摘要
A highly selective, sensitive a rid robust LC-MS/MS method was developed for the simultaneous quantification of cortisol, cortisone, prednisolone and prednisone in human plasma. Prednisolone. cortisol and cortisone have similar fragmentation pattern. These three compounds were chromatographically separated, thus eliminating the inherent interference that fragments derived from the M + 2 and M isotopes of prednisolone contribute in the MRM channels of cortisol and cortisone, respectively. Additionally, by using a small particle (1.8 mu m) analytical column, interferences present in the plasma samples from post-transplant recipients were Successfully resolved from cortisol after a simple extraction consisting of protein precipitation. evaporation and reconstitution, The chromatographic separation was achieved oil a Zorbax-SB Phenyl column under isocratic conditions during a Full time of 8 min. intra-run and inter-run precision and accuracy Within +/- 15% were achieved during a 3-run validation for quality control samples at five concentration levels in charcoal-stripped plasma as well as in normal plasma, over a 500-fold dynamic concentration range. The lower limit of quantitation was 0.500 ng/mL for cortisone and prednisone. 1.00 ng/mL for cortisol and 2.00 ng/mL for prednisolone. The performance of I lie small particle column was maintained during more than 1200 injections ill terms of peak retention time, symmetry and backpressure. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:765 / 772
页数:8
相关论文
共 25 条
[1]  
[Anonymous], TRANSPLANTATION S168
[2]   RADIOIMMUNOASSAY FOR PREDNISOLONE [J].
COLBURN, WA ;
BULLER, RH .
STEROIDS, 1973, 21 (06) :833-846
[3]   Corticosteroids and kidney transplantation [J].
Curtis, John .
CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2006, 1 (05) :907-908
[4]   Quantitative determination of several synthetic corticosteroids by gas chromatography mass spectrometry after purification by immunoaffinity chromatography [J].
Delahaut, P ;
Jacquemin, P ;
Colemonts, Y ;
Dubois, M ;
DeGraeve, J ;
Deluyker, H .
JOURNAL OF CHROMATOGRAPHY B, 1997, 696 (02) :203-215
[5]   Simultaneous determination of cortisol, dexamethasone, methylprednisolone, prednisone, prednisolone, mycophenolic acid and mycophenolic acid glucuronide, in human plasma utilizing liquid chromatography-tandem mass spectrometry [J].
DiFrancesco, Robin ;
Frerichs, Valerie ;
Donnelly, Julie ;
Hagler, Colleen ;
Hochreiter, Jill ;
Tornatore, Kathleen M. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 859 (01) :42-51
[6]   Determination of the glucocorticoids prednisone, prednisolone, dexamethasone, and cortisol in human serum using liquid chromatography coupled to tandem mass spectrometry [J].
Frerichs, VA ;
Tornatore, KM .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 802 (02) :329-338
[7]  
FREY FJ, 1979, CLIN CHEM, V25, P1944
[8]  
GOODWIN W E, 1962, Trans Am Assoc Genitourin Surg, V54, P116
[9]  
HARIHARAN M, 1992, CLIN CHEM, V38, P346
[10]  
Jusko W.J., 1992, Applied Pharmacokinetics: Principles of Therapeutics Drug Monitoring