An automated cell analysis sensing system based on a microfabricated rheoscope for the study of red blood cells physiology

被引:18
作者
Bransky, Avishay [1 ]
Korin, Natanel
Nemirovski, Yael
Dinnar, Uri
机构
[1] Technion Israel Inst Technol, Biomed Dept, Haifa, Israel
[2] Technion Israel Inst Technol, Dept Elect Engn, Haifa, Israel
关键词
rheoscope; ektacytometer; erythrocytes; RBC; microchannel;
D O I
10.1016/j.bios.2005.12.006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An automated rheoscope has been developed, utilizing a microfabricated glass flow cell, high speed camera and advanced image-processing software. RBCs suspended in a high viscosity medium were filmed flowing through a microchannel. Under these conditions, RBCs exhibit different orientations and deformations according to their location in the velocity profile. The rheoscope system produces valuable data such as velocity profile of RBCs, spatial distribution within a microchannel and deformation index (DI) curves. The variation of DI across the channel height, due to change in shear stress, was measured carrying implications for diffractometry methods. These curves of DI were taken at a constant flow rate and cover most of the relevant shear stress spectrum. This is an improvement of the existing techniques for deformability measurements and may serve as a diagnostic tool for certain blood disorders. The DI curves were compared to measurements of the flowing RBCs velocity profile. In addition, we found that RBCs flowing in a microchannel are mostly gathered in the center of the flow and maintain a characteristic spatial distribution. The spatial distribution in this region changes slightly with increasing flow rate. Hence, the system described, provides means for examining the behavior of individual RBCs, and may serve as a microfabricated diagnostic device for deformability measurement. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:165 / 169
页数:5
相关论文
共 26 条
[1]  
ALTENDORF EA, 1978, P SOC PHOTO-OPT INS, V2978, P136
[2]  
Baskurt OK, 2004, CLIN HEMORHEOL MICRO, V31, P23
[3]  
BESSIS M, 1975, BLOOD CELLS, V1, P307
[4]  
BESSIS M, 1980, BLOOD CELLS, V6, P315
[5]   Biotechnology at low Reynolds numbers [J].
Brody, JP ;
Yager, P ;
Goldstein, RE ;
Austin, RH .
BIOPHYSICAL JOURNAL, 1996, 71 (06) :3430-3441
[6]   Measurement of the distribution of red blood cell deformability using an automated rheoscope [J].
Dobbe, JGG ;
Streekstra, GJ ;
Hardeman, MR ;
Ince, C ;
Grimbergen, CA .
CYTOMETRY, 2002, 50 (06) :313-325
[7]  
Dobbe JGG, 2004, BIORHEOLOGY, V41, P65
[8]   Analyzing red blood cell-deformability distributions [J].
Dobbe, JGG ;
Hardeman, MR ;
Streekstra, GJ ;
Strackee, J ;
Ince, C ;
Grimbergen, CA .
BLOOD CELLS MOLECULES AND DISEASES, 2002, 28 (03) :373-384
[9]  
Engstrom K. G., 1995, Biorheology, V32, P115, DOI 10.1016/0006-355X(95)91960-B
[10]  
FISCHER T, 1977, BLOOD CELLS, V3, P351