Hardware-sensitive optimization for intensity modulated radiotherapy

被引:38
作者
Cho, PS
Marks, RJ
机构
[1] Univ Washington, Dept Radiat Oncol, Seattle, WA 98195 USA
[2] Univ Washington, Dept Elect Engn, Seattle, WA 98195 USA
关键词
D O I
10.1088/0031-9155/45/2/312
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The multileaf collimator (MLC) hardware constraints are usually neglected in the process of intensity-modulated beam optimization. Consequently, it is not always possible to deliver planned beam modulation using dynamic MLC. Beam optimization is significantly diminished if the results must be approximated due to limitations imposed by the delivery device. To overcome this problem, an inverse beam optimization method which incorporates the hardware constraints has been developed. The hardware constraints, including the leaf velocity, the dose rate and the minimum required gap between opposing and adjacent leaves, were considered. An iterative search for feasible modulation was conducted alternately in the dosimetric space and the MLC position-time space. The optimization algorithm was designed for a unidirectional leaf trajectory and a constant dose rate. A scheme to reduce tongue-and-groove underdosage during optimization was also implemented. Comparisons were made between the solutions produced by this method and conventional optimization disregarding the hardware restrictions. The beam profiles generated by the conventional method were modified to satisfy the hardware specifications. The results indicate that inclusion of MLC constraints during optimization can improve the degree of conformity that is deliverable.
引用
收藏
页码:429 / 440
页数:12
相关论文
共 29 条
[1]   METHOD OF LINEAR PROGRAMMING APPLIED TO RADIATION TREATMENT PLANNING [J].
BAHR, GK ;
KEREIAKES, JG ;
HORWITZ, H ;
FINNEY, R ;
GALVIN, J ;
GOODE, K .
RADIOLOGY, 1968, 91 (04) :686-+
[2]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[3]  
Boyer AL, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P13
[4]   ON THE USE OF CIMMINO SIMULTANEOUS PROJECTIONS METHOD FOR COMPUTING A SOLUTION OF THE INVERSE PROBLEM IN RADIATION-THERAPY TREATMENT PLANNING [J].
CENSOR, Y ;
ALTSCHULER, MD ;
POWLIS, WD .
INVERSE PROBLEMS, 1988, 4 (03) :607-623
[5]  
Cho PS, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P310
[6]   Optimization of intensity modulated beams with volume constraints using two methods: Cost function minimization and projections onto convex sets [J].
Cho, PS ;
Lee, S ;
Marks, RJ ;
Oh, SH ;
Sutlief, SG ;
Phillips, MH .
MEDICAL PHYSICS, 1998, 25 (04) :435-443
[7]  
Convery D, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P350
[8]   Generation of discrete beam-intensity modulation by dynamic multileaf collimation under minimum leaf separation constraints [J].
Convery, DJ ;
Webb, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09) :2521-2538
[9]  
Dirkx MLP, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P357
[10]   Leaf trajectory calculation for dynamic multileaf collimation to realize optimized fluence profiles [J].
Dirkx, MLP ;
Heijmen, BJM ;
van Santvoort, JPC .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (05) :1171-1184