Automatic inverse treatment planning of Gamma Knife radiosurgery via deep reinforcement learning

被引:4
|
作者
Liu, Yingzi [1 ]
Shen, Chenyang [2 ]
Wang, Tonghe [1 ]
Zhang, Jiahan [1 ]
Yang, Xiaofeng [1 ]
Liu, Tian [1 ]
Kahn, Shannon [1 ]
Shu, Hui-Kuo [1 ]
Tian, Zhen [1 ]
机构
[1] Emory Univ, Dept Radiat Oncol, Atlanta, GA USA
[2] Univ Texas Southwestern Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX 75390 USA
关键词
automatic inverse treatment planning; automatic priority tuning; deep reinforcement learning; Gamma Knife radiosurgery; MULTIOBJECTIVE OPTIMIZATION; EVOLUTIONARY ALGORITHM; WEIGHTS; SYSTEM;
D O I
10.1002/mp.15576
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Several inverse planning algorithms have been developed for Gamma Knife (GK) radiosurgery to determine a large number of plan parameters by solving an optimization problem, which typically consists of multiple objectives. The priorities among these objectives need to be repetitively adjusted to achieve a clinically good plan for each patient. This study aimed to achieve automatic and intelligent priority tuning by developing a deep reinforcement learning (DRL)-based method to model the tuning behaviors of human planners. Methods We built a priority-tuning policy network using deep convolutional neural networks. Its input was a vector composed of multiple plan metrics that were used in our institution for GK plan evaluation. The network can determine which tuning action to take based on the observed quality of the intermediate plan. We trained the network using an end-to-end DRL framework to approximate the optimal action-value function. A scoring function was designed to measure the plan quality to calculate the received reward of a tuning action. Results Vestibular schwannoma was chosen as the test bed in this study. The number of training, validation and testing cases were 5, 5, and 16, respectively. For these three datasets, the average scores of the initial plans obtained with the same initial priority set were 3.63 +/- 1.34, 3.83 +/- 0.86 and 4.20 +/- 0.78, respectively, while they were improved to 5.28 +/- 0.23, 4.97 +/- 0.44 and 5.22 +/- 0.26 through manual priority tuning by human expert planners. Our network achieved competitive results with 5.42 +/- 0.11, 5.10 +/- 0. 42, 5.28 +/- 0.20, respectively. Conclusions Our network can generate GK plans of comparable or slightly higher quality than the plans generated by human planners via manual priority tuning for vestibular schwannoma cases. The network can potentially be incorporated into the clinical workflow as planning assistance to improve GK planning efficiency and help to reduce plan quality variation caused by interplanner variability. We also hope that our method can reduce the workload of GK planners and allow them to spend more time on more challenging cases.
引用
收藏
页码:2877 / 2889
页数:13
相关论文
共 50 条
  • [31] Gamma Knife Stereotactic Radiosurgery for the Treatment of Primary and Metastatic Ocular Malignancies
    Arnett, Andrea L. H.
    Reynolds, Margaret M.
    Pulido, Jose S.
    Parney, Ian F.
    Laack, Nadia N.
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2017, 95 (06) : 363 - 368
  • [32] Treatment of sellar metastases with gamma knife radiosurgery in patients with advanced cancer
    Benjamin, Carolina
    Ashayeri, Kimberly
    Golfinos, John G.
    Placantonakis, Dimitris G.
    Silverman, Joshua
    Kondziolka, Douglas
    PITUITARY, 2020, 23 (06) : 665 - 671
  • [33] Patterns of failure after treatment of atypical meningioma with gamma knife radiosurgery
    Attia, Albert
    Chan, Michael D.
    Mott, Ryan T.
    Russell, Greg B.
    Seif, David
    Bourland, J. Daniel
    Deguzman, Allan F.
    Ellis, Thomas L.
    McMullen, Kevin P.
    Munley, Michael T.
    Tatter, Stephen B.
    Shaw, Edward G.
    JOURNAL OF NEURO-ONCOLOGY, 2012, 108 (01) : 179 - 185
  • [34] Patterns of failure after treatment of atypical meningioma with gamma knife radiosurgery
    Albert Attia
    Michael D. Chan
    Ryan T. Mott
    Greg B. Russell
    David Seif
    J. Daniel Bourland
    Allan F. Deguzman
    Thomas L. Ellis
    Kevin P. McMullen
    Michael T. Munley
    Stephen B. Tatter
    Edward G. Shaw
    Journal of Neuro-Oncology, 2012, 108 : 179 - 185
  • [35] Gamma knife radiosurgery for treatment of trigeminal neuralgia in multiple sclerosis patients
    Huang, E
    Teh, BS
    Zeck, O
    Woo, SY
    Lu, HH
    Chiu, JK
    Butler, EB
    Gormley, WB
    Carpenter, LS
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2002, 79 (01) : 44 - 50
  • [36] Treatment of sellar metastases with gamma knife radiosurgery in patients with advanced cancer
    Carolina Benjamin
    Kimberly Ashayeri
    John G. Golfinos
    Dimitris G. Placantonakis
    Joshua Silverman
    Douglas Kondziolka
    Pituitary, 2020, 23 : 665 - 671
  • [37] Gamma Knife Radiosurgery in the Treatment of Tumor-Related Facial Pain
    Squire, Sarah E.
    Chan, Michael D.
    Furr, R. Michael
    Lowell, Dorothy A.
    Tatter, Stephen B.
    Ellis, Thomas L.
    Bourland, J. Daniel
    deGuzman, Allan F.
    Munley, Michael T.
    Ekstrand, Kenneth E.
    Shaw, Edward G.
    McMullen, Kevin P.
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2012, 90 (03) : 145 - 150
  • [38] An integrated solution of deep reinforcement learning for automatic IMRT treatment planning in non-small-cell lung cancer
    Wang, Hanlin
    Bai, Xue
    Wang, Yajuan
    Lu, Yanfei
    Wang, Binbing
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [39] Path Planning for UAV Ground Target Tracking via Deep Reinforcement Learning
    Li, Bohao
    Wu, Yunjie
    IEEE ACCESS, 2020, 8 (29064-29074) : 29064 - 29074
  • [40] Deep reinforcement learning trajectory planning for vibration suppression via jerk control
    Park, Sung Gwan
    Rhim, Sungsoo
    2023 20TH INTERNATIONAL CONFERENCE ON UBIQUITOUS ROBOTS, UR, 2023, : 818 - 824