Holmium:Yttrium-Aluminum-Garnet Laser Pulse Type Affects Irrigation Temperatures in a Benchtop Ureteral Model

被引:28
作者
Winship, Brenton [1 ]
Terry, Russell [1 ]
Boydston, Kohldon [1 ]
Carlos, Evan [1 ]
Wollin, Daniel [2 ]
Peters, Chloe [3 ]
Li, Jingqiu [4 ]
Preminger, Glenn [1 ]
Lipkin, Michael [1 ]
机构
[1] Duke Univ, Med Ctr, Div Urol Surg, 1573 White Zone,Box 3167, Durham, NC 27710 USA
[2] Brigham & Womens Hosp, Div Urol, 75 Francis St, Boston, MA 02115 USA
[3] Duke Univ, Sch Med, Durham, NC USA
[4] Duke NUS Med Sch, Singapore, Singapore
关键词
laser; ureteroscopy; urolithiasis; thermal injury; LITHOTRIPSY; FRAGMENTATION; RETROPULSION; SETTINGS;
D O I
10.1089/end.2019.0496
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction: MOSES (TM) technology is a holmium:yttrium-aluminum-garnet laser pulse mode shown to minimize stone retropulsion. This may facilitate lithotripsy at higher power settings. However, power and heat production are proportional, and temperatures capable of tissue injury may occur during ureteroscopic lithotripsy. Although previous in vitro studies demonstrate the importance of irrigation and activation time on heat production, the impact of pulse type has not been evaluated. Methods: A flexible ureteroscope with a 365 mu m laser fiber was placed in an 11/13 F access sheath inserted into a 50 mL saline bag to simulate a ureter, renal pelvis, and antegrade irrigant flow. A thermocouple was placed adjacent to the laser tip, and the laser fired for 30 seconds at 0.6 J/6 Hz, 0.8 J/8 Hz, 1 J/10 Hz, 1 J/20 Hz, and 0.2 J/70 Hz at irrigation pressure of 100 mmHg. Four runs were tested per setting using short pulse, long pulse (LP), MOSES-contact (MC), and MOSES-distance (MD) modes. The mean temperature changes (dT) were compared and thermal dose was calculated in cumulative equivalent minutes at 43 degrees C (CEM43) using an adjusted baseline of 37 degrees C. CEM43 >= 120 minutes was considered the tissue injury threshold. Results: At 0.8 J/8 Hz, LP produced the greatest dT, significantly higher than MC (p = 0.041). CEM43 did not exceed the injury threshold. At 1 J/10 Hz, dT was significantly higher for LP versus MC and MD (p = 0.024 and 0.045, respectively). Thermal dose remained below the injury threshold. No differences in dT were seen between pulse types at 0.6 J/6 Hz, 0.2 J/70 Hz, or 1 J/20 Hz. At 1 J/20 Hz, thermal dose exceeded the injury threshold for all pulse types within 3 seconds. Conclusions: Laser pulse type and length seemed to impact heat production in our ureteral model. LP produced significantly greater temperatures at 0.8 J/8 Hz and 1 J/10 Hz relative to MOSES settings. Fortunately, thermal dose remained safe at these settings. Both LP and MOSES technology have been shown to reduce stone retropulsion. At power <= 10 W, the latter may confer this advantage with decreased heat production.
引用
收藏
页码:896 / 901
页数:6
相关论文
共 18 条
[1]   Thermal Response to High-Power Holmium Laser Lithotripsy [J].
Aldoukhi, Ali H. ;
Ghani, Khurshid R. ;
Hall, Timothy L. ;
Roberts, William W. .
JOURNAL OF ENDOUROLOGY, 2017, 31 (12) :1308-1312
[2]   Caliceal Fluid Temperature During High-Power Holmium Laser Lithotripsy in an In Vivo Porcine Model [J].
Aldoukhi, Ali H. ;
Hall, Timothy L. ;
Ghani, Khurshid R. ;
Maxwell, Adam D. ;
MacConaghy, Brian ;
Roberts, William W. .
JOURNAL OF ENDOUROLOGY, 2018, 32 (08) :724-729
[3]   In Vitro Comparison of Stone Fragmentation When Using Various Settings with Modern Variable Pulse Holmium Lasers [J].
Bell, John Roger ;
Penniston, Kristina L. ;
Nakada, Stephen Y. .
JOURNAL OF ENDOUROLOGY, 2017, 31 (10) :1067-1072
[4]   In Vitro Comparison of Holmium Lasers: Evidence for Shorter Fragmentation Time and Decreased Retropulsion Using a Modern Variable-pulse Laser [J].
Bell, John Roger ;
Penniston, Kristina L. ;
Nakada, Stephen Y. .
UROLOGY, 2017, 107 :37-42
[5]   Temperature Changes Inside the Kidney: What Happens During Holmium:Yttrium-Aluminium-Garnet Laser Usage? [J].
Buttice, Salvatore ;
Sener, Tarik Emre ;
Proietti, Silvia ;
Dragos, Laurian ;
Tefik, Tzevat ;
Doizi, Steeve ;
Traxer, Olivier .
JOURNAL OF ENDOUROLOGY, 2016, 30 (05) :574-579
[6]   Use of the Moses Technology to Improve Holmium Laser Lithotripsy Outcomes: A Preclinical Study [J].
Elhilali, Mostafa M. ;
Badaan, Shadie ;
Ibrahim, Ahmed ;
Andonian, Sero .
JOURNAL OF ENDOUROLOGY, 2017, 31 (06) :598-604
[7]   High power holmium:YAG versus thulium fiber laser treatment of kidney stones in dusting mode: ablation rate and fragment size studies [J].
Hardy, Luke A. ;
Vinnichenko, Viktoriya ;
Fried, Nathaniel M. .
LASERS IN SURGERY AND MEDICINE, 2019, 51 (06) :522-530
[8]   Thermal effects of Ho: YAG laser lithotripsy: real-time evaluation in an in vitro model [J].
Hein, Simon ;
Petzold, Ralf ;
Schoenthaler, Martin ;
Wetterauer, Ulrich ;
Miernik, Arkadiusz .
WORLD JOURNAL OF UROLOGY, 2018, 36 (09) :1469-1475
[9]  
Lin X, 1999, LASER PULSE HEATING, P8120
[10]   Pathophysiological aspects of ureterorenoscopic management of upper urinary tract calculi [J].
Osther, Palle J. S. ;
Pedersen, Katja V. ;
Lildal, Soren K. ;
Pless, Maria S. ;
Andreassen, Kim H. ;
Osther, Susanne S. ;
Jung, Helene U. .
CURRENT OPINION IN UROLOGY, 2016, 26 (01) :63-69