Evaluation of CdZnTe as neutron detector around medical accelerators

被引:15
作者
Martin-Martin, A. [1 ]
Iniguez, M. P. [1 ]
Luke, P. N. [2 ]
Barquero, R. [3 ]
Lorente, A. [4 ]
Morchon, J. [1 ]
Gallego, E. [4 ]
Quincoces, G. [5 ]
Marti-Climent, J. M. [5 ]
机构
[1] Univ Valladolid, Dept Fis Teor Atom & Opt, E-47011 Valladolid, Spain
[2] Univ Calif Berkeley, Lawrence Berkeley Natl Lab, Berkeley, CA 94720 USA
[3] Hosp Univ Rio Hortega, Valladolid, Spain
[4] Univ Politecn Madrid, Dept Ingn Nucl, Madrid, Spain
[5] Univ Navarra Clin, Pamplona, Spain
关键词
CYCLOTRON VAULT ROOM; PET CYCLOTRON; SEMICONDUCTOR-DETECTORS; SPECTRA; LINAC; SPECTROMETER;
D O I
10.1093/rpd/ncp038
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The operation of electron linear accelerators (LINACs) and cyclotrons can produce a mixed gamma-neutron field composed of energetic neutrons coming directly from the source and scattered lower energy neutrons. The thermal neutron detection properties of a non-moderated coplanar-grid CdZnTe (CZT) gamma-ray detector close to an 18 MV electron LINAC and an 18 MeV proton cyclotron producing the radioisotope F-18 for positron emission tomography are investigated. The two accelerators are operated at conditions producing similar thermal neutron fluence rates of the order of 10(4) cm(-2) s(-1) at the measurement locations. The counting efficiency of the CZT detector using the prompt 558 keV photopeak following Cd-113 thermal neutron capture is evaluated and a good neutron detection performance is found at the two installations.
引用
收藏
页码:193 / 199
页数:7
相关论文
共 27 条
[1]   Monte Carlo neutron doses estimations inside a PET cyclotron vault room [J].
Barquero, R. ;
Mendez, R. ;
Marti-Climent, J. M. ;
Quincoces, G. .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :477-481
[2]   Monte Carlo simulation estimates of neutron doses to critical organs of a patient undergoing 18 MV x-ray LINAC-based radiotherapy [J].
Barquero, R ;
Edwards, TM ;
Iñiguez, MP ;
Vega-Carrillo, HR .
MEDICAL PHYSICS, 2005, 32 (12) :3579-3588
[3]   Neutron spectra and dosimetric features around an 18 MV Linac accelerator [J].
Barquero, R ;
Mendez, R ;
Vega-Carrillo, HR ;
Iñiguez, MP ;
Edwards, TM .
HEALTH PHYSICS, 2005, 88 (01) :48-58
[4]   Neutron detection with cryogenics and semiconductors [J].
Bell, ZW ;
Carpenter, DA ;
Cristy, SS ;
Lamberti, VE ;
Burger, A ;
Woodfield, BF ;
Niedermayr, T ;
Han, ID ;
Labov, SE ;
Friedrich, S ;
West, WG ;
Pohl, KR ;
van den Berg, L .
PHYSICA STATUS SOLIDI C - CONFERENCES AND CRITICAL REVIEWS, VOL 2, NO 5, 2005, 2 (05) :1592-1605
[5]   Progress in dosimetry of neutrons and light nuclei [J].
Blomgren, Jan ;
Lindborg, Lennart ;
Golnik, Natalia ;
Jones, Dan ;
Schuhmacher, Helmut ;
Spurny, Frantisek ;
Stenerlow, Bo .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :1-2
[6]   A NEW TYPE OF NEUTRON SPECTROMETER [J].
BRAMBLETT, RL ;
EWING, RI ;
BONNER, TW .
NUCLEAR INSTRUMENTS & METHODS, 1960, 9 (01) :1-12
[7]  
BRIESMEISTER J, 2000, LA13709M LOS AL NAC
[8]   Neutron spectrometry - historical review and present status [J].
Brooks, FD ;
Klein, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 476 (1-2) :1-11
[9]   The neutron fluence and H*(10) response of the new LB 6411 rem counter [J].
Burgkhardt, B ;
Fieg, G ;
Klett, A ;
Plewnia, A ;
Siebert, BRL .
RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) :361-364
[10]  
CARROLL LR, 2004, P 9 INT WORKSH TARG