Positron emission mammographic instrument: Initial results

被引:49
作者
Murthy, K
Aznar, M
Bergman, AM
Thompson, CJ
Robar, JL
Lisbona, R
Loutfi, A
Gagnon, JH
机构
[1] McGill Univ, Royal Victoria Hosp, Montreal Neurol Inst, Med Phys Unit, Montreal, PQ H3A 2B4, Canada
[2] McGill Univ, Royal Victoria Hosp, Dept Nucl Med, Montreal, PQ H3A 2B4, Canada
[3] McGill Univ, Royal Victoria Hosp, Dept Surg Oncol, Montreal, PQ H3A 2B4, Canada
[4] McGill Univ, Royal Victoria Hosp, Dept Radiol, Montreal, PQ H3A 2B4, Canada
关键词
breast radiography; technology;
D O I
10.1148/radiology.215.1.r00ap03280
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Performance characteristics of a positron emission mammographic (PEM) instrument were studied. This dedicated metabolic breast imaging system has spatial resolution of 2.8-mm full width at half maximum (FWHM), coincidence resolving time of 12-nsec FWHM, and absolute efficiency of 3%. Hot spots with diameter of 16 mm in a phantom with signal-to-background activity ratio of 6:1 were distinguishable with a scanning time of 5 minutes.
引用
收藏
页码:280 / 285
页数:6
相关论文
共 19 条
[1]  
ADAM LE, 1996, IEEE NUCL SCI S MED, V2, P1270
[2]   NEW METHODS FOR IMAGING THE BREAST - TECHNIQUES, FINDINGS, AND POTENTIAL [J].
ADLER, DD ;
WAHL, RL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (01) :19-30
[3]   Technique to obtain positron emission mammography images in registration with x-ray mammograms [J].
Bergman, AM ;
Thompson, CJ ;
Murthy, K ;
Robar, JL ;
Clancy, RL ;
English, MJ ;
Loutfi, A ;
Lisbona, R ;
Gagnon, J .
MEDICAL PHYSICS, 1998, 25 (11) :2119-2129
[4]   Simple technique to increase the linearity and field-of-view in position sensitive photomultiplier tubes [J].
Clancy, RL ;
Thompson, CJ ;
Robar, JL ;
Bergman, AM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :494-498
[5]   PERFORMANCE EVALUATION OF THE PC-2048 - A NEW 15-SLICE ENCODED-CRYSTAL PET SCANNER FOR NEUROLOGICAL STUDIES [J].
EVANS, AC ;
THOMPSON, CJ ;
MARRETT, S ;
MEYER, E ;
MAZZA, M .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (01) :90-98
[6]  
HENSON DE, 1990, CANCER, V65, P2155, DOI 10.1002/1097-0142(19900501)65:9+<2155::AID-CNCR2820651412>3.0.CO
[7]  
2-O
[8]  
HOFFMAN EJ, 1986, IEEE T NUCL SCI, V33, P420
[9]   POSITRON EMISSION TOMOGRAPHY STUDIES IN PATIENTS WITH LOCALLY ADVANCED AND/OR METASTATIC BREAST-CANCER - A METHOD FOR EARLY THERAPY EVALUATION [J].
JANSSON, T ;
WESTLIN, JE ;
AHLSTROM, H ;
LILJA, A ;
LANGSTROM, B ;
BERGH, J .
JOURNAL OF CLINICAL ONCOLOGY, 1995, 13 (06) :1470-1477
[10]  
MURTHY K, 1998, IN PRESS IEEE T NUCL