Quantification of polarization in biological tissue by optical coherence tomography

被引:1
|
作者
Jiao, SL [1 ]
Wang, LV [1 ]
机构
[1] Texas A&M Univ, Biomed Engn Program, Opt Imaging Lab, College Stn, TX 77843 USA
来源
LASER TISSUE INTERACTION XIII: PHOTOCHEMICAL, PHOTOTHERMAL, AND PHOTOMECHANICAL | 2002年 / 4617卷
关键词
ellipsometry and polarimetry; optical coherence tomography;
D O I
10.1117/12.472525
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A multiple-channel polarization-sensitive system based on optical coherence tomography was built to measure the Mueller matrix of scattering biological tissue. The Jones matrix of a sample can be determined with a single scan and subsequently converted into an equivalent non-depolarizing Mueller matrix. As a result, the system can be used to measure the Mueller matrix of soft tissue. Birefringence, axis orientation, and diattenuation can be extracted.