Load Transfer Analysis at the Interface Between a Steel Post and Polyethylene Matrix Using Pull-Out Test: Experimental and Theoretical Parametric Study

被引:2
作者
Abid, M. [1 ,2 ]
Kharrat, M. [1 ,3 ]
Dammak, M. [1 ,3 ]
Maalej, A. [1 ]
机构
[1] Ecole Natl Ingn Sfax, Lab Syst Electromecan, Sfax 3038, Tunisia
[2] Inst Super Etudes Technol Sfax, El Bustan 3099, Sfax, Tunisia
[3] Inst Preparatoire Etudes Ingn Sfax, Dept Technol, Sfax 3018, Tunisia
关键词
Total Knee Arthroplasty (TKA); fixation; post; interface; load transfer; interference fit; pull-out; High Density Polyethylene (HDPE); Coulomb friction; TOTAL KNEE ARTHROPLASTY; FINITE-ELEMENT; FIXATION; COMPONENTS; IMPLANTS; DESIGNS; SCREWS;
D O I
10.1163/156856108X389433
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
In various kinds of orthopedic or dental prostheses, adequate fixation of prosthetic components to the host bone represents a great challenge. The short- and long-term fixation stabilities of the implants are often provided by the use of posts. In this work, a cylindrical steel post inserted in a predrilled High Density Polyethylene (HDPE) cylinder with an initial interference fit was used as a model to analyze the contribution of post fixation to the initial stability of Total Knee Arthroplasty (TKA) tibial component. To characterize the ability of the steel post/HDPE cylinder interface for transferring mechanical load, pull-out tests were carried out on the model post fixation for different initial interference fits and implantation lengths. Experimental ultimate pull-out force was found to increase in a quasi-linear fashion with increase in the interference fit or the implantation length. On the basis of Coulomb friction at the post/HDPE interface, an analytical model has been developed for the analysis of the experimental ultimate pull-out force results. For the ranges of initial interference fit and implantation length considered, the analytical model was fitted to the experimental results by adjusting the friction coefficient. The mean value of the friction coefficient obtained was 0.043. Finite element simulation was also carried out for the pull-out test using the ABAQUS program. It was found that the finite element simulation describes well the experimental ultimate pull-out forces than the analytical calculation. (C) Koninklijke Brill NV, Leiden, 2009
引用
收藏
页码:259 / 267
页数:9
相关论文
共 15 条
[1]   In vitro assessment of retention of four esthetic dowels to resin core foundation and teeth [J].
Al-harbi, F ;
Nathanson, D .
JOURNAL OF PROSTHETIC DENTISTRY, 2003, 90 (06) :547-555
[2]   Tibial component fixation in total knee arthroplasty - A comparison of pegged and stemmed designs [J].
Bertin, Kim C. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (05) :670-678
[3]   Mechanics of the tapered interference fit in dental implants [J].
Bozkaya, D ;
Müftü, S .
JOURNAL OF BIOMECHANICS, 2003, 36 (11) :1649-1658
[4]   Tibial fixation without screws in cementless total knee arthroplasty [J].
Cooke, C ;
Walter, WK ;
Zicat, B .
JOURNAL OF ARTHROPLASTY, 2006, 21 (02) :237-241
[5]  
Dammak M, 2003, Technol Health Care, V11, P245
[6]   Effect of torque loading on the pull-out response of a steel post inserted in a polyethylene cylinder [J].
Dammak, M. ;
Kharrat, M. ;
Abid, M. ;
Maalej, A. .
JOURNAL OF ADHESION SCIENCE AND TECHNOLOGY, 2006, 20 (11) :1189-1198
[7]   Analysis of cementless implants using interface nonlinear friction - Experimental and finite element studies [J].
Dammak, M ;
ShiraziAdl, A ;
Zukor, DJ .
JOURNAL OF BIOMECHANICS, 1997, 30 (02) :121-129
[8]   Fibrous composite materials in dentistry and orthopaedics: review and applications [J].
Fujihara, K ;
Teo, K ;
Gopal, R ;
Loh, PL ;
Ganesh, VK ;
Ramakrishna, S ;
Foong, KWC ;
Chew, CL .
COMPOSITES SCIENCE AND TECHNOLOGY, 2004, 64 (06) :775-788
[9]   Surface cementation of stemmed tibial components in primary total knee arthroplasty - Minimum 5-year follow-up [J].
Hofmann, AA ;
Goldberg, TD ;
Tanner, AM ;
Cook, TM .
JOURNAL OF ARTHROPLASTY, 2006, 21 (03) :353-357
[10]   The effect of stem modularity and mode of fixation on tibial component stability in revision total knee arthroplasty [J].
Jazrawi, LM ;
Bai, B ;
Kummer, FJ ;
Hiebert, R ;
Stuchin, SA .
JOURNAL OF ARTHROPLASTY, 2001, 16 (06) :759-767