In Vitro, Time-Resolved PIV Comparison of the Effect of Stent Design on Wall Shear Stress

被引:52
作者
Charonko, John [1 ]
Karri, Satyaprakash [1 ]
Schmieg, Jaime [1 ]
Prabhu, Santosh [2 ]
Vlachos, Pavlos [1 ]
机构
[1] Virginia Tech, Blacksburg, VA 24061 USA
[2] Abbott Vasc, Santa Clara, CA 95054 USA
基金
美国国家科学基金会;
关键词
Coronary arteries; Blood flow; Phase offset; Oscillatory shear index; Drug-eluting stent; Bare metal stent; Endothelial cells; CORONARY-ARTERY-DISEASE; NEWTONIAN BLOOD-FLOW; FLUID-DYNAMICS; IMPLANTATION; MODEL; RESTENOSIS; PLACEMENT; IMMEDIATE; HUMANS;
D O I
10.1007/s10439-009-9697-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effect of stent design on wall shear stress (WSS) and oscillatory shear index (OSI) was studied in vitro using time-resolved digital particle image velocimetry (DPIV). Four drug-eluting stents [XIENCE V-A (R) (Abbott Vascular), TAXUS(A (R)) Libert,(A (R)) (Boston Scientific), Endeavor(A (R)) (Medtronic), and Cypher(A (R)) (J&J Cordis)] and a bare-metal stent [VISIONA (R) (Abbott Vascular)] were implanted into compliant vessel models, and the flow was measured in physiologically accurate coronary conditions featuring reversal and realistic offsets between pressure and flowrate. DPIV measurements were made at three locations under two different flow rates (resting: Re = 160, f = 70 bpm and exercise: Re = 300, f = 120 bpm). It was observed that design substantially affected the WSS experienced at the vessel walls. Averaged values between struts ranged from 2.05 dynes/cm(2) (Cypher(A (R))) to 8.52 dynes/cm(2) (XIENCE V-A (R)) in resting conditions, and from 3.72 dynes/cm(2) (Cypher(A (R))) to 14.66 dynes/cm(2) (VISIONA (R)) for the exercise state. Within the stent, the WSS dropped and the OSI increased immediately distal to each strut. In addition, an inverse correlation between average WSS and OSI existed. Comparisons with recently published results from animal studies show strong correlation between the measured WSS and observed endothelial cell coverage. These results suggest the importance of stent design on the WSS experienced by endothelial cells in coronary arteries.
引用
收藏
页码:1310 / 1321
页数:12
相关论文
共 49 条
[21]   Restenosis after coronary placement of various stent types [J].
Kastrati, A ;
Mehilli, J ;
Dirschinger, J ;
Pache, J ;
Ulm, K ;
Schühlen, H ;
Seyfarth, M ;
Schmitt, C ;
Blasini, R ;
Neumann, FJ ;
Schömig, A .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 87 (01) :34-39
[22]  
Kastrati A, 2001, CIRCULATION, V103, P2816
[23]  
Kleinstreuer C, 2001, CRIT REV BIOMED ENG, V29, P1, DOI 10.1615/CritRevBiomedEng.v29.i1.10
[24]   Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with time-dependent 3D computational fluid dynamics models [J].
LaDisa, JF ;
Olson, LE ;
Guler, I ;
Hettrick, DA ;
Kersten, JR ;
Warltier, DC ;
Pagel, PS .
JOURNAL OF APPLIED PHYSIOLOGY, 2005, 98 (03) :947-957
[25]   Stent design properties and deployment ratio influence indexes of wall shear stress: a three-dimensional computational fluid dynamics investigation within a normal artery [J].
LaDisa, JF ;
Olson, LE ;
Guler, I ;
Hettrick, DA ;
Audi, SH ;
Kersten, JR ;
Warltier, DC ;
Pagel, PS .
JOURNAL OF APPLIED PHYSIOLOGY, 2004, 97 (01) :424-430
[26]   Three-dimensional computational fluid dynamics modeling of alterations in coronary wall shear stress produced by stent implantation [J].
LaDisa, JF ;
Guler, I ;
Olson, LE ;
Hettrick, DA ;
Kersten, JR ;
Warltier, DC ;
Pagel, PS .
ANNALS OF BIOMEDICAL ENGINEERING, 2003, 31 (08) :972-980
[27]  
LEE SW, 2008, P 2008 SUMM BIOENG C
[28]   Materials, fluid dynamics, and solid mechanics aspects of coronary artery stents: A state-of-the-art review [J].
Lewis, Gladius .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2008, 86B (02) :569-590
[29]   Hemodynamic shear stress and its role in atherosclerosis [J].
Malek, AM ;
Alper, SL ;
Izumo, S .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 282 (21) :2035-2042
[30]   Drug-eluting stent pathology - should we still be cautious? [J].
Nakazawa, Gaku ;
Finn, Aloke V. ;
Virmani, Renu .
NATURE CLINICAL PRACTICE CARDIOVASCULAR MEDICINE, 2008, 5 (01) :1-1