Pipeline Embolization Device With Adjunctive Coils for the Treatment of Unruptured Large or Giant Vertebrobasilar Aneurysms: A Single-Center Experience

被引:8
作者
Zhou, Yangyang [1 ]
Wu, Xinzhi [1 ]
Tian, Zhongbin [1 ]
Yang, Xinjian [1 ]
Mu, Shiqing [1 ]
机构
[1] Capital Med Univ, Beijing Neurosurg Inst, Beijing Tiantan Hosp, Dept Intervent Neuroradiol, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
pipeline embolization device; coils; large aneurysms; giant aneurysms; vertebrobasilar artery; INTRACRANIAL ANEURYSMS; ENDOVASCULAR TREATMENT; POSTERIOR CIRCULATION; FLOW DIVERTORS; FUSIFORM ANEURYSMS; DIVERSION; STENTS;
D O I
10.3389/fneur.2020.522583
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To evaluate effectiveness and safety of Pipeline embolization device (PED) for large or giant verterbrobasilar aneurysms (LGVBAs), and to compare the therapeutic effects of PED with and without adjunctive coils. Methods: We retrospectively analyzed 21 cases of unruptured LGVBAs who were treated in our hospital with PED. These cases were divided into "PED group" and "PED with adjunctive coils group." We compared the aneurysm characteristics and treatment outcomes between the two groups. Results: The overall neurological complication rate was 28.6% (6/21) and the mortality rate was 4.8% (1/21). There were 12 patients in the PED group and nine in the PED with adjunctive coils group. There were no significant differences in age, smoking, hypertension, aneurysm size, aneurysm location, or operation time between the two groups. The complete aneurysm embolization rate and favorable outcome rate (modified Rankin Scale = 0,1) of the PED with adjunctive coils group was 78% (7/9) and 100% (9/9), respectively, which were both better compared with the PED group with 63.6% (7/11) and 83% (10/12), respectively. However, these differences were not statistically significant. Conclusion: The effectiveness and safety of PED for LGVBAs is acceptable. Treatment results did not differ between the PED and PED with adjunctive coils groups; therefore, whether coils should be used may depend the operator. Our results suggest that correct use of the coils does not increase complications. We suggest that PED with adjunctive coils should be used for some selected LGVBAs.
引用
收藏
页数:11
相关论文
共 30 条
[1]   Flow-Diverter Devices for Intracranial Aneurysms: Systematic Review and Meta-analysis [J].
Arrese, Ignacio ;
Sarabia, Rosario ;
Pintado, Rebeca ;
Delgado-Rodriguez, Miguel .
NEUROSURGERY, 2013, 73 (02) :193-199
[2]   Endovascular management of fusiform aneurysms in the posterior circulation: the era of flow diversion [J].
Awad, Ahmed J. ;
Mascitelli, Justin R. ;
Haroun, Reham R. ;
De Leacy, Reade A. ;
Fifi, Johanna T. ;
Mocco, J. .
NEUROSURGICAL FOCUS, 2017, 42 (06)
[3]   Smoking is not associated with recurrence and retreatment of intracranial aneurysms after endovascular coiling [J].
Brinjikji, Waleed ;
Lingineni, Ravi K. ;
Gu, Chris N. ;
Lanzino, Giuseppe ;
Cloft, Harry J. ;
Ilish, Lauren ;
Koeller, Kristen ;
Kallmes, David F. .
JOURNAL OF NEUROSURGERY, 2015, 122 (01) :95-100
[4]   Endovascular Treatment of Intracranial Aneurysms With Flow Diverters A Meta-Analysis [J].
Brinjikji, Waleed ;
Murad, Mohammad H. ;
Lanzino, Giuseppe ;
Cloft, Harry J. ;
Kallmes, David F. .
STROKE, 2013, 44 (02) :442-447
[5]   Spontaneous Delayed Migration/Shortening of the Pipeline Embolization Device: Report of 5 Cases [J].
Chalouhi, N. ;
Tjoumakaris, S. I. ;
Gonzalez, L. F. ;
Hasan, D. ;
Pema, P. J. ;
Gould, G. ;
Rosenwasser, R. H. ;
Jabbour, P. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (12) :2326-2330
[6]   Patency of the posterior communicating artery following treatment with the Pipeline Embolization Device [J].
Daou, Badih ;
Valle-Giler, Edison P. ;
Chalouhi, Nohra ;
Starke, Robert M. ;
Tjoumakaris, Stavropoula ;
Hasan, David ;
Rosenwasser, Robert H. ;
Hebert, Ryan ;
Jabbour, Pascal .
JOURNAL OF NEUROSURGERY, 2017, 126 (02) :564-569
[7]   Giant fusiform intracranial aneurysms: review of 120 patients treated surgically from 1965 to 1992 [J].
Drake, CG ;
Peerless, SJ .
JOURNAL OF NEUROSURGERY, 1997, 87 (02) :141-162
[8]   Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections [J].
Fischer, Sebastian ;
Vajda, Zsolt ;
Perez, Marta Aguilar ;
Schmid, Elisabeth ;
Hopf, Nikolai ;
Baezner, Hansjoerg ;
Henkes, Hans .
NEURORADIOLOGY, 2012, 54 (04) :369-382
[9]  
Fox B, 2014, J NEUROINTERV SURG, V7, pe37
[10]   Aneurysms of the vertebral artery (VA) [J].
Hernesniemi, J ;
Karatas, A ;
Niemelä, M ;
Ishii, K ;
Shen, H ;
Kivisaari, R ;
Rinne, J ;
Lehecka, M ;
Kivipelto, L .
ZENTRALBLATT FUR NEUROCHIRURGIE, 2005, 66 (04) :223-229