A twin ionization chamber setup as detector for light charged particles with energies around 1 MeV applied to the 10B(n, α)7Li reaction

被引:21
作者
Göpfert, A [1 ]
Hambsch, FJ [1 ]
Bax, H [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
关键词
alpha-particle spectroscopy; ionization chamber; standard; electronic filters; background suppression;
D O I
10.1016/S0168-9002(99)00952-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The setup and operating conditions of a gridded twin ionization chamber with sample change facility to study light charged particle properties in the 1 MeV region is described. Detailed studies of different grid geometries in connection with the choice of an eligible counting gas mixture and the applied high voltage have been performed. Due to the high overall amplification of the small electrical chamber signals obtained from such low-energy particles, special filters have been developed in order to increase the signal-to-noise ratio. Timing properties of the chamber signals are discussed in detail. Information available from chamber signals and encoding methods are elucidated by spectra of alpha particles created by U-234,U-235 spontaneous alpha decay. The detector permits the independent and simultaneous measurement of energy and angular distribution of particles in both sides of the chamber. Finally, preliminary results and related analysis methods will be presented for the investigation of the B-10(n, alpha(0))/B-10(n, alpha(1)y) branching ratios. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:438 / 451
页数:14
相关论文
empty
未找到相关数据