A comprehensive and efficient daily quality assurance for PBS proton therapy

被引:34
作者
Actis, O. [1 ]
Meer, D. [1 ]
Koenig, S. [1 ]
Weber, D. C. [1 ,2 ]
Mayor, A. [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, CH-5232 Villigen, Switzerland
[2] Univ Zurich Hosp, Dept Radiat Oncol, Zurich, Switzerland
关键词
proton therapy; quality assurance; pencil beam scanning; dosimetric equipment; verification of beam parameters; efficient workflow; ANDERSON CANCER CENTER; IONIZATION-CHAMBER; BEAM; SYSTEM; GANTRY; PSI; DOSIMETRY; HOUSTON; DESIGN; HADRON;
D O I
10.1088/1361-6560/aa5131
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There are several general recommendations for quality assurance (QA) measures, which have to be performed at proton therapy centres. However, almost each centre uses a different therapy system. In particular, there is no standard procedure for centres employing pencil beam scanning and each centre applies a specific QA program. Gantry 2 is an operating therapy system which was developed at PSI and relies on the most advanced technological innovations. We developed a comprehensive daily QA program in order to verify the main beam characteristics to assure the functionality of the therapy delivery system and the patient safety system. The daily QA program entails new hardware and software solutions for a highly efficient clinical operation. In this paper, we describe a dosimetric phantom used for verifying the most critical beam parameters and the software architecture developed for a fully automated QA procedure. The connection between our QA software and the database allows us to store the data collected on a daily basis and use it for trend analysis over longer periods of time. All the data presented here have been collected during a time span of over two years, since the beginning of the Gantry 2 clinical operation in 2013. Our procedure operates in a stable way and delivers the expected beam quality. The daily QA program takes only 20 min. At the same time, the comprehensive approach allows us to avoid most of the weekly and monthly QA checks and increases the clinical beam availability.
引用
收藏
页码:1661 / 1675
页数:15
相关论文
共 35 条
[1]   Precise on-line position measurement for particle therapy [J].
Actis, O. ;
Meer, D. ;
Koenig, S. .
JOURNAL OF INSTRUMENTATION, 2014, 9
[2]   Dosimetric characterization of a large area pixel-segmented ionization chamber [J].
Amerio, S ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Donetti, M ;
Fidanzio, A ;
Garelli, E ;
Giordanengo, S ;
Madon, E ;
Marchetto, F ;
Nastasi, U ;
Peroni, C ;
Piermattei, A ;
Freire, CJS ;
Sardo, A ;
Trevisiol, E .
MEDICAL PHYSICS, 2004, 31 (02) :414-420
[3]  
[Anonymous], 2000, TECHNICAL REPORT SER
[4]   An overview of the comprehensive proton therapy machine quality assurance procedures implemented at The University of Texas M. D. Anderson Cancer Center Proton Therapy Center-Houston [J].
Arjomandy, Bijan ;
Sahoo, Narayan ;
Zhu, X. Ronald ;
Zullo, John R. ;
Wu, Richard Y. ;
Zhu, Mingping ;
Ding, Xiaoning ;
Martin, Craig ;
Ciangaru, George ;
Gillin, Michael T. .
MEDICAL PHYSICS, 2009, 36 (06) :2269-2282
[5]   Segmented ionization chambers for beam monitoring in hadrontherapy [J].
Braccini, Saverio ;
Cirio, Roberto ;
Donetti, Marco ;
Marchetto, Flavio ;
Pitta, Giuseppe ;
Lavagno, Marco ;
La Rosa, Vanessa .
MODERN PHYSICS LETTERS A, 2015, 30 (17)
[6]   Strip ionization chambers as 3-D detector for hadron therapy [J].
Brusasco, C ;
Cattai, A ;
Cirio, R ;
Dellacasa, G ;
Donetti, M ;
Isoardi, P ;
Marchetto, F ;
Peroni, C ;
Rolando, V ;
Ruspa, M ;
Solano, A ;
Zambernardi, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 389 (03) :499-512
[7]   A dosimetry system for fast measurement of 3D depth-dose profiles in charged-particle tumor therapy with scanning techniques [J].
Brusasco, C ;
Voss, B ;
Schardt, D ;
Krämer, M ;
Kraft, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2000, 168 (04) :578-592
[8]   Two-dimensional and quasi-three-dimensional dosimetry of hadron and photon beams with the Magic Cube and the Pixel Ionization Chamber [J].
Cirio, R ;
Garelli, E ;
Schulte, R ;
Amerio, S ;
Boriano, A ;
Bourhaleb, F ;
Coutrakon, G ;
Donetti, M ;
Giordanengo, S ;
Koss, P ;
Madon, E ;
Marchetto, F ;
Nastasi, U ;
Peroni, C ;
Santuari, D ;
Sardo, A ;
Scielzo, G ;
Stasi, M ;
Trevisiol, E .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (16) :3713-3724
[9]  
De. Tec. Tor, 2016, DEVICES TECHNOLOGIES
[10]   A novel daily QA system for proton therapy [J].
Ding, Xiaoning ;
Zheng, Yuanshui ;
Zeidan, Omar ;
Mascia, Anthony ;
Hsi, Wen ;
Kang, Yixiu ;
Ramirez, Eric ;
Schreuder, Niek ;
Harris, Ben .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :115-126