A study of the accuracy of CyberKnife spinal radiosurgery using skeletal structure tracking

被引:103
作者
Ho, Anthony K.
Fu, Dongshan
Cotrutz, Cristian
Hancock, Steven L.
Chang, Steven D.
Gibbs, Iris C.
Maurer, Calvin R., Jr.
Adler, John R., Jr.
机构
[1] Stanford Univ, Med Ctr, Dept Radiat Oncol, Stanford, CA 94305 USA
[2] Stanford Univ, Med Ctr, Dept Neurosurg, Stanford, CA 94305 USA
[3] Accuray Inc, Sunnyvale, CA USA
关键词
CyberKnife; image-guided robotic radiosurgery; spinal tumors; spine;
D O I
10.1227/01.NEU.0000249248.55923.EC
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: New technology has enabled the increasing use of radiosurgery to ablate spinal lesions. The first generation of the CyberKnife (Accuray, Inc., Sunnyvale, CA) image-guided radiosurgery system required implanted radiopaque markers (fiducials) to localize spinal targets. A recently developed and now commercially available spine tracking technology called Xsight (Accuray, Inc.) tracks skeletal structures and eliminates the need for implanted fiducials. The Xsight system localizes spinal targets by direct reference to the adjacent vertebral elements. This study sought to measure the accuracy of Xsight spine tracking and provide a qualitative assessment of overall system performance. METHODS: Total system error, which is defined as the distance between the centroids of the planned and delivered dose distributions and represents all possible treatment planning and delivery errors, was measured using a realistic, anthropomorphic head-and-neck phantom. The Xsight tracking system error component of total system error was also computed by retrospectively analyzing image data obtained from eleven patients with a total of 44 implanted fiducials who underwent CyberKnife spinal radiosurgery. RESULTS: The total system error of the Xsight targeting technology was measured to be 0.61 mm. The tracking system error component was found to be 0.49 mm. CONCLUSION: The Xsight spine tracking system is practically important because it is accurate and eliminates the use of implanted fiducials. Experience has shown this technology to be robust under a wide range of clinical circumstances.
引用
收藏
页码:147 / 156
页数:10
相关论文
共 38 条
[31]   Validation of a two- to three-dimensional registration algorithm for aligning preoperative CT images and intraoperative fluoroscopy images [J].
Penney, GP ;
Batchelor, PG ;
Hill, DLG ;
Hawkes, DJ ;
Weese, J .
MEDICAL PHYSICS, 2001, 28 (06) :1024-1032
[32]   Preliminary visual field preservation after staged CyberKnife radiosurgery for perioptic lesions [J].
Pham, CJ ;
Chang, SD ;
Gibbs, IC ;
Jones, P ;
Heilbrun, MP ;
Adler, JR .
NEUROSURGERY, 2004, 54 (04) :799-810
[33]   Fast generation of digitally reconstructed radiographs using attenuation fields with application to 2D-3D image registration [J].
Russakoff, DB ;
Rohlfing, T ;
Mori, K ;
Rueckert, D ;
Ho, A ;
Adler, JR ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (11) :1441-1454
[34]   Image-guided hypo-fractionated stereotactic radiosurgery to spinal lesions [J].
Ryu, SI ;
Chang, SD ;
Kim, DH ;
Murphy, MJ ;
Le, QT ;
Martin, DP ;
Adler, JR .
NEUROSURGERY, 2001, 49 (04) :838-846
[35]   Multisession cyberknife radiosurgery for intramedullary spinal cord arteriovenous malformations [J].
Sinclair, John ;
Chang, Steven D. ;
Gibbs, Iris C. ;
Adler, John R., Jr. .
NEUROSURGERY, 2006, 58 (06) :1081-1088
[36]  
Weese J, 1997, IEEE Trans Inf Technol Biomed, V1, P284, DOI 10.1109/4233.681173
[37]   Accuray CyberKnife® image-guided radiosurgical system [J].
Welch, WC ;
Gerszten, PC .
EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (02) :141-147
[38]   Comparison and evaluation of retrospective intermodality brain image registration techniques [J].
West, J ;
Fitzpatrick, JM ;
Wang, MY ;
Dawant, BM ;
Maurer, CR ;
Kessler, RM ;
Maciunas, RJ ;
Barillot, C ;
Lemoine, D ;
Collignon, A ;
Maes, F ;
Suetens, P ;
Vandermeulen, D ;
vandenElsen, PA ;
Napel, S ;
Sumanaweera, TS ;
Harkness, B ;
Hemler, PF ;
Hill, DLG ;
Hawkes, DJ ;
Studholme, C ;
Maintz, JBA ;
Viergever, MA ;
Malandain, G ;
Pennec, X ;
Noz, ME ;
Maguire, GQ ;
Pollack, M ;
Pelizzari, CA ;
Robb, RA ;
Hanson, D ;
Woods, RP .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (04) :554-566