Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry

被引:162
作者
Paelinck, L. [1 ]
De Neve, W. [1 ]
DeWagter, C. [1 ]
机构
[1] Ghent Univ Hosp, Dept Radiotherapy, B-9000 Ghent, Belgium
关键词
D O I
10.1088/0031-9155/52/1/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the value of a commercially available flatbed scanner for film dosimetry with radiochromic film for external radiotherapy. The EPSON Pro 1680 Expression scanner was examined as a densitometer for two-dimensional film dosimetry with Gafchromic EBT film. An accurate and efficient scanning procedure was established. Possible drift and warm-up effects of the scanner were studied and the direct physical influence of the scanner light on the radiochromic film was assessed. Next, we investigated the scan field uniformity. Also, we examined if the accuracy of radiochromic film was improved by subtracting the optical density of the unirradiated blank film from the optical density of the irradiated film. To assess the accuracy of Gafchromic EBT film when the EPSON scanner was used as a densitometer, the depth dose of a 2 x 15 cm(2) field and the in-plane and cross-plane profiles of a 15 x 15 cm(2) field were measured and compared with diamond detector measurements. When taking consecutive scans, we found that the optical density taken from the first scan was about 1% higher than the optical density taken from subsequent scans. We attribute this to the warming up of the lamp of the scanner. Longer-term drift of the scanner was found to be absent. We found that the use of a correction matrix was necessary to correct for the non-uniform scanner response over the scan field. Subtracting the optical density of the unirradiated blank film from the irradiated film improves the precision of the Gafchromic EBT film. Depth dose and profile measurements with Gafchromic EBT film and the diamond detector are in agreement within 2.5%. The EPSON Pro 1680 Expression scanner is an excellent tool for accurate two-dimensional film dosimetry with Gafchromic EBT film provided that some precautions and corrections are taken into account.
引用
收藏
页码:231 / 242
页数:12
相关论文
共 50 条
  • [41] A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry
    Poppinga, D.
    Schoenfeld, A. A.
    Doerner, K. J.
    Blanck, O.
    Harder, D.
    Poppe, B.
    MEDICAL PHYSICS, 2014, 41 (02) : 021707
  • [42] Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure
    Menegotti, L.
    Delana, A.
    Martignano, A.
    MEDICAL PHYSICS, 2008, 35 (07) : 3078 - 3085
  • [43] Correction of lateral response artifacts from flatbed scanners for dual-channel radiochromic film dosimetry
    Akino, Yuichi
    Shiomi, Hiroya
    Isohashi, Fumiaki
    Suzuki, Osamu
    Seo, Yuji
    Tamari, Keisuke
    Hirata, Takero
    Mizuno, Hirokazu
    Ogawa, Kazuhiko
    JOURNAL OF RADIATION RESEARCH, 2021, 62 (02) : 319 - 328
  • [44] RADIOCHROMIC FILM DOSIMETRY WITH FLATBED SCANNERS: A FAST AND ACCURATE METHOD FOR DOSE CALIBRATION AND UNIFORMITY CORRECTION WITH SINGLE FILM EXPOSURE
    Menegotti, L.
    Delana, A.
    Vanoni, V.
    Martignano, A.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S391 - S391
  • [45] Radiochromic Film Dosimetry Update
    Niroomand-Rad, A.
    Chiu-Tsao, S.
    Trichter, S.
    Das, I.
    MEDICAL PHYSICS, 2015, 42 (06) : 3626 - 3626
  • [46] Eigencolor radiochromic film dosimetry
    Bouchard, Hugo
    Billas, Ilias
    Subiel, Anna
    Duane, Simon
    MEDICAL PHYSICS, 2021, 48 (05) : 2592 - 2603
  • [47] Improving Radiochromic Film's Sensitivity Limit with an Optical Common-mode Rejection and a Flatbed Color Scanner
    Ohuchi, Hiroko
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2008, : 167 - 170
  • [48] The performance of an optical cone-beam CT scanner adapted for radiochromic film dosimetry
    Babic, Steven
    Jordan, Kevin
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21) : N377 - N389
  • [49] In Situ Radiochromic Film Dosimetry Using Portable Spectrophotometer
    Jung, S.
    Cho, J.
    Han, J.
    Kim, J.
    Park, J.
    Choi, C.
    Son, J.
    Kim, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [50] Characterisation of EPSONV700 flatbed scanner for EBT3 Gafchromic film dosimetry
    Nelson, V.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S737 - S738