Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses

被引:75
作者
Galvin, Alison L. [1 ]
Kang, Lu [1 ]
Udofia, Itoro [1 ]
Jennings, Louise M. [1 ]
McEwen, Hannah M. J. [2 ]
Jin, Zhongmin [1 ]
Fisher, John [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] DePuy Int Ltd, Leeds, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
Total knee replacement; Polyethylene wear; MOLECULAR-WEIGHT POLYETHYLENE; TOTAL JOINT REPLACEMENTS; TIBIAL COMPONENTS; PRESSURE PREDICTION; DELAMINATION; KINEMATICS; UHMWPE; DESIGN;
D O I
10.1016/j.jbiomech.2009.05.010
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Ultra high molecular weight polyethylene (PE) remains the primary bearing surface of choice in total knee replacements (TKR). Wear is controlled by levels of cross-shear motion and contact stress. The aim of this study was to compare the wear of fixed-bearing total knee replacements with curved and flat inserts and to test the hypothesis that the flat inserts which give higher contact stresses and smaller contact areas would lead to lower levels of surface wear. A low-conforming, high contact stress knee with a low-medium level of cross shear resulted in significantly lower wear rates in comparison to a standard cruciate sacrificing fixed-bearing knee. The low wear solution found in the knee simulator was supported by fundamental studies of wear as a function of pressure and cross shear in the pin on plate system. Current designs of fixed-bearing knees do not offer this low wear solution due to their medium cross shear, moderate conformity and medium contact stress. (C) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1898 / 1902
页数:5
相关论文
共 26 条
  • [1] CONTACT AND RUBBING OF FLAT SURFACES
    ARCHARD, JF
    [J]. JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) : 981 - 988
  • [2] THE INFLUENCE OF CONTACT STRESS ON THE WEAR OF UHMWPE FOR TOTAL REPLACEMENT HIP PROSTHESES
    BARBOUR, PSM
    BARTON, DC
    FISHER, J
    [J]. WEAR, 1995, 181 : 250 - 257
  • [3] Investigation of wear of knee prostheses in a new displacement/force-controlled simulator
    Barnett, PI
    McEwen, HMJ
    Auger, DD
    Stone, MH
    Ingham, E
    Fisher, J
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H1) : 51 - 61
  • [4] Comparison of wear in a total knee replacement under different kinematic conditions
    Barnett, PI
    Fisher, J
    Auger, DD
    Stone, MH
    Ingham, E
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2001, 12 (10-12) : 1039 - 1042
  • [5] THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT
    BARTEL, DL
    BICKNELL, VL
    WRIGHT, TM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) : 1041 - 1051
  • [6] BARTEL DL, 1995, CLIN ORTHOP RELAT R, P76
  • [7] Effect of oxidation on delamination of ultrahigh-molecular-weight polyethylene tibial components
    Bell, CJ
    Walker, PS
    Abeysundera, MR
    Simmons, JMH
    King, PM
    Blunn, GW
    [J]. JOURNAL OF ARTHROPLASTY, 1998, 13 (03) : 280 - 290
  • [8] POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES
    ENGH, GA
    DWYER, KA
    HANES, CK
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 9 - 17
  • [9] Posterior cruciate-retaining modular total knee arthroplasty - A 9- to 12-year follow-up investigation
    Fetzer, GB
    Callaghan, JJ
    Templeton, JE
    Goetz, DD
    Sullivan, PM
    Kelley, SS
    [J]. JOURNAL OF ARTHROPLASTY, 2002, 17 (08) : 961 - 966
  • [10] Howling GI, 2001, J BIOMED MATER RES, V58, P415, DOI 10.1002/jbm.1036