogram tracings) were analyzed. For each backup breathing frequency (spontaneous mode, low and high backup breathing frequencies), AHI values were 62 +/- 7/h, 26 +/- 7/h, and 17 +/- 5/h (mean +/- SD), respectively. Correlation between AHI(NIV) and AHI(PSG) was highly significant (r(2) = 0.89, P < .001). As determined by Bland-Altman analysis, mean bias was 6.5 events/h, and limits of agreement were +26.0 and -12.9 events/h. Bias increased significantly with higher AHI values. Using a threshold AHI value of 10/h to define appropriate control of respiratory events, the ventilator software had a sensitivity of 90.9%, a specificity and positive predictive value of 100%, and a negative predictive value of 71%. CONCLUSIONS: In stable subjects with obesity-hypoventilation syndrome, the home ventilator software tested was appropriate for determining if control of respiratory events was satisfactory on NIV or if further testing or adjustment of ventilator settings was required.
引用
收藏
页码:1051 / 1056
页数:6
相关论文
共 27 条
[1]   Pulse wave amplitude reduction: A surrogate marker of micro-arousals associated with respiratory events occurring under non-invasive ventilation? [J].
Adler, Dan ;
Bridevaux, Pierre Olivier ;
Contal, Olivier ;
Georges, Marjolaine ;
Dupuis-Lozeron, Elise ;
Claudel, Elisabeth ;
Pepin, Jean Louis ;
Janssens, Jean Paul .
RESPIRATORY MEDICINE, 2013, 107 (12) :2053-2060
[2]   Performance characteristics of 10 home mechanical ventilators in pressure-support mode - A comparative bench study [J].
Battisti, A ;
Tassaux, D ;
Janssens, JP ;
Michotte, JB ;
Jaber, S ;
Jolliet, P .
CHEST, 2005, 127 (05) :1784-1792
[3]  
Berry RB, 2020, AASM MANUAL SCORING
[4]   Respiratory Event Detection by a Positive Airway Pressure Device [J].
Berry, Richard B. ;
Kushida, Clete A. ;
Kryger, Meir H. ;
Soto-Calderon, Haideliza ;
Staley, Bethany ;
Kuna, Samuel T. .
SLEEP, 2012, 35 (03) :361-367
[5]  
Berry RB, 2010, J CLIN SLEEP MED, V6, P491
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]   Noninvasive Ventilation in Mild Obesity Hypoventilation Syndrome A Randomized Controlled Trial [J].
Borel, Jean-Christian ;
Tamisier, Renaud ;
Gonzalez-Bermejo, Jesus ;
Baguet, Jean-Philippe ;
Monneret, Denis ;
Arnol, Nathalie ;
Roux-Lombard, Pascale ;
Wuyam, Bernard ;
Levy, Patrick ;
Pepin, Jean-Louis .
CHEST, 2012, 141 (03) :692-702
[8]   The accuracy of autotitrating CPAP-determined residual apnea-hypopnea index [J].
Cilli, Aykut ;
Uzun, Rusen ;
Bilge, Ugur .
SLEEP AND BREATHING, 2013, 17 (01) :189-193
[9]   Impact of Different Backup Respiratory Rates on the Efficacy of Noninvasive Positive Pressure Ventilation in Obesity Hypoventilation Syndrome A Randomized Trial [J].
Contal, Olivier ;
Adler, Dan ;
Borel, Jean-Christian ;
Espa, Fabrice ;
Perrig, Stephen ;
Rodenstein, Daniel ;
Pepin, Jean-Louis ;
Janssens, Jean-Paul .
CHEST, 2013, 143 (01) :37-46
[10]   Monitoring of Noninvasive Ventilation by Built-in Software of Home Bilevel Ventilators A Bench Study [J].
Contal, Olivier ;
Vignaux, Laurence ;
Combescure, Christophe ;
Pepin, Jean-Louis ;
Jolliet, Philippe ;
Janssens, Jean-Paul .
CHEST, 2012, 141 (02) :469-476