Reverse total shoulder arthroplasty component center of rotation affects muscle function

被引:38
作者
Hoenecke, Heinz R., Jr. [1 ]
Flores-Hernandez, Cesar [2 ]
D'Lima, Darryl D. [2 ]
机构
[1] Scripps Clin, Div Orthopaed Surg, La Jolla, CA 92037 USA
[2] Scripps Clin, Shiley Ctr Orthopaed Res & Educ, La Jolla, CA 92037 USA
关键词
Shoulder arthroplasty; reverse shoulder arthroplasty; glenosphere; deltoid function; glenohumeral forces; shoulder biomechanics; CUFF DEFICIENCY; PROSTHESIS; IMPINGEMENT; ARTHRITIS; ABDUCTION; DESIGN; MOTION; RANGE;
D O I
10.1016/j.jse.2013.11.025
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Medialization of the glenohumeral center of rotation alters the moment arm of the deltoid, can affect muscle function, and increases the risk for scapular notching due to impingement. The objective of this study was to determine the effect of position of the glenosphere on deltoid efficiency and the range of glenohumeral adduction. Methods: Scapulohumeral bone models were reconstructed from computed tomography scans and virtually implanted with primary or reverse total shoulder arthroplasty implants. The placement of the glenosphere was varied to simulate differing degrees of "medialization'' and inferior placement relative to the glenoid. Muscle and joint forces were computed during shoulder abduction in OpenSim musculoskeletal modeling software. Results: The average glenohumeral joint reaction forces for the primary total shoulder arthroplasty were within 5% of those previously reported in vivo. Superior placement or full lateralization of the glenosphere increased glenohumeral joint reaction forces by 10% and 18%, respectively, relative to the recommended reverse total shoulder arthroplasty position. The moment arm of the deltoid muscle was the highest at the recommended baseline surgical position. The baseline glenosphere position resulted in a glenohumeral adduction deficit averaging more than 10 degrees that increased to more than 25 degrees when the glenosphere was placed superiorly. Only with full lateralization was glenohumeral adduction unaffected by superoinferior placement. Discussion/Conclusion: Selecting optimum placement of the glenosphere involves tradeoffs in bending moment at the implant-bone interface, risk for impingement, and deltoid efficiency. A viable option is partially medializing the glenosphere, which retains most of the benefits of deltoid efficiency and reduces the risk for scapular notching. (C) 2014 Journal of Shoulder and Elbow Surgery Board of Trustees.
引用
收藏
页码:1128 / 1135
页数:8
相关论文
共 21 条
[1]   In vivo glenohumeral contact forces -: Measurements in the first patient 7 months postoperatively [J].
Bergmann, G. ;
Graichen, F. ;
Bender, A. ;
Kaab, M. ;
Rohlmann, A. ;
Westerhoff, P. .
JOURNAL OF BIOMECHANICS, 2007, 40 (10) :2139-2149
[2]   Grammont reverse prosthesis: Design, rationale, and biomechanics [J].
Boileau, P ;
Watkinson, DJ ;
Hatzidakis, AM ;
Balg, F .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2005, 14 (01) :147S-161S
[3]   TENDONS, LIGAMENTS, AND CAPSULE OF THE ROTATOR CUFF - GROSS AND MICROSCOPIC ANATOMY [J].
CLARK, JM ;
HARRYMAN, DT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (05) :713-725
[4]   Reverse shoulder arthroplasty for the treatment of rotator cuff deficiency [J].
Cuff, Derek ;
Pupello, Derek ;
Virani, Nazeem ;
Levy, Jonathan ;
Frankle, Mark .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (06) :1244-1251
[5]   Shoulder prostheses treating cuff tear arthropathy: a comparative biomechanical study [J].
De Wilde, LF ;
Audenaert, EA ;
Berghs, BM .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (06) :1222-1230
[6]   OpenSim: open-source software to create and analyze dynamic Simulations of movement [J].
Delp, Scott L. ;
Anderson, Frank C. ;
Arnold, Allison S. ;
Loan, Peter ;
Habib, Ayman ;
John, Chand T. ;
Guendelman, Eran ;
Thelen, Darryl G. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (11) :1940-1950
[7]   The reverse shoulder prosthesis for glenohumeral arthritis associated with severe rotator cuff deficiency - A minimum two-year follow-up study of sixty patients [J].
Frankle, M ;
Siegal, S ;
Pupello, D ;
Saleem, A ;
Mighell, M ;
Vasey, M .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08) :1697-1705
[8]   DELTA-SHOULDER PROSTHESIS FOR ROTATOR CUFF RUPTURE [J].
GRAMMONT, PM ;
BAULOT, E .
ORTHOPEDICS, 1993, 16 (01) :65-68
[9]   Evaluation of abduction range of motion and avoidance of inferior scapular impingement in a reverse shoulder model [J].
Gutierrez, Sergio ;
Levy, Jonathan C. ;
Frankle, Mark A. ;
Cuff, Derek ;
Keller, Tony S. ;
Pupello, Derek R. ;
Lee, William E., III .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2008, 17 (04) :608-615
[10]   Range of Impingement-Free Abduction and Adduction Deficit After Reverse Shoulder Arthroplasty Hierarchy of Surgical and Implant-Design-Related Factors [J].
Gutierrez, Sergio ;
Comiskey, Charles A., IV ;
Luo, Zong-Ping ;
Pupello, Derek R. ;
Frankle, Mark A. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (12) :2606-2615