Determination of Pb, Cr, Cd, and As in Aluminum-Plastic Packaging Materials via Inductively Coupled Plasma-Mass Spectrometry with Microwave Digestion

被引:3
作者
Mi Hai-peng [1 ]
Zhu Hong-ming [1 ]
Li Gen-rong [1 ]
Yu Xiang [1 ]
Ma Bing-bing [2 ]
Zhou Xi-lin [2 ]
Wang Ya-sen [2 ]
Su Zhong-hua [2 ]
Deng Xiong [2 ]
机构
[1] Chongqing Acad Metrol & Qual Inspect, Chongqing 401123, Peoples R China
[2] Natl Qual Supervis & Inspect Ctr Aluminum Magnesi, Chongqing 408403, Peoples R China
关键词
Microwave digestion; ICP-MS; Aluminum-plastic packaging; Lead; Chromium; Cadmium; Arsenic; ICP-MS; ELEMENTS;
D O I
10.3964/j.issn.1000-0593(2017)02-0646-05
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Microwave digestion was performed to study the pretreatment methods of aluminum-plastic packaging materials (APPMs). Five different digestion reagent combinations and proportions were thoroughly considered. Digestion results indicated that the most suitable reagent combination was sulfuric and nitric acids with the optimal proportion of 1 7 after the orthogonal experiment. Moreover, the possible reasons of the experimental phenomenon were analyzed. The contents of Pb, Cr, Cd, and As in APPMs were subsequently determined via inductively coupled plasma-mass spectrometry (ICP-MS). The satisfactory linearity of calibration curves was obtained with the linear correlation coefficients above 0. 999 5, and the instrument detection limits of Pb, Cr, Cd, and As for the current method were 0. 215, 0. 067, 0. 006 and 0. 020 ng.mL(-1), respectively. Furthermore, the recoveries of standard addition ranged from 83. 8% to 111. 6%, and the relative standard deviations ranged from 0. 5% to 7. 4%. Two independent parallel determination results of Pb, Cr, Cd, and As in APPMs were approaching, and the student's t-test (confidence level, alpha=0. 05) showed that the determination results had no significant differences. In conclusion, the present method exhibited fine linearity, low detection limit, high recovery, and good precision, which can accurately be utilized to analyze Pb, Cr, Cd, and As elements in APPMs or other similar materials.
引用
收藏
页码:646 / 650
页数:5
相关论文
共 7 条
[1]   Determination of trace element concentrations in tomato samples at different stages of maturation by ICP OES and ICP-MS following microwave-assisted digestion [J].
Bressy, Fernanda C. ;
Brito, Geysa B. ;
Barbosa, Isa S. ;
Teixeira, Leonardo S. G. ;
Korn, Maria Gracas A. .
MICROCHEMICAL JOURNAL, 2013, 109 :145-149
[2]   Determining trace metal elements in the tooth enamel from Hui and Han Ethnic groups in China using microwave digestion and inductively coupled plasma mass spectrometry (ICP-MS) [J].
He, Maoyong ;
Lu, Hai ;
Luo, Chongguang ;
Ren, Tongxiang .
MICROCHEMICAL JOURNAL, 2016, 127 :142-144
[3]  
Julshamn K, 2007, J AOAC INT, V90, P844
[4]   Investigation by ICP-MS of trace element levels in vegetable edible oils produced in Spain [J].
Llorent-Martinez, E. J. ;
Ortega-Barrales, P. ;
Fernandez-de Cordova, M. L. ;
Dominguez-Vidal, A. ;
Ruiz-Medina, A. .
FOOD CHEMISTRY, 2011, 127 (03) :1257-1262
[5]  
Song M, 2014, ABSTR APPL ANAL, V2014, P3, DOI DOI 10.1016/J.NEULET.2014.03.071
[6]   Determination of Trace Element Silver in Animal Serum, Tissues and Organs by Microwave Digestion-ICP-MS [J].
Yuan Jun-jie ;
Xie You-zhuan ;
Han Chen ;
Sun Wei ;
Zhang Kai ;
Zhao Jie ;
Lu Xiao ;
Lu Jian-xi ;
Ren Wei .
SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34 (09) :2533-2537
[7]   Determination of free amino acids and 18 elements in freeze-dried strawberry and blueberry fruit using an Amino Acid Analyzer and ICP-MS with micro-wave digestion [J].
Zhang Hua ;
Wang Zhen-Yu ;
Yang Xin ;
Zhao Hai-Tian ;
Zhang Ying-Chun ;
Dong Ai-Jun ;
Jing Jing ;
Wang Jing .