Medina embolization device for the treatment of intracranial aneurysms: 18 months' angiographic results

被引:17
作者
Haffaf, Idriss [1 ]
Clarencon, Frederic [1 ,2 ]
Shotar, Eimad [1 ,2 ]
Rolla-Bigliani, Claudia [1 ]
Perre, Saskia Vande [1 ]
Mathon, Bertrand [2 ,3 ]
Drir, Mehdi [4 ]
Sourour, Nader-Antoine [1 ]
机构
[1] Hop La Pitie Salpetriere, Dept Neuroradiol, F-75013 Paris, France
[2] Sorbonne Univ, Paris, France
[3] Hop La Pitie Salpetriere, Dept Neurosurg, Paris, France
[4] Hop La Pitie Salpetriere, Dept Anesthesiol, Paris, France
关键词
ENDOVASCULAR TREATMENT; EXPERIENCE; SAFETY;
D O I
10.1136/neurintsurg-2018-014110
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Background and purpose The Medina embolization device (MED) is a new flow disruption device combining the design of a detachable coil with an intrasaccular flow disrupter. Safety and short-term angiographic effectiveness of this device have recently been reported. However, long-term angiographic results are lacking. We report herein the 18 months' angiographic outcome in patients treated for a wide-neck intracranial aneurysm with the MED. Materials and methods Nineteen patients (17 female, mean age 50 years) with 20 wide-neck intracranial aneurysms (six ruptured; 14 unruptured) were treated by the MED between January 2015 and June 2016. Procedure-related complications were systematically recorded; discharge and 6-9 months' follow-up modified Rankin Scale scores were assessed. Angiographic mid-term and long-term follow-up were performed with a mean delay of 6.4 +/- 1.5 months (n=16 aneurysms) and 17.7 +/- 4.2 months (n=15 aneurysms), respectively. Occlusion rates were evaluated after the procedure and at the mid-term and long-term follow-up using the Roy-Raymond scale. Results Embolization with the MED was feasible in all except two cases (2/20, 10%). One per-procedural perforation was recorded (1/20, 5%) and one MED deployment failed because of the aneurysm's shape (1/20, 5%). Three cases of thromboembolic complications were observed (3/20, 15%). Only one thromboembolic complication was responsible for clinical sequelae. Grade A occlusion rate was 61% (11/18) after the procedure, 75% at 6 months' follow-up (12/16), and 80% (12/15) at long-term follow-up. Two cases (2/18, 11%) of recanalization at mid-term were documented angiographically. No recanalization occurred between the mid-term and long-term follow-up. Conclusion MED is a hybrid embolization device, combining properties of a conventional coil with those of an intrasaccular flow disrupter. Our series focusing on long-term angiographic follow-up shows a satisfactory long-term occlusion rate. Larger series with longer angiographic follow-up times are warranted to confirm these preliminary results.
引用
收藏
页码:516 / 522
页数:7
相关论文
共 15 条
[1]   The Medina Embolic Device: Karolinska experience [J].
Bhogal, P. ;
Brouwer, P. A. ;
Yeo, L. ;
Svensson, M. ;
Soderman, M. .
INTERVENTIONAL NEURORADIOLOGY, 2018, 24 (01) :4-13
[2]   Single-center experience in the endovascular treatment of wide-necked intracranial aneurysms with a bridging intra-/extra-aneurysm implant (pCONus) [J].
Fischer, S. ;
Weber, A. ;
Titschert, A. ;
Brenke, C. ;
Kowoll, A. ;
Weber, W. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (11) :1186-1191
[3]   Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coverage [J].
Froelich, Andreas Maximilian ;
Nawka, Marie Teresa ;
Ernst, Marielle ;
Frischmuth, Isabell ;
Fiehler, Jens ;
Buhk, Jan-Hendrik .
PLOS ONE, 2018, 13 (02)
[4]   Clinical and angiographic outcomes after stent-assisted coiling of cerebral aneurysms with Enterprise and Neuroform stents: a comparative analysis of the literature [J].
King, Brent ;
Vaziri, Sasha ;
Singla, Amit ;
Fargen, Kyle M. ;
Mocco, J. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (12) :905-909
[5]  
Koebbe Christopher J, 2006, Neurosurgery, V59, pS93
[6]  
Moret J, 1997, Interv Neuroradiol, V3, P21
[7]   The Medina Embolic Device: early clinical experience from a single center [J].
Perez, Marta Aguilar ;
Bhogal, Pervinder ;
Moreno, Rosa Martinez ;
Baezner, Hansjorg ;
Ganslandt, Oliver ;
Henkes, Hans .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (01) :77-+
[8]   Safety and Efficacy of Balloon Remodeling Technique during Endovascular Treatment of Intracranial Aneurysms: Critical Review of the Literature [J].
Pierot, L. ;
Cognard, C. ;
Spelle, L. ;
Moret, J. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (01) :12-15
[9]   Remodeling Technique for Endovascular Treatment of Ruptured Intracranial Aneurysms Had a Higher Rate of Adequate Postoperative Occlusion than Did Conventional Coil Embolization with Comparable Safety [J].
Pierot, Laurent ;
Cognard, Christophe ;
Anxionnat, Rene ;
Ricolfi, Frederic .
RADIOLOGY, 2011, 258 (02) :546-553
[10]   Stent-Assisted Coiling of Intracranial Aneurysms Clinical and Angiographic Results in 216 Consecutive Aneurysms [J].
Piotin, Michel ;
Blanc, Raphael ;
Spelle, Laurent ;
Mounayer, Charbel ;
Piantino, Rhelen ;
Schmidt, Paul J. ;
Moret, Jacques .
STROKE, 2010, 41 (01) :110-115