A system to use electromagnetic tracking for the quality assurance of brachytherapy catheter digitization

被引:52
|
作者
Damato, Antonio L. [1 ]
Viswanathan, Akila N. [1 ]
Don, Sarah M. [1 ]
Hansen, Jorgen L. [1 ]
Cormack, Robert A. [1 ]
机构
[1] Brigham & Womens Hosp, Dana Farber Canc Inst, Dept Radiat Oncol, Boston, MA 02115 USA
关键词
brachytherapy; quality assurance; electromagnetic tracking; UNCERTAINTIES; ACCURACY;
D O I
10.1118/1.4894710
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the use of a system using electromagnetic tracking (EMT), post-processing and an error-detection algorithm for detecting errors and resolving uncertainties in high-dose-rate brachytherapy catheter digitization for treatment planning. Methods: EMT was used to localize 15 catheters inserted into a phantom using a stepwise acquisition technique. Five distinct acquisition experiments were performed. Noise associated with the acquisition was calculated. The dwell location configuration was extracted from the EMT data. A CT scan of the phantom was performed, and five distinct catheter digitization sessions were performed. No a priori registration of the CT scan coordinate system with the EMT coordinate system was performed. CT-based digitization was automatically extracted from the brachytherapy plan DICOM files (CT), and rigid registration was performed between EMT and CT dwell positions. EMT registration error was characterized in terms of the mean and maximum distance between corresponding EMT and CT dwell positions per catheter. An algorithm for error detection and identification was presented. Three types of errors were systematically simulated: swap of two catheter numbers, partial swap of catheter number identification for parts of the catheters (mix), and catheter-tip shift. Error-detection sensitivity (number of simulated scenarios correctly identified as containing an error/number of simulated scenarios containing an error) and specificity (number of scenarios correctly identified as not containing errors/number of correct scenarios) were calculated. Catheter identification sensitivity (number of catheters correctly identified as erroneous across all scenarios/number of erroneous catheters across all scenarios) and specificity (number of catheters correctly identified as correct across all scenarios/number of correct catheters across all scenarios) were calculated. The mean detected and identified shift was calculated. Results: The maximum noise +/- 1 standard deviation associated with the EMT acquisitions was 1.0 +/- 0.1 mm, and the mean noise was 0.6 +/- 0.1 mm. Registration of all the EMT and CT dwell positions was associated with a mean catheter error of 0.6 +/- 0.2 mm, a maximum catheter error of 0.9 +/- 0.4 mm, a mean dwell error of 1.0 +/- 0.3 mm, and a maximum dwell error of 1.3 +/- 0.7 mm. Error detection and catheter identification sensitivity and specificity of 100% were observed for swap, mix and shift (>= 2.6 mm for error detection; >= 2.7 mm for catheter identification) errors. A mean detected shift of 1.8 +/- 0.4 mm and a mean identified shift of 1.9 +/- 0.4 mm were observed. Conclusions: Registration of the EMT dwell positions to the CT dwell positions was possible with a residual mean error per catheter of 0.6 +/- 0.2 mm and a maximum error for any dwell of 1.3 +/- 0.7 mm. These low residual registration errors show that quality assurance of the general characteristics of the catheters and of possible errors affecting one specific dwell position is possible. The sensitivity and specificity of the catheter digitization verification algorithm was 100% for swap and mix errors and for shifts >= 2.6 mm. On average, shifts >= 1.8 mm were detected, and shifts >= 1.9 mm were detected and identified. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Localization of reference points in electromagnetic tracking data and their application for treatment error detection in interstitial breast brachytherapy
    Duerrbeck, Christopher
    Schuster, Sabrina
    Sauer, Birte Christina
    Abu-Hossin, Nadin
    Strnad, Vratislav
    Fietkau, Rainer
    Bert, Christoph
    MEDICAL PHYSICS, 2023, 50 (09) : 5772 - 5783
  • [22] Assessment of integrated electromagnetic tracking for dwell position monitoring in a clinical HDR brachytherapy setting for prostate cancer
    Androulakis, Ioannis
    Schiphof-Godart, Jeremy
    van Heerden, Laura E.
    Luthart, Lorne
    Rijnsdorp, Rene
    Hoogeman, Mischa S.
    Westerveld, Henrike
    Christianen, Miranda E. M. C.
    Mens, Jan Willem M.
    van Paassen, Rosemarijn
    Negenman, Eva M.
    Nout, Remi A.
    Kolkman-Deurloo, Inger Karine K.
    RADIOTHERAPY AND ONCOLOGY, 2024, 200
  • [23] Feasibility of a quality assurance system for electromagnetic field therapy
    Heehun Sung
    Jong Hyun Kim
    Jaehyeon Seo
    Geon Oh
    Yunhui Jo
    Yongha Gi
    Hyunwoo Kim
    Myonggeun Yoon
    Journal of the Korean Physical Society, 2022, 81 : 1029 - 1038
  • [24] Feasibility of a quality assurance system for electromagnetic field therapy
    Sung, Heehun
    Kim, Jong Hyun
    Seo, Jaehyeon
    Oh, Geon
    Jo, Yunhui
    Gi, Yongha
    Kim, Hyunwoo
    Yoon, Myonggeun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 81 (11) : 1029 - 1038
  • [25] An anthropomorphic phantom for quality assurance and training in gynaecological brachytherapy
    de Almeida, CE
    Rodriguez, M
    Vianello, E
    Ferreira, IH
    Sibata, C
    RADIOTHERAPY AND ONCOLOGY, 2002, 63 (01) : 75 - 81
  • [26] Improved electromagnetic tracking for catheter path reconstruction with application in high-dose-rate brachytherapy
    Elodie Lugez
    Hossein Sadjadi
    Chandra P. Joshi
    Selim G. Akl
    Gabor Fichtinger
    International Journal of Computer Assisted Radiology and Surgery, 2017, 12 : 681 - 689
  • [27] Electromagnetic tracking for catheter localization
    Schneider, MR
    BIOMEDICAL DIAGNOSTIC, GUIDANCE, AND SURGICAL-ASSIST SYSTEMS, 1999, 3595 : 61 - 68
  • [28] Medical aspects of quality assurance in brachytherapy
    Kovacs, G
    STRAHLENTHERAPIE UND ONKOLOGIE, 1998, 174 : 47 - 49
  • [29] Positional and angular tracking of HDR 192Ir source for brachytherapy quality assurance using radiochromic film dosimetry
    Aldelaijan, Saad
    Devic, Slobodan
    Bekerat, Hamed
    Papaconstadopoulos, Pavlos
    Schneider, James
    Seuntjens, Jan
    Cormack, Robert A.
    Buzurovic, Ivan M.
    MEDICAL PHYSICS, 2020, 47 (12) : 6122 - 6139
  • [30] Electromagnetic tracking for catheter reconstruction in ultrasound-guided high-dose-rate brachytherapy of the prostate
    Bharat, Shyam
    Kung, Cynthia
    Dehghan, Ehsan
    Ravi, Ananth
    Venugopal, Niranjan
    Bonillas, Antonio
    Stanton, Doug
    Kruecker, Jochen
    BRACHYTHERAPY, 2014, 13 (06) : 640 - 650