Prospective comparative study of the EUS guided 25-gauge FNA needle with the 19-gauge Trucut needle and 22-gauge FNA needle in patients with solid pancreatic masses

被引:178
作者
Sakamoto, Hiroki [1 ]
Kitano, Masayuki [1 ]
Komaki, Takamitsu [1 ]
Noda, Kazu [1 ]
Chikugo, Takaaki [2 ]
Dote, Kensaku [2 ]
Takeyama, Yoshifumi [3 ]
Das, Kunal [4 ]
Yamao, Kenji [5 ]
Kudo, Masatoshi [1 ]
机构
[1] Kinki Univ, Sch Med, Div Gastroenterol & Hepatol, Dept Internal Med, Osaka 5898511, Japan
[2] Kinki Univ, Sch Med, Dept Pathol, Osaka 5898511, Japan
[3] Kinki Univ, Sch Med, Dept Surg, Osaka 5898511, Japan
[4] Sir Ganga Ram Hosp New Delhi, Dept Gastroenterol, New Delhi, India
[5] Aichi Canc Ctr Hosp, Dept Gastroenterol, Aichi, Japan
关键词
EUS-FNA; 25-gauge FNA needle; 22-gauge FNA needle; 19-gauge Trucut needle; Pancreatic masses; ENDOSCOPIC ULTRASOUND; ASPIRATION BIOPSY; DIAGNOSIS; ULTRASONOGRAPHY; COMPLICATIONS; ACCURACY; NUMBER; PASSES; TUMORS;
D O I
10.1111/j.1440-1746.2008.05636.x
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
The aim of this prospective study was to compare fine-needle aspiration guided by endoscopic ultrasonography (EUS-FNA) using 25-gauge and 22-gauge needles with the EUS-guided 19-gauge Trucut needle biopsy (EUS-TNB) in patients with solid pancreatic mass. Twenty-four consecutive patients with pancreatic mass underwent biopsies by both EUS-FNA and EUS-TNB. Three needles were compared with respect to technical success rate, tissue size obtained, overall diagnostic accuracy and accuracy for histological and cytological diagnosis. The 25-gauge EUS-FNA was technically easier and obtained superior overall diagnostic accuracy than the 22-gauge and Trucut needles, especially in lesions of the pancreas head and uncinate process. Overall accuracy for the 25-gauge, 22-gauge and Trucut needle was 91.7%, 79.7% and 54.1%, respectively. Accuracy for cytological diagnosis irrespective the site of lesions with 25-gauge, 22-gauge and Trucut needles was 91.7%, 75.0%, and 45.8%, respectively. For uncinate masses, it was 100%, 33.3%, and 0.0%, respectively. These differences were significant. Among technically successful patients, the accuracy for histological diagnosis using the 25-gauge was significantly inferior (P < 0.05) to 22-gauge and Trucut needles and the rates were 45.8%, 78.9% and 83.3%. The 25-gauge FNA needle was significantly superior in terms of technical success rate and overall diagnostic accuracy, especially for the head and uncinate lesions, compared to the 22-gauge and Trucut needles and could be considered 'the best choice needle for cytological diagnosis' of solid pancreatic lesions. If histological diagnosis is required, the 22-gauge FNA needle and Trucut needle may be advantageous for use in head/uncinate and body/tail lesions, respectively.
引用
收藏
页码:384 / 390
页数:7
相关论文
共 27 条
[1]  
Bentz JS, 1998, DIAGN CYTOPATHOL, V18, P98, DOI 10.1002/(SICI)1097-0339(199802)18:2<98::AID-DC4>3.0.CO
[2]  
2-P
[3]   Endoscopic ultrasound guided fine needle aspiration of malignant pancreatic lesions [J].
Bhutani, MS ;
Hawes, RH ;
Baron, PL ;
Sanders-Cliette, A ;
van Velse, A ;
Osborne, JF ;
Hoffman, BJ .
ENDOSCOPY, 1997, 29 (09) :854-858
[4]   Cytologic diagnosis of pancreatic endocrine tumors by endoscopic ultrasound-guided fine-needle aspiration: A review [J].
Chang, Fuju ;
Chandra, Ashish ;
Culora, Giuseppe ;
Mahadeva, Ula ;
Meenan, John ;
Herbert, Amanda .
DIAGNOSTIC CYTOPATHOLOGY, 2006, 34 (09) :649-658
[5]   The clinical utility of endoscopic ultrasound-guided fine-needle aspiration in the diagnosis and staging of pancreatic carcinoma [J].
Chang, KJ ;
Nguyen, P ;
Erickson, RA ;
Durbin, TE ;
Katz, KD .
GASTROINTESTINAL ENDOSCOPY, 1997, 45 (05) :387-393
[6]   Factors predicting the number of EUS-guided fine-needle passes for diagnosis of pancreatic malignancies [J].
Erickson, RA ;
Sayage-Rabie, L ;
Beissner, RS .
GASTROINTESTINAL ENDOSCOPY, 2000, 51 (02) :184-190
[7]   FINE-NEEDLE ASPIRATION CYTOLOGY GUIDED BY ENDOSCOPIC ULTRASONOGRAPHY - RESULTS IN 141 PATIENTS [J].
GIOVANNINI, M ;
SEITZ, JF ;
MONGES, G ;
PERRIER, H ;
RABBIA, I .
ENDOSCOPY, 1995, 27 (02) :171-177
[8]   Endoscopic ultrasound-guided fine-needle aspiration biopsy using linear array and radial scanning endosonography [J].
Gress, FG ;
Hawes, RH ;
Savides, TJ ;
Ikenberry, SO ;
Lehman, GA .
GASTROINTESTINAL ENDOSCOPY, 1997, 45 (03) :243-250
[9]   Puncture of solid pancreatic tumors guided by endoscopic ultrasonography: A pilot study series comparing Trucut and 19-gauge and 22-gauge aspiration needles [J].
Itoi, T ;
Itokawa, F ;
Sofuni, A ;
Nakamura, K ;
Tsuchida, A ;
Yamao, K ;
Kawai, T ;
Moriyasu, F .
ENDOSCOPY, 2005, 37 (04) :362-366
[10]  
Johnson DE, 1997, AM SURGEON, V63, P675