Characterization and use of EBT radiochromic film for IMRT dose verification

被引:136
作者
Zeidan, Omar A.
Stephenson, Stacy Ann L.
Meeks, Sanford L.
Wagner, Thomas H.
Willoughby, Twyla R.
Kupelian, Patrick A.
Langen, Katja M.
机构
[1] MD Anderson Canc Ctr, Dept Radiat Phys, Orlando, FL 32806 USA
[2] Wesleyan Coll, Dept Chem & Phys, Macon, GA 31210 USA
关键词
radiochromic film dosimetry; GAFChromic film; TomoTherapy; diode array; IMRT verification;
D O I
10.1118/1.2360012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present an evaluation of a new and improved radiochromic film, type EBT, for its implementation to IMRT dose verification. Using a characterized flat bed color CCD scanner, the film's dose sensitivity, uniformity, and speed of development post exposure were shown to be superior to previous types of radiochromic films. The film's dose response was found to be very similar to ion chamber scans in water through comparisons of depth dose and lateral dose profiles. The effect of EBT film polarization with delivered dose and film scan orientation was shown to have a significant effect on the scanner's OD readout. In addition, the film's large size, flexibility, and the ability to submerge it in water for relatively short periods of time allowed for its use in both water and solid water phantoms to verify TomoTherapy IMRT dose distributions in flat and curved dose planes. Dose verification in 2D was performed on ten IMRT plans (five head and neck and five prostate) by comparing measured EBT dose distributions to TomoTherapy treatment planning system calculated dose. The quality of agreement was quantified by the gamma index for four sets of dose difference and distance to agreement criteria. Based on this study, we show that EBT film has several favorable features that allow for its use in routine IMRT patient-specific QA. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:4064 / 4072
页数:9
相关论文
共 26 条
[1]   Effects of read-out light sources and ambient light on radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2407-2412
[2]   Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[3]   Radiochromic film dosimetry in water phantoms [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01) :N27-N31
[4]   Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions [J].
Dempsey, JF ;
Low, DA ;
Mutic, S ;
Markman, J ;
Kirov, AS ;
Nussbaum, GH ;
Williamson, JF .
MEDICAL PHYSICS, 2000, 27 (10) :2462-2475
[5]   Dosimetric properties of improved GafChromic films for seven different digitizers [J].
Devic, S ;
Seuntjens, J ;
Hegyi, G ;
Podgorsak, EB ;
Soares, CG ;
Kirov, AS ;
Ali, I ;
Williamson, JF ;
Elizondo, A .
MEDICAL PHYSICS, 2004, 31 (09) :2392-2401
[6]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[7]  
*ICRU, 1987, 42 ICRU
[8]  
*ISP, GAFCHROMIC EBT WHIT
[9]   GafChromic MD-55: Investigated as a precision dosimeter [J].
Klassen, NV ;
van der Zwan, L ;
Cygler, J .
MEDICAL PHYSICS, 1997, 24 (12) :1924-1934
[10]   Evaluation of a diode array for QA measurements on a helical tomotherapy unit [J].
Langen, KM ;
Meeks, SL ;
Poole, DO ;
Wagner, TH ;
Willoughby, TR ;
Zeidan, OA ;
Kupelian, PA ;
Ruchala, KJ ;
Olivera, GH .
MEDICAL PHYSICS, 2005, 32 (11) :3424-3430