Small volume flow probe for automated direct-injection NMR analysis: Design and performance

被引:25
作者
Haner, RL
Llanos, W
Mueller, L
机构
[1] Varian NMR Syst, Palo Alto, CA 94304 USA
[2] Bristol Myers Squibb, Princeton, NJ 08543 USA
关键词
direct-injection NMR; SAR by NMR; high-throughput screening; LC-NMR; NMR probe;
D O I
10.1006/jmre.1999.1983
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A detailed characterization of an NMR how probe for use in direct-injection sample analysis is presented. A 600-MHz, indirect detection NMR flow probe with a 120-mu l active volume is evaluated in two configurations: first as a stand-alone small volume probe for the analysis of static, nonflowing solutions, and second as a component in an integrated liquids-handling system used for high-throughput NMR analysis. In the stand-alone mode,H-1 lineshape, sensitivity, radiofrequency (RF) homogeneity, and heat transfer characteristics are measured and compared to conventional-format NMR probes of related design. Commonly used descriptive terminology for the hardware, sample regions, and RF coils are reviewed or defined, and test procedures developed for how probes are described. The flow probe displayed general performance that is competitive with standard probes. Key advantages of the flow probe include high molar sensitivity, ease of use in an automation setup, and superior reproducibility of magnetic field homogeneity which enables the practical implementation of 1D T2-edited analysis of protein-ligand interactions. (C) 2000 Academic Press.
引用
收藏
页码:69 / 78
页数:10
相关论文
共 14 条
  • [1] REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY NUCLEAR MAGNETIC-RESONANCE ONLINE COUPLING WITH SOLVENT NON-EXCITATION
    ALBERT, K
    KUNST, M
    BAYER, E
    SPRAUL, M
    BERMEL, W
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 463 (02): : 355 - 363
  • [2] Adaptation of commercial 500 mHz probes for LCNMR
    Barjat, H
    Morris, GA
    Newman, MJ
    Swanson, AG
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES A, 1996, 119 (01) : 115 - 119
  • [3] ONLINE COUPLING OF LIQUID-CHROMATOGRAPHY AND HIGH-FIELD NUCLEAR MAGNETIC-RESONANCE SPECTROMETRY
    BAYER, E
    ALBERT, K
    NIEDER, M
    GROM, E
    WOLFF, G
    RINDLISBACHER, M
    [J]. ANALYTICAL CHEMISTRY, 1982, 54 (11) : 1747 - 1750
  • [4] One-dimensional relaxation- and diffusion-edited NMR methods for screening compounds that bind to macromolecules
    Hajduk, PJ
    Olejniczak, ET
    Fesik, SW
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1997, 119 (50) : 12257 - 12261
  • [5] Haner R. L., 1999, U.S. patent, Patent No. [U.S. 5867026 A, 5867026]
  • [6] CONTINUOUS-FLOW HIGH-FIELD NUCLEAR MAGNETIC-RESONANCE DETECTOR FOR LIQUID-CHROMATOGRAPHIC ANALYSIS OF FUEL SAMPLES
    HAW, JF
    GLASS, TE
    DORN, HC
    [J]. ANALYTICAL CHEMISTRY, 1981, 53 (14) : 2327 - 2332
  • [7] Hill H.D.W., 1993, US 5192,911, Patent No. 5192911
  • [8] LIMITS OF MEASUREMENT IN MAGNETIC RESONANCE
    HILL, HDW
    RICHARDS, RE
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1968, 1 (10): : 977 - &
  • [9] High-resolution NMR techniques for solid-phase synthesis and combinatorial chemistry
    Keifer, PA
    [J]. DRUG DISCOVERY TODAY, 1997, 2 (11) : 468 - 478
  • [10] Keifer Paul A., 1998, Drugs of the Future, V23, P301