How flatbed scanners upset accurate film dosimetry

被引:56
作者
van Battum, L. J. [1 ]
Huizenga, H. [2 ]
Verdaasdonk, R. M. [1 ]
Heukelom, S. [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Phys & Med Technol, Support Radiotherapy, POB 7057, NL-1007 MB Amsterdam, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Radiotherapy, NL-6525 ED Nijmegen, Netherlands
关键词
film dosimetry; Gafchromic; film; flatbed scanners; lateral correction; polarization; EBT; EBT2; EBT3; EBT2 RADIOCHROMIC FILM; GAFCHROMIC EBT2; PHOTON; ORIENTATION; SENSITIVITY; CALIBRATION; ARTIFACTS; PROTOTYPE; IMRT;
D O I
10.1088/0031-9155/61/2/625
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2-2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un) irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner's transmission mode, with red-green-blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner's optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the
引用
收藏
页码:625 / 649
页数:25
相关论文
共 50 条
  • [1] Aalbersm A H L, 2012, NCS18
  • [2] Scanning orientation and polarization effects for XRQA radiochromic film
    Alnawaf, Hani
    Butson, Martin J.
    Cheung, Tsang
    Yu, Peter K. N.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2010, 26 (04): : 216 - 219
  • [3] Improving the energy response of external beam therapy (EBT) GafChromic™ dosimetry films at low energies (≤100 keV)
    Bekerat, H.
    Devic, S.
    DeBlois, F.
    Singh, K.
    Sarfehnia, A.
    Seuntjens, J.
    Shih, Shelley
    Yu, Xiang
    Lewis, D.
    [J]. MEDICAL PHYSICS, 2014, 41 (02)
  • [4] Scanning orientation effects on Gafchromic EBT film dosimetry
    Butson M.J.
    Cheung T.
    Yu P.K.N.
    [J]. Australasian Physical and Engineering Sciences in Medicine, 2006, 29 (03) : 281 - 284
  • [5] Evaluation of the magnitude of EBT Gafchromic film polarization effects
    Butson, M. J.
    Cheung, T.
    Yu, P. K. N.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (01) : 21 - 25
  • [6] Absorption spectra variations of EBT radiochromic film from radiation exposure
    Butson, MJ
    Cheung, T
    Yu, PKN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13) : N135 - N140
  • [7] Polarization effects on a high-sensitivity radiochromic film
    Butson, MJ
    Yu, PKN
    Cheung, T
    Inwood, D
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (15) : N207 - N211
  • [8] Comparison of four commercial devices for RapidArc and sliding window IMRT QA
    Chandraraj, Varatharaj
    Stathakis, Sotirios
    Manickam, Ravikumar
    Esquivel, Carlos
    Supe, Sanjay S.
    Papanikolaou, Nikos
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02): : 338 - 349
  • [9] Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique
    Chang, Liyun
    Ho, Sheng-Yow
    Lee, Tsair-Fwu
    Yeh, Shyh-An
    Ding, Hueisch-Jy
    Chen, Pang-Yu
    [J]. MEDICAL PHYSICS, 2015, 42 (10) : 5838 - 5847
  • [10] Dose response characteristics of new models of GAFCHROMIC films: Dependence on densitometer light source and radiation energy
    Chiu-Tsao, ST
    Duckworth, T
    Zhang, CF
    Patel, NS
    Hsiung, CY
    Lin, W
    Shih, JA
    Harrison, LB
    [J]. MEDICAL PHYSICS, 2004, 31 (09) : 2501 - 2508