3-D potential distribution;
3-D power loss density distribution;
bipolar transurethral resection;
electrosurgery;
multielectrode measuring system;
TRANSURETHRAL RESECTION;
PROSTATE;
BIPOLAR;
D O I:
10.1109/TIM.2014.2310112
中图分类号:
TM [电工技术];
TN [电子技术、通信技术];
学科分类号:
0808 ;
0809 ;
摘要:
To analyze the electric field around bipolar resectoscopes, used in urology, in terms of reasons for late complications after a surgical treatment a flexible multielectrode system was developed to measure the 3-D potential distribution. A high spatial resolution is achieved with the least possible individual measurements under the conditions of a quasi-static electric field. A flexible arrangement and positioning of the measuring points in the vertical direction of the experimental environment enable an adjustable spatial resolution and the selection of the region of interest. The existing influence of the multielectrode system on the measuring results is described and a correction method is presented to achieve significant results. Thus, the multielectrode system is usable for a comparative study of bipolar resectoscopes varying in the arrangement of resection and return electrode.