cobas 8000 Modular Analyzer Series Evaluated under Routine-like Conditions at 14 Sites in Australia, Europe, and the United States

被引:24
作者
von Eckardstein, Arnold [1 ]
Roth, Hans Juergen [2 ]
Jones, Graham [3 ]
Preston, Sharon [4 ]
Szekeres, Thomas [5 ]
Imdahl, Roland [6 ]
Conti, Marc [7 ]
Blanckaert, Norbert [8 ]
Jose, Darren [9 ]
Thiery, Joachim [10 ]
Feldmann, Lisa [11 ]
von Ahsen, Nicolas [12 ]
Locatelli, Massimo [13 ]
Kremastinou, Jenni [14 ]
Kunst, Albert [15 ]
Hubbuch, Arnulf [15 ]
McGovern, Margaret [15 ]
机构
[1] Univ Spital Zurich, Inst Klin Chem, Zurich, Switzerland
[2] Labor Dr Limbach & Kollegen, Heidelberg, Germany
[3] St Vincents Publ Hosp, Sydney, NSW, Australia
[4] Huntsville Dist Mem Hosp, Huntsville, AL USA
[5] AKH Wien, Klin Inst Med & Chem Labordiagnost, Vienna, Austria
[6] Labor Schottdorf MVZ GmbH, Augsburg, Germany
[7] CHU Bicetre, Le Kremlin Bicetre, France
[8] Univ Hosp KU, Louvain, Belgium
[9] NHS Fdn Trust, Poole Hosp, Poole, Dorset, England
[10] Univ Klinikum Leipzig, Leipzig, Germany
[11] Mid Amer Clin Labs, Indianapolis, IN USA
[12] Univ Gottingen, Univ Med, D-37073 Gottingen, Germany
[13] Laboraf Diagnost & Ric San Raffaele SpA, Milan, Italy
[14] Biomed Bioiatriki, Athens, Greece
[15] Roche Diagnost GmbH, Mannheim, Germany
来源
JALA | 2013年 / 18卷 / 04期
关键词
analyzers; analytical modules; analytical performance; practicability assessment; workflow analysis; PERFORMANCE EVALUATION; CLINICAL-CHEMISTRY; ANALYTICS;
D O I
10.1177/2211068212472183
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Clinical laboratories need to test patient samples precisely, accurately, and efficiently. The latest member of the Roche cobas modular platform family, the cobas 8000 modular analyzer series allows compact and convenient consolidation of clinical chemistry and immunochemistry assays in high-workload laboratories with a throughput of 3 to 15 million tests annually. Here we present the results of studies designed to test the overall system performance under routine-like conditions that were conducted at 14 laboratories over 2 y. Experiments that test analytical performance of the new module were integrated with overall system functionality testing of all modules in different configurations. More than two million results were generated and evaluated for similar to 100 applications using serum/plasma, urine, or EDTA blood samples. During the workflow studies, eight configurations of the possible 38 combinations were used, covering all available analytical modules. The versatility of the module combinations makes the system customizable to fit the needs of diverse laboratories, allowing precise and accurate analysis of a broad spectrum of clinical chemistry and immunochemistry parameters with short turnaround times. This new system will contribute to the ability of clinical laboratories to offer better service to their customers and support vital clinical decision making.
引用
收藏
页码:306 / 327
页数:22
相关论文
共 11 条
[1]  
Bablok W., 1995, QUIMICA CLIN S, V14, P239
[2]   Multicentre performance evaluation of the E170 module for MODULAR ANALYTICS [J].
Bieglmayer, C ;
Chan, DW ;
Sokoll, L ;
Imdahl, R ;
Kobayashi, M ;
Yamada, E ;
Lilje, DJ ;
Luthe, H ;
Meissner, J ;
Messeri, G ;
Celli, A ;
Tozzi, P ;
Roth, HJ ;
Schmidt, FP ;
Mächler, ML ;
Schuff-Werner, P ;
Zingler, C ;
Smitz, J ;
Schiettecatte, J ;
Vonderschmitt, DJ ;
Pei, P ;
Ng, K ;
Ebert, C ;
Kirch, P ;
Wanger, M ;
McGovern, M ;
Stockmann, W ;
Kunst, A .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2004, 42 (10) :1186-1202
[3]   REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[4]   MODULAR ANALYTICS:: A new approach to automation in the clinical laboratory [J].
Horowitz, GL ;
Zaman, Z ;
Blanckaert, NJC ;
Chan, DW ;
DuBois, JA ;
Golaz, O ;
Mensi, N ;
Keller, F ;
Stolz, H ;
Klingler, K ;
Marocchi, A ;
Prencipe, L ;
McLawhon, RW ;
Nilsen, OL ;
Oellerich, M ;
Luthe, H ;
Orsonneau, JL ;
Richeux, G ;
Recio, F ;
Roldan, E ;
Rymo, L ;
Wicktorsson, AC ;
Welch, SL ;
Wieland, H ;
Grawitz, AB ;
Mitsumaki, H ;
McGovern, M ;
Ng, K ;
Stockmann, W .
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY, 2005, (01) :8-25
[5]  
Kunst A., 2005, CLIN CHIM ACTA, p355S
[6]  
Mocarelli P., 2008, J AUTOM METHOD MANAG, V2008, P1
[7]  
PASSING H, 1983, J CLIN CHEM CLIN BIO, V21, P709
[8]  
Peake Michael, 2006, Clin Biochem Rev, V27, P173
[9]  
STOCKMANN W, 1993, EVALUATION METHODS L, P185
[10]   An innovative approach to functionality testing of analysers in the clinical laboratory [J].
Stockmann, Wolfgang ;
Engeldinger, Werner ;
Kunst, Albert ;
McGovern, Margaret .
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY, 2008,