Variation of kQclin, Qmsrfclin, fmsr for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio

被引:72
作者
Francescon, Paolo [1 ]
Beddar, Sam [2 ]
Satariano, Ninfa [1 ]
Das, Indra J. [3 ]
机构
[1] Osped Vicenza, Dept Radiat Oncol, I-36100 Vicenza, Italy
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77005 USA
[3] Indiana Univ, Sch Med, Dept Radiat Oncol, Indianapolis, IN 46202 USA
关键词
small beams; correction factor; output factor; Monte Carlo; detector; PLASTIC SCINTILLATION DETECTORS; MONTE-CARLO SIMULATIONS; THIMBLE IONIZATION CHAMBERS; QUALITY CORRECTION FACTORS; OUTPUT FACTOR MEASUREMENT; ENERGY BEAM DOSIMETRY; PHOTON BEAMS; CYBERKNIFE SYSTEM; ION CHAMBERS; RADIOSURGERY;
D O I
10.1118/1.4895978
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Evaluate the ability of different dosimeters to correctly measure the dosimetric parameters percentage depth dose (PDD), tissue-maximum ratio (TMR), and off-axis ratio (OAR) in water for small fields. Methods: Monte Carlo (MC) simulations were used to estimate the variation of k(Qclin, Qmsr)(fclin, fmsr) for several types of microdetectors as a function of depth and distance from the central axis for PDD, TMR, and OAR measurements. The variation of k(Qclin, Qmsr)(fclin, fmsr) enables one to evaluate the ability of a detector to reproduce the PDD, TMR, and OAR in water and consequently determine whether it is necessary to apply correction factors. The correctness of the simulations was verified by assessing the ratios between the PDDs and OARs of 5- and 25-mm circular collimators used with a linear accelerator measured with two different types of dosimeters (the PTW 60012 diode and PTW PinPoint 31014 microchamber) and the PDDs and the OARs measured with the Exradin W1 plastic scintillator detector (PSD) and comparing those ratios with the corresponding ratios predicted by the MC simulations. Results: MC simulations reproduced results with acceptable accuracy compared to the experimental results; therefore, MC simulations can be used to successfully predict the behavior of different dosimeters in small fields. The Exradin W1 PSD was the only dosimeter that reproduced the PDDs, TMRs, and OARs in water with high accuracy. With the exception of the EDGE diode, the stereotactic diodes reproduced the PDDs and the TMRs in water with a systematic error of less than 2% at depths of up to 25 cm; however, they produced OAR values that were significantly different from those in water, especially in the tail region (lower than 20% in some cases). The microchambers could be used for PDD measurements for fields greater than those produced using a 10-mm collimator. However, with the detector stem parallel to the beam axis, the microchambers could be used for TMR measurements for all field sizes. The microchambers could not be used for OAR measurements for small fields. Conclusions: Compared with MC simulation, the Exradin W1 PSD can reproduce the PDDs, TMRs, and OARs in water with a high degree of accuracy; thus, the correction used for converting dose is very close to unity. The stereotactic diode is a viable alternative because it shows an acceptable systematic error in the measurement of PDDs and TMRs and a significant underestimation in only the tail region of the OAR measurements, where the dose is low and differences in dose may not be therapeutically meaningful. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:14
相关论文
共 42 条
  • [11] An EGSnrc Monte Carlo study of the microionization chamber for reference dosimetry of narrow irregular IMRT beamlets
    Capote, R
    Sánchez-Doblaclo, F
    Leal, A
    Lagares, JI
    Arráns, R
    Hartmann, GH
    [J]. MEDICAL PHYSICS, 2004, 31 (09) : 2416 - 2422
  • [12] Experimental small field 6 MV output ratio analysis for various diode detector and accelerator combinations
    Cranmer-Sargison, Gavin
    Weston, Steve
    Sidhu, Narinder P.
    Thwaites, David I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) : 429 - 435
  • [13] The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambers
    Crop, F.
    Reynaert, N.
    Pittomvils, G.
    Paelinck, L.
    De Wagter, C.
    Vakaet, L.
    Thierens, H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) : 2951 - 2969
  • [14] Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields
    Czarnecki, D.
    Zink, K.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08) : 2431 - 2444
  • [15] Das I.J., 2000, J RADIOSURGERY, V3, P177, DOI DOI 10.1023/A:1009594509115
  • [16] Das I. J., 2014, MED PHYS IN PRESS
  • [17] Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM
    Das, Indra J.
    Cheng, Chee-Wai
    Watts, Ronald J.
    Ahnesjo, Anders
    Gibbons, John
    Li, X. Allen
    Lowenstein, Jessica
    Mitra, Raj K.
    Simon, William E.
    Zhu, Timothy C.
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 4186 - 4215
  • [18] Small fields:: Nonequilibrium radiation dosimetry
    Das, Indra J.
    Ding, George X.
    Ahnesjo, Anders
    [J]. MEDICAL PHYSICS, 2008, 35 (01) : 206 - 215
  • [19] Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams
    De Vlamynck, K
    Palmans, H
    Verhaegen, F
    De Wagter, C
    De Neve, W
    Thierens, H
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1874 - 1882
  • [20] Monte Carlo simulated correction factors for output factor measurement with the CyberKnife system-results for new detectors and correction factor dependence on measurement distance and detector orientation
    Francescon, P.
    Kilby, W.
    Satariano, N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (06) : N11 - N17