Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance

被引:20
作者
Tao, Ashley [1 ,2 ]
Huang, Richard [1 ,3 ]
Tao, Shengzhen [1 ]
Michalak, Gregory J. [1 ]
McCollough, Cynthia H. [1 ]
Leng, Shuai [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Gundersen Hlth Syst, La Crosse, WI 54601 USA
[3] Univ Southern Calif, Los Angeles, CA 90007 USA
关键词
multi-energy CT; photon counting detector CT; iodine quantification; dual-source CT; spectral separation; tin filter; COMPUTED-TOMOGRAPHY; ENERGY CT; EXPERIMENTAL FEASIBILITY; MATERIAL SEPARATION;
D O I
10.1088/1361-6560/ab1c34
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photon counting detectors (PCD) can provide spectral information to enable iodine quantification through multi-energy imaging but performance is limited by current PCD technology. The purpose of this work is to evaluate iodine quantification in a phantom study using dual-source PCD-CT (DS-PCD-CT), and compare to single-source (SS)-PCD-CT and traditional DS energy integrating detector (EID)-based dual-energy CT. A multi-energy CT phantom with iodine inserts (0 to 15 mg ml(-1) concentration) was imaged on a research SS-PCD-CT scanner (CTDIvol = 18 mGy). A DS-PCD-CT was emulated by acquiring two sequential scans (CTDIvol = 9 mGy each) using tube potentials: 140 kVp/80 kVp, 140 kVp/100 kVp and 140 kVp/120 kVp. For each kVp, 1 or 2 energy bins were reconstructed to achieve either dual-energy or quadruple energy CT. In addition to these energy combinations, a Sn filter was used for the high tube potential (140 kVp) of each kVp pair. For comparison, the same phantom was also scanned on a commercially available DS-EID-CT with matched radiation dose (CTDIvol = 18 mGy). Material decomposition was performed in image space using a standard least-squares based approach to generate iodine and water-specific images. The root-mean-square-error (RMSE) measured over each insert from the iodine image was used to determine iodine accuracy. The iodine RMSE from SS-PCD (140 kVp with 2 energy bins) was 2.72 mg ml(-1). The use of a DS configuration with 1 energy bin per kVp (140 kVp/80 kVp) resulted in a RMSE of 2.29 mg ml(-1). Two energy bins per kVp further reduced iodine RMSE to 1.83 mg ml(-1). The addition of a Sn filter to the latter quadruple energy mode reduced RMSE to 1.48 mg ml(-1). RMSE for DS-PCD-CT (2 energy bins per kVp) decreased by 1.3% (Sn140 kVp/80 kVp) and 15% (Sn140 kVp/100 kVp) as compared to DS-EID-CT. DS-PCD-CT with a Sn filter improved iodine quantification as compared to both SS-PCD-CT and DS-EID-CT.
引用
收藏
页数:10
相关论文
共 30 条
  • [1] Spectroscopic (multi-energy) CT distinguishes iodine and barium contrast material in MICE
    Anderson, N. G.
    Butler, A. P.
    Scott, N. J. A.
    Cook, N. J.
    Butzer, J. S.
    Schleich, N.
    Firsching, M.
    Grasset, R.
    de Ruiter, N.
    Campbell, M.
    Butler, P. H.
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (09) : 2126 - 2134
  • [2] Carmi R, 2005, IEEE NUCL SCI CONF R, P1876
  • [3] Lesion insertion in the projection domain: Methods and initial results
    Chen, Baiyu
    Leng, Shuai
    Yu, Lifeng
    Yu, Zhicong
    Ma, Chi
    McCollough, Cynthia
    [J]. MEDICAL PHYSICS, 2015, 42 (12) : 7034 - 7042
  • [4] Experimental feasibility of spectral photon-counting computed tomography with two contrast agents for the detection of endoleaks following endovascular aortic repair
    Dangelmaier, Julia
    Bar-Ness, Daniel
    Daerr, Heiner
    Muenzel, Daniela
    Si-Mohamed, Salim
    Ehn, Sebastian
    Fingerle, Alexander A.
    Kimm, Melanie A.
    Kopp, Felix K.
    Boussel, Loic
    Roessl, Ewald
    Pfeiffer, Franz
    Rummeny, Ernst J.
    Proksa, Roland
    Douek, Philippe
    Noel, Peter B.
    [J]. EUROPEAN RADIOLOGY, 2018, 28 (08) : 3318 - 3325
  • [5] Performance of today's dual energy CT and future multi energy CT in virtual non-contrast imaging and in iodine quantification: A simulation study
    Faby, Sebastian
    Kuchenbecker, Stefan
    Sawall, Stefan
    Simons, David
    Schlemmer, Heinz-Peter
    Lell, Michael
    Kachelriess, Marc
    [J]. MEDICAL PHYSICS, 2015, 42 (07) : 4349 - 4366
  • [6] Characterization of Urinary Stone Composition by Use of Whole-body, Photon-counting Detector CT
    Ferrero, Andrea
    Gutjahr, Ralf
    Halaweish, Ahmed F.
    Leng, Shuai
    McCollough, Cynthia H.
    [J]. ACADEMIC RADIOLOGY, 2018, 25 (10) : 1270 - 1276
  • [7] First performance evaluation of a dual-source CT (DSCT) system
    Flohr, TG
    McCollough, CH
    Bruder, H
    Petersilka, M
    Gruber, K
    Süss, C
    Grasruck, M
    Stierstorfer, K
    Krauss, B
    Raupach, R
    Primak, AN
    Küttner, A
    Achenbach, S
    Becker, C
    Kopp, A
    Ohnesorge, BM
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (02) : 256 - 268
  • [8] Human Imaging With Photon Counting-Based Computed Tomography at Clinical Dose Levels: Contrast-to-Noise Ratio and Cadaver Studies
    Gutjahr, Ralf
    Halaweish, Ahmed F.
    Yu, Zhicong
    Leng, Shuai
    Yu, Lifeng
    Li, Zhoubo
    Jorgensen, Steven M.
    Ritman, Erik L.
    Kappler, Steffen
    McCollough, Cynthia H.
    [J]. INVESTIGATIVE RADIOLOGY, 2016, 51 (07) : 421 - 429
  • [9] Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging
    Iwanczyk, Jan S.
    Nygard, Einar
    Meirav, Oded
    Arenson, Jerry
    Barber, William C.
    Hartsough, Neal E.
    Malakhov, Nail
    Wessel, Jan C.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 535 - 542
  • [10] A research prototype system for quantum-counting clinical CT
    Kappler, S.
    Glasser, F.
    Janssen, S.
    Kraft, E.
    Reinwand, M.
    [J]. MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622