PMN-PT Single-Crystal High-Frequency Kerfless Phased Array

被引:38
作者
Chen, Ruimin [1 ]
Cabrera-Munoz, Nestor E. [1 ]
Lam, Kwok Ho [1 ,2 ]
Hsu, Hsiu-sheng [1 ]
Zheng, Fan [1 ]
Zhou, Qifa [1 ]
Shung, K. Kirk [1 ]
机构
[1] Univ So Calif, Dept Biomed Engn, NIH Resource Ctr Med Ultrason Transducer Technol, Los Angeles, CA 90089 USA
[2] Hong Kong Polytech Univ, Dept Elect Engn, Hunghom, Hong Kong, Peoples R China
基金
美国国家卫生研究院;
关键词
INTRAVASCULAR ULTRASOUND CATHETER; FORWARD-LOOKING IVUS; LINEAR-ARRAY; TRANSDUCER; DESIGN;
D O I
10.1109/TUFFC.2014.2999
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper reports the design, fabrication, and characterization of a miniature high-frequency kerfless phased array prepared from a PMN-PT single crystal for forward-looking intravascular or endoscopic imaging applications. After lapping down to around 40 mu m, the PMN-PT material was utilized to fabricate 32-element kerfless phased arrays using micromachining techniques. The aperture size of the active area was only 1.0 x 1.0 mm. The measured results showed that the array had a center frequency of 40 MHz, a bandwidth of 34% at -6 dB with a polymer matching layer, and an insertion loss of 20 dB at the center frequency. Phantom images were acquired and compared with simulated images. The results suggest that the feasibility of developing a phased array mounted at the tip of a forward-looking intravascular catheter or endoscope. The fabricated array exhibits much higher sensitivity than PZT ceramic-based arrays and demonstrates that PMN-PT is well suited for this application.
引用
收藏
页码:1033 / 1041
页数:9
相关论文
共 39 条
[1]  
BEZANSON A, 2012, P IEEE ULTR S, P2114
[2]  
Cannata JM, 2005, ULTRASON, P109
[3]   Development of a 35-MHz piezo-composite ultrasound array for medical imaging [J].
Cannata, JM ;
Williams, JA ;
Zhou, QF ;
Ritter, TA ;
Shung, KK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (01) :224-236
[4]   High frequency PMN-PT single crystal focusing transducer fabricated by a mechanical dimpling technique [J].
Chen, Y. ;
Lam, K. H. ;
Zhou, D. ;
Cheng, W. F. ;
Dai, J. Y. ;
Luo, H. S. ;
Chan, H. L. W. .
ULTRASONICS, 2013, 53 (02) :345-349
[5]   Annular-ring CMUT arrays for forward-looking IVUS: Transducer characterization and imaging [J].
Degertekin, FL ;
Guldiken, RO ;
Karaman, M .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (02) :474-482
[6]  
Dmore C. E. M., 2013, IEEE T ULTRASON FERR, V53, P1046
[7]   PMN-PT single-crystal transducer for non-destructive evaluation [J].
Edwards, G. ;
Chan, H. L. W. ;
Batten, A. ;
Lam, K. H. ;
Luo, H. S. ;
Scott, D. A. .
SENSORS AND ACTUATORS A-PHYSICAL, 2006, 132 (02) :434-440
[8]   ARTERIAL IMAGING WITH A NEW FORWARD-VIEWING INTRAVASCULAR ULTRASOUND CATHETER .1. INITIAL STUDIES [J].
EVANS, JL ;
NG, KH ;
VONESH, MJ ;
KRAMER, BL ;
MEYERS, SN ;
MILLS, TA ;
KANE, BJ ;
ALDRICH, WN ;
JANG, YT ;
YOCK, PG ;
ROLD, MD ;
ROTH, SI ;
MCPHERSON, DD .
CIRCULATION, 1994, 89 (02) :712-717
[9]   A NEW 15-50 MHz ARRAY-BASED MICRO-ULTRASOUND SCANNER FOR PRECLINICAL IMAGING [J].
Foster, F. Stuart ;
Mehi, James ;
Lukacs, Marc ;
Hirson, Desmond ;
White, Chris ;
Chaggares, Chris ;
Needles, Andrew .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (10) :1700-1708
[10]   Advances in ultrasound biomicroscopy [J].
Foster, FS ;
Pavlin, CJ ;
Harasiewicz, KA ;
Christopher, DA ;
Turnbull, DH .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (01) :1-27