Calibration of a wide-field frequency-domain fluorescence lifetime microscopy system using light emitting diodes as light sources

被引:34
作者
Elder, A. D.
Frank, J. H.
Swartling, J.
Dai, X.
Kaminski, C. F.
机构
[1] Univ Cambridge, Dept Chem Engn, Cambridge CB2 3RA, England
[2] Sandia Natl Labs, Livermore, CA USA
关键词
D O I
10.1111/j.1365-2818.2006.01689.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
High brightness light emitting diodes are an inexpensive and versatile light source for wide-field frequency-domain fluorescence lifetime imaging microscopy. In this paper a full calibration of an LED based fluorescence lifetime imaging microscopy system is presented for the first time. A radio-frequency generator was used for simultaneous modulation of light emitting diode (LED) intensity and the gain of an intensified charge coupled device (CCD) camera. A homodyne detection scheme was employed to measure the demodulation and phase shift of the emitted fluorescence, from which phase and modulation lifetimes were determined at each image pixel. The system was characterized both in terms of its sensitivity to measure short lifetimes (500 ps to 4 ns), and its capability to distinguish image features with small lifetime differences. Calibration measurements were performed in quenched solutions containing Rhodamine 6G dye and the results compared to several independent measurements performed with other measurement methodologies, including time correlated single photon counting, time gated detection, and acousto optical modulator (AOM) based modulation of excitation sources. Results are presented from measurements and simulations. The effects of limited signal-to-noise ratios, baseline drifts and calibration errors are discussed in detail. The implications of limited modulation bandwidth of high brightness, large area LED devices (similar to 40 MHz for devices used here) are presented. The results show that phase lifetime measurements are robust down to sub ns levels, whereas modulation lifetimes are prone to errors even at large signal-to-noise ratios. Strategies for optimizing measurement fidelity are discussed. Application of the fluorescence lifetime imaging microscopy system is illustrated with examples from studies of molecular mixing in microfluidic devices and targeted drug delivery research.
引用
收藏
页码:166 / 180
页数:15
相关论文
共 50 条
[41]   Wide-field photon counting fluorescence lifetime imaging microscopy: application to photosynthesizing systems [J].
Zdeněk Petrášek ;
Hann-Jörg Eckert ;
Klaus Kemnitz .
Photosynthesis Research, 2009, 102 :157-168
[42]   Optically sectioned wide-field fluorescence lifetime imaging microscopy enabled by structured illumination [J].
Hinsdale, Taylor ;
Olsovsky, Cory ;
Rico-Jimenez, Jose J. ;
Maitland, Kristen C. ;
Jo, Javier A. ;
Malik, Bilal H. .
BIOMEDICAL OPTICS EXPRESS, 2017, 8 (03) :1455-1465
[43]   Spatially resolved electric fields in polymer light-emitting diodes using fluorescence lifetime imaging [J].
Koeberg, M ;
Elson, DS ;
French, PMW ;
Bradley, DDC .
SYNTHETIC METALS, 2003, 139 (03) :925-928
[44]   Microscopy using micro-pixellated light emitting diodes [J].
Poher, V. ;
Kennedy, G. T. ;
Elson, D. S. ;
French, P. M. W. ;
Neil, A. A. A. .
THREE-DIMENSIONAL AND MULTIDIMENSIONAL MICROSCOPY: IMAGE ACQUISITION AND PROCESSING XIV, 2007, 6443
[45]   wFLFM: enhancing the resolution of Fourier light-field microscopy using a hybrid wide-field image [J].
Liu, Wenhao ;
Jia, Shu .
APPLIED PHYSICS EXPRESS, 2021, 14 (01)
[46]   A wide-field time-domain fluorescence lifetime imaging microscope with optical sectioning [J].
Webb, SED ;
Gu, Y ;
Léveque-Fort, S ;
Siegel, J ;
Cole, MJ ;
Dowling, K ;
Jones, R ;
French, PMW ;
Neil, MAA ;
Juskaitis, R ;
Sucharov, LOD ;
Wilson, T ;
Lever, MJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (04) :1898-1907
[47]   Frequency-domain fluorescence spectroscopy using 280-nm and 300-nm light-emitting diodes: Measurement of proteins and protein-related fluorophores [J].
Barbieri, B ;
Terpetschnig, E ;
Jameson, DM .
ANALYTICAL BIOCHEMISTRY, 2005, 344 (02) :298-300
[48]   FLUORESCENCE LIFETIME RESOLUTION OF SPECTRA IN THE FREQUENCY-DOMAIN USING MULTIWAY ANALYSIS [J].
MILLICAN, DW ;
MCGOWN, LB .
ANALYTICAL CHEMISTRY, 1990, 62 (20) :2242-2247
[49]   Fluorescence microscopy investigation of InGaN-based light-emitting diodes [J].
Presa, Silvino ;
Maaskant, Pleun P. ;
Kappers, Menno J. ;
Corbett, Brian .
IET OPTOELECTRONICS, 2016, 10 (02) :39-43
[50]   Exploiting new excitation strategies in fluorescence microscopy operated by light emitting diodes [J].
Angelini, M. ;
Mazzini, G. .
CYTOMETRY PART A, 2008, 73A (01) :98-98