Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry

被引:161
|
作者
Paelinck, L. [1 ]
De Neve, W. [1 ]
DeWagter, C. [1 ]
机构
[1] Ghent Univ Hosp, Dept Radiotherapy, B-9000 Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 01期
关键词
D O I
10.1088/0031-9155/52/1/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the value of a commercially available flatbed scanner for film dosimetry with radiochromic film for external radiotherapy. The EPSON Pro 1680 Expression scanner was examined as a densitometer for two-dimensional film dosimetry with Gafchromic EBT film. An accurate and efficient scanning procedure was established. Possible drift and warm-up effects of the scanner were studied and the direct physical influence of the scanner light on the radiochromic film was assessed. Next, we investigated the scan field uniformity. Also, we examined if the accuracy of radiochromic film was improved by subtracting the optical density of the unirradiated blank film from the optical density of the irradiated film. To assess the accuracy of Gafchromic EBT film when the EPSON scanner was used as a densitometer, the depth dose of a 2 x 15 cm(2) field and the in-plane and cross-plane profiles of a 15 x 15 cm(2) field were measured and compared with diamond detector measurements. When taking consecutive scans, we found that the optical density taken from the first scan was about 1% higher than the optical density taken from subsequent scans. We attribute this to the warming up of the lamp of the scanner. Longer-term drift of the scanner was found to be absent. We found that the use of a correction matrix was necessary to correct for the non-uniform scanner response over the scan field. Subtracting the optical density of the unirradiated blank film from the irradiated film improves the precision of the Gafchromic EBT film. Depth dose and profile measurements with Gafchromic EBT film and the diamond detector are in agreement within 2.5%. The EPSON Pro 1680 Expression scanner is an excellent tool for accurate two-dimensional film dosimetry with Gafchromic EBT film provided that some precautions and corrections are taken into account.
引用
收藏
页码:231 / 242
页数:12
相关论文
共 50 条
  • [21] On Improving the Accuracy of EBT2 Film Dosimetry Using a Flatbed Scanner
    Pawlicki, T.
    Whitaker, M.
    Kim, G.
    MEDICAL PHYSICS, 2010, 37 (06) : 3248 - 3249
  • [22] Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC® film
    Lynch, Bart D.
    Kozelka, Jakub
    Ranade, Manisha K.
    Li, Jonathan G.
    Simon, William E.
    Dempsey, James F.
    MEDICAL PHYSICS, 2006, 33 (12) : 4551 - 4556
  • [23] Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry
    Lewis, David
    Chan, Maria F.
    MEDICAL PHYSICS, 2015, 42 (01) : 416 - 429
  • [24] Precise radiochromic film dosimetry using a flat-bed document scanner
    Devic, S
    Seuntjens, J
    Sham, E
    Podgorsak, EB
    Schmidtlein, CR
    Kirov, AS
    Soares, CG
    MEDICAL PHYSICS, 2005, 32 (07) : 2245 - 2253
  • [25] Temporal Characterization of the Flat-Bed Scanner Influencing Dosimetry Using Radiochromic Film
    Gotanda, Rumi
    Katsuda, Toshizo
    Gotanda, Tatsuhiro
    Tanki, Nobuyoshi
    Yatake, Hidetoshi
    Kawaji, Yasuyuki
    Kuwano, Tadao
    Akagawa, Takuya
    Tabuchi, Akihiko
    Ono, Atsushi
    Arao, Shinichi
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 537 - 539
  • [26] Evaluation of spatial resolution in image acquisition by optical flatbed scanners for radiochromic film dosimetry
    Asero, G.
    Greco, C.
    Gueli, A. M.
    Raffaele, L.
    Spampinato, S.
    JOURNAL OF INSTRUMENTATION, 2016, 11
  • [27] Reference dosimetry using radiochromic film
    Girard, Frederic
    Bouchard, Hugo
    Lacroix, Frederic
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (06): : 339 - 353
  • [28] Evaluation of a new ractiochromic Film/Flatbed scanner system for planar clinical dosimetry
    Goss, M.
    Oldham, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2449 - 2449
  • [29] Effect of scanner lens on lateral response artefact in radiochromic film dosimetry
    Tarafder Shameem
    Nick Bennie
    Martin Butson
    David Thwaites
    Physical and Engineering Sciences in Medicine, 2022, 45 : 721 - 727
  • [30] Effect of scanner lens on lateral response artefact in radiochromic film dosimetry
    Shameem, Tarafder
    Bennie, Nick
    Butson, Martin
    Thwaites, David
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (3) : 721 - 727