CT and MR image fusion of tandem and ring applicator using rigid registration in intracavitary brachytherapy planning

被引:7
作者
Oinam, Arun S. [1 ]
Tomar, Parsee [1 ]
Patel, Firuza D. [1 ]
Singh, Lakhwant [2 ]
Rai, Bhavana [1 ]
Bahl, Amit [1 ]
机构
[1] Post Grad Inst Med Educ & Res, Dept Radiotherapy, Chandigarh, Ut Chandigarh, India
[2] Guru Nanak Dev Univ Inst, Dept Phys, Amritsar, Punjab, India
关键词
autoradiograph; dwell position; MPR; intracavitary; interstitial brachytherapy; impact of registration error; CERVICAL-CANCER; GUIDED BRACHYTHERAPY; WORKING GROUP; RECONSTRUCTION; GUIDELINES; CARCINOMA; BLADDER; IMPACT; TRIAL;
D O I
10.1120/jacmp.v15i2.4206
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study is to find the uncertainties in the reconstruction of MR compatible ring-tandem intracavitary applicators of high-dose rate image-based brachytherapy treatment planning using rigid registration of 3D MR and CT image fusion. Tandem and ring reconstruction in MR image based brachytherapy planning was done using rigid registration of CT and MR applicator geometries. Verifications of registration for applicator fusion were performed in six verification steps at three different sites of tandem ring applicator set. The first site consists of three errors at the level of ring plane in (1) cranio-caudal shift (Cranial Shift) of ring plane along tandem axis, (2) antero-posterior shift (AP Shift) perpendicular to tandem axis on the plane containing the tandem, and (3) lateral shift (Lat Shift) perpendicular to the plane containing the tandem at the level of ring plane. The other two sites are the verifications at the tip of tandem and neck of the ring. The verification at the tip of tandem consists of two errors in (1) antero-posterior shift (AP Shift) perpendicular to tandem axis on the plane containing the tandem, and (2) lateral shift (Lat Shift) perpendicular to the plane containing the tandem. The third site of verification at the neck of the ring is the error due to the rotation of ring about tandem axis. The impact of translational errors from -5 mm to 5 mm in the step of 1 mm along x-, y-, and z-axis and three rotational errors about these axes from -19.1 degrees to 19.1 degrees in the step of 3.28 on dose-volume histogram parameters (D-2cc, D-1cc, D-0.1cc, and D-5cc of bladder, rectum, and sigmoid, and D-90 and D-98 of HRCTV were also analyzed. Maximum registration errors along cranio-caudal direction was 2.2 mm (1 case), whereas the errors of 31 out of 34 cases of registration were found within 1.5 mm, and those of two cases were less than 2 mm but greater than 1.5 mm. Maximum rotational error of ring about tandem axis was 3.15 degrees (1.1 mm). In other direction and different sites of the ring applicator set, the errors were within 1.5 mm. The impacts of registration errors on DVH parameters of bladder, rectum, and sigmoid were very sensitive to antero-posterior shift. Cranio-caudal errors of registration also largely affected the rectum DVH parameters. Largest change of 17.95% per mm and 20.65% per mm in all the DVH parameters of all OARs and HRCTV were observed for phi and Psi rotational errors as compare to other translational and rotational errors. Catheter reconstruction in MR image using rigid registration of applicator geometries of CT and MR images is a feasible technique for MR image-based intracavitary brachytherapy planning. The applicator regis-tration using the contours of tandem and neck of the ring of CT and MR images decreased the rotational error about tandem axis. Verification of CT MR image fusion using applicator registration which consists of six steps of verification at three different sites in ring applicator set can report all the errors due to translation and rotational shift along theta, phi, and Psi. phi and Psi rotational errors, which produced potential changes in DVH parameters, can be tackled using AP Shift and Lat Shift at the tip of tandem. The maximum shift was still found along the tandem axis in this technique.
引用
收藏
页码:191 / 204
页数:14
相关论文
共 16 条
[1]   Direct reconstruction of the Vienna applicator on MR images [J].
Berger, Daniel ;
Dimopoulos, Johannes ;
Poetter, Richard ;
Kirisits, Christian .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :347-351
[2]   Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of cervical cancer [J].
De Leeuw, Astrid A. C. ;
Moerland, Marinus A. ;
Nomden, Christel ;
Tersteeg, Robert H. A. ;
Roesink, Judith M. ;
Jurgenliemk-Schulz, Ira M. .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :341-346
[3]   Applicator reconstruction in MRI 3D image-based dose planning of brachytherapy for cervical cancer [J].
Haack, Soren ;
Nielsen, Soren Kynde ;
Lindegaard, Jacob Christian ;
Gelineck, John ;
Tanderup, Kari .
RADIOTHERAPY AND ONCOLOGY, 2009, 91 (02) :187-193
[4]   Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group* (I):: concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV [J].
Haie-Meder, C ;
Pötter, R ;
Van Limbergen, E ;
Briot, E ;
De Brabandere, M ;
Dimopoulos, J ;
Dumas, I ;
Hellebust, TP ;
Kirisits, C ;
Lang, SF ;
Muschitz, S ;
Nevinson, J ;
Nulens, A ;
Petrow, P ;
Wachter-Gerstner, N .
RADIOTHERAPY AND ONCOLOGY, 2005, 74 (03) :235-245
[5]   Reconstruction of a ring applicator using CT imaging: impact of the reconstruction method and applicator orientation [J].
Hellebust, Taran Paulsen ;
Tanderup, Kari ;
Bergstrand, Eva Stabell ;
Knutsen, Bjorn Helge ;
Roislien, Jo ;
Olsen, Dag Rune .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (16) :4893-4904
[6]   The Vienna applicator for combined intracavitary and interstitial brachytherapy of cervical cancer:: Design, application, treatment planning, and dosimetric results [J].
Kirisits, C ;
Lang, S ;
Dimopoulos, J ;
Berger, D ;
Georg, D ;
Pötter, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (02) :624-630
[7]   Dose and volume parameters for MRI-based treatment planning in intracavitary brachytherapy for cervical cancer [J].
Kirisits, C ;
Pötter, R ;
Lang, S ;
Dimopoulos, J ;
Wachter-Gerstner, N ;
Georg, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (03) :901-911
[8]  
Krempien TP, 2003, BRACHYTHERAPY, V2, P164
[9]   Proposed guidelines for image-based intracavitary brachytherapy for cervical carcinoma: Report from Image-Guided Brachytherapy Working Group [J].
Nag, S ;
Cardenes, H ;
Chang, S ;
Das, IJ ;
Erickson, B ;
Ibbott, GS ;
Lowenstein, J ;
Roll, J ;
Thomadsen, B ;
Varia, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (04) :1160-1172
[10]   Adaptive image guided brachytherapy for cervical cancer: A combined MRI-/CT-planning technique with MRI only at first fraction [J].
Nesvacil, Nicole ;
Poetter, Richard ;
Sturdza, Alina ;
Hegazy, Neamat ;
Federico, Mario ;
Kirisits, Christian .
RADIOTHERAPY AND ONCOLOGY, 2013, 107 (01) :75-81