Establishment of reference interval for immature platelet fraction using Sysmex XN9000 hematology analyzer in Anhui province in China

被引:0
|
作者
Yang, Baixia [1 ]
Zhou, Yindi
Ding, Shoufu [1 ]
Zha, Helin [1 ]
Zhang, Fan [1 ]
He, Wentao [1 ]
Xu, Yuanhong [1 ]
机构
[1] Anhui Med Univ, Affiliated Hosp 1, Dept Clin Lab, 218 Jixi Rd, Hefei 230022, Peoples R China
关键词
Reticulated platelet; reference interval; immature platelet fraction; UMBILICAL-CORD BLOOD; PERFORMANCE EVALUATION; TERM;
D O I
暂无
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The aim of this study was to establish reference intervals for immature platelet fraction (IPF) for healthy adults and term newborns in the area of Anhui province. Except outliers, a total of 2179 health adult examination subjects and 153 healthy term newborns were included in this study. The percentage (IPF%) and absolute value (IPF#) of IPF in adult peripheral blood and neonatal umbilical cord blood were measured using the Sysmex XN9000 fully automatic hematology analyzer. The reference intervals were calculated by non-parametric percentile method according to the Clinical and Laboratory Standard Institute guideline C28-A3. Our results showed that, for adults, no significant differences were found between males and females, while significant difference existed between different age groups for IPF% and IPF#. A significant difference also existed between adults and term newborns for all test parameters. The reference intervals of IPF% and IPF# for adults and term newborns were 0.7-8.4% and 2.9-16.9x109/L, 0.9-4.8% and 3.1-13.5x109/L, respectively. The reference intervals determined by the Sysmex XN9000 hematology analyzer for healthy adults and term newborns in the area of Anhui province would be useful for future clinical diagnosis as well as treatment.
引用
收藏
页码:4214 / 4220
页数:7
相关论文
共 29 条
  • [1] Reference interval for immature platelet fraction on Sysmex XN hematology analyzer: a comparison study with Sysmex XE-2100
    Ko, Young Jin
    Hur, Mina
    Kim, Hanah
    Choi, Sang Gyeu
    Moon, Hee-Won
    Yun, Yeo-Min
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (07) : 1091 - 1097
  • [2] Application of the Fluorescence Method on Sysmex XN9000 Hematology Analyzer for Correcting Platelet Count in Individuals with Microcytosis
    Deng, Jiankai
    Xie, Shuhua
    Chen, Yaoming
    Ma, Qinghua
    He, Yuting
    Liu, Min
    Wang, Dong
    Yu, Xuegao
    LABORATORY MEDICINE, 2023, 54 (01) : E10 - E13
  • [3] REFERENCE INTERVAL FOR IMMATURE PLATELET FRACTION IN SYSMEX XN MODULAR SYSTEM
    Kim, Hanah
    Ko, Young Jin
    Hur, Mina
    Choi, Sang Gyeu
    Moon, Hee-Won
    Yun, Yeo-Min
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 84 - 84
  • [4] Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
    Imperiali, Claudia E.
    Arbiol-Roca, Ariadna
    Sanchez-Navarro, Lourdes
    Dastis-Arias, Macarena
    Lopez-Delgado, Juan C.
    Cortes-Bosch, Anna
    Sancho-Cerro, Ana
    Dot-Bach, Dolors
    BIOCHEMIA MEDICA, 2018, 28 (01)
  • [5] Performance of the Sysmex XN-V hematology analyzer in determining the immature platelet fraction in dogs: A preliminary study and reference values
    Jornet-Rius, Oriol
    Mesalles-Naranjo, Montse
    Pastor, Josep
    VETERINARY CLINICAL PATHOLOGY, 2023, 52 (03) : 433 - 442
  • [6] ESTABLISHMENT OF REFERENCE INTERVAL FOR IMMATURE PLATELET FRACTION
    Ko, Young Jin
    Kim, Hanah
    Hur, Mina
    Choi, Sang-Gyeu
    Moon, Hee-Won
    Park, Chul-Min
    Yun, Yeo-Min
    Hong, Sung Noh
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 95 - 95
  • [7] Establishment of reference interval for immature platelet fraction
    Ko, Y. J.
    Kim, H.
    Hur, M.
    Choi, S. G.
    Moon, H. -W.
    Yun, Y. -M.
    Hong, S. N.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2013, 35 (05) : 528 - 533
  • [8] Reference intervals for mean platelet volume and immature platelet fraction determined on a sysmex XE5000 hematology analyzer
    Joergensen, Mikala Klok
    Bathum, Lise
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2016, 76 (02): : 172 - 176
  • [9] Abnormal leukocyte scattergrams and immature platelet fraction on Sysmex XN-9000 analyzer: a new diagnostic tool for altered megakaryopoiesis?
    Buoro, Sabrina
    Manenti, Barbara
    Seghezzi, Michela
    Dominoni, Paola
    Vavassori, Mauro
    Trezzi, Rosangela
    Galli, Monica
    Finazzi, Guido
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2017, 77 (01): : 73 - 75
  • [10] Measurement uncertainty of platelet concentration using the Sysmex XN automated hematology analyzer
    Lim, Yong Kwan
    Kweon, Oh Joo
    Choi, Jee-Hye
    Lee, Woonhyoung
    Park, Ae Ja
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2018, 78 (03): : 224 - 229