Electrostatic axially harmonic orbital trapping: A high-performance technique of mass analysis

被引:545
作者
Makarov, A [1 ]
机构
[1] HD Technol Ltd, Manchester M22 5PP, Lancs, England
关键词
D O I
10.1021/ac991131p
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work describes a new type of mass analyzer which employs trapping in an electrostatic field. The potential distribution of the held can be represented as a combination of quadrupole and logarithmic potentials. In the absence of any magnetic or rf fields, ion stability is achieved only due to ions orbiting around an axial electrode, Orbiting ions also perform harmonic oscillations along the electrode with frequency proportional to (m/z)(-1/2). These oscillations are detected using image current detection and are transformed into mass spectra using fast FT, similarly to FT ICR. Practical aspects of the trap design are presented. High-mass resolution up to 150 000 for ions produced by laser ablation has been demonstrated, along with high-energy acceptance and wide mass range.
引用
收藏
页码:1156 / 1162
页数:7
相关论文
共 27 条
  • [1] ANDERSON GA, 1995, P 43 ANN C MASS SPEC, P1093
  • [2] [Anonymous], 1965, PROBLEMS ASYMPTOTIC
  • [3] A gated electrostatic ion trap to repetitiously measure the charge and m/z of large electrospray ions
    Benner, WH
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (20) : 4162 - 4168
  • [4] Bryant P.J., 1993, The Principles of Circular Accelerators and Storage Rings
  • [5] ULTRAHIGH-RESOLUTION MATRIX-ASSISTED LASER-DESORPTION IONIZATION OF SMALL PROTEINS BY FOURIER-TRANSFORM MASS-SPECTROMETRY
    CASTORO, JA
    WILKINS, CL
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (19) : 2621 - 2627
  • [6] New type of electrostatic ion trap for storage of fast ion beams
    Dahan, M
    Fishman, R
    Heber, O
    Rappaport, M
    Altstein, N
    Zajfman, D
    van der Zande, WJ
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) : 76 - 83
  • [7] Dienes T, 1996, MASS SPECTROM REV, V15, P163, DOI 10.1002/(SICI)1098-2787(1996)15:3<163::AID-MAS2>3.0.CO
  • [8] 2-G
  • [9] Routine part-per-million mass accuracy for high-mass ions: Space-charge effects in MALDI FT-ICR
    Easterling, ML
    Mize, TH
    Amster, IJ
    [J]. ANALYTICAL CHEMISTRY, 1999, 71 (03) : 624 - 632
  • [10] Gall L.N., 1986, Soviet Union Pat., Patent No. 1247973