Simulated Radiation Dose Reduction in Whole-Body CT on a 3rd Generation Dual-Source Scanner: An Intraindividual Comparison

被引:6
|
作者
Brendlin, Andreas S. [1 ]
Winkelmann, Moritz T. [1 ]
Do, Phuong Linh [2 ]
Schwarze, Vincent [3 ]
Peisen, Felix [1 ]
Almansour, Haidara [1 ]
Bongers, Malte N. [1 ]
Artzner, Christoph P. [1 ]
Weiss, Jakob [4 ]
Kim, Jong Hyo [5 ,6 ]
Othman, Ahmed E. [1 ]
Afat, Saif [1 ]
机构
[1] Univ Hosp Tuebingen, Dept Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
[2] St Gertrauden Hosp Berlin, Dept Cardiol, D-10713 Berlin, Germany
[3] Univ Hosp LMU, Dept Radiol, D-80336 Munich, Germany
[4] Univ Hosp Freiburg, Dept Diagnost & Intervent Radiol, D-79106 Freiburg, Germany
[5] Seoul Natl Univ, Dept Radiol, Coll Med, Seoul 03080, South Korea
[6] Adv Inst Convergence Technol, Suwon 16229, South Korea
关键词
CT; whole-body staging; radiation dose; 3rd generation dual-source scanner; iterative reconstruction (IR); Filtered Back Projection (FBP); Advanced Modeled Iterative Reconstruction (ADMIRE); image reconstruction; MODELED ITERATIVE RECONSTRUCTION; TUBE VOLTAGE SELECTION; IMAGE QUALITY; COMPUTED-TOMOGRAPHY; DIAGNOSTIC-ACCURACY; ANGIOGRAPHY; ALGORITHM; RISK;
D O I
10.3390/diagnostics11010118
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To evaluate the effect of radiation dose reduction on image quality and diagnostic confidence in contrast-enhanced whole-body computed tomography (WBCT) staging. We randomly selected March 2016 for retrospective inclusion of 18 consecutive patients (14 female, 60 +/- 15 years) with clinically indicated WBCT staging on the same 3rd generation dual-source CT. Using low-dose simulations, we created data sets with 100, 80, 60, 40, and 20% of the original radiation dose. Each set was reconstructed using filtered back projection (FBP) and Advanced Modeled Iterative Reconstruction (ADMIRE(R), Siemens Healthineers, Forchheim, Germany) strength 1-5, resulting in 540 datasets total. ADMIRE 2 was the reference standard for intraindividual comparison. The effective radiation dose was calculated using commercially available software. For comparison of objective image quality, noise assessments of subcutaneous adipose tissue regions were performed automatically using the software. Three radiologists blinded to the study evaluated image quality and diagnostic confidence independently on an equidistant 5-point Likert scale (1 = poor to 5 = excellent). At 100%, the effective radiation dose in our population was 13.3 +/- 9.1 mSv. At 20% radiation dose, it was possible to obtain comparably low noise levels when using ADMIRE 5 (p = 1.000, r = 0.29). We identified ADMIRE 3 at 40% radiation dose (5.3 +/- 3.6 mSv) as the lowest achievable radiation dose with image quality and diagnostic confidence equal to our reference standard (p = 1.000, r > 0.4). The inter-rater agreement for this result was almost perfect (ICC >= 0.958, 95% CI 0.909-0.983). On a 3rd generation scanner, it is feasible to maintain good subjective image quality, diagnostic confidence, and image noise in single-energy WBCT staging at dose levels as low as 40% of the original dose (5.3 +/- 3.6 mSv), when using ADMIRE 3.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Effects of radiation dose reduction on diagnostic performance of 3rd generation Dual Source CT pulmonary angiography
    Winkelmann, Moritz T.
    Walter, Sven S.
    Stock, Eva
    Brendlin, Andreas
    Kolb, Manuel
    Othman, Ahmed E.
    Afat, Saif
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 134
  • [2] Effects of radiation dose reduction on diagnostic performance of 3rd generation Dual Source CT pulmonary angiography
    Winkelmann, Moritz T.
    Walter, Sven S.
    Stock, Eva
    Brendlin, Andreas
    Kolb, Manuel
    Othman, Ahmed E.
    Afat, Saif
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 134
  • [3] Approaches to ultra-low radiation dose coronary artery calcium scoring based on 3rd generation dual-source CT: A phantom study
    McQuiston, Andrew D.
    Muscogiuri, Giuseppe
    Schoepf, U. Joseph
    Meinel, Felix G.
    Canstein, Christian
    Varga-Szemes, Akos
    Cannao, Paola M.
    Wichmann, Julian L.
    Allmendinger, Thomas
    Vliegenthart, Rozemarijn
    De Cecco, Carlo N.
    EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (01) : 39 - 47
  • [4] Diagnosis of Pulmonary Artery Embolism: Comparison of Single-Source CT and 3rd Generation Dual-Source CT using a Dual-Energy Protocol Regarding Image Quality and Radiation Dose
    Petritsch, Bernhard
    Kosmala, Aleksander
    Gassenmaier, Tobias
    Weng, Andreas Max
    Veldhoen, Simon
    Kunz, Andreas Steven
    Bley, Thorsten Alexander
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2017, 189 (06): : 527 - 536
  • [5] Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3rd generation dual-source CT
    Mangold, Stefanie
    Wichmann, Julian L.
    Schoepf, U. Joseph
    Poole, Zachary B.
    Canstein, Christian
    Varga-Szemes, Akos
    Caruso, Damiano
    Bamberg, Fabian
    Nikolaou, Konstantin
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2016, 26 (10) : 3608 - 3616
  • [6] Automated tube voltage selection for radiation dose and contrast medium reduction at coronary CT angiography using 3rd generation dual-source CT
    Stefanie Mangold
    Julian L. Wichmann
    U. Joseph Schoepf
    Zachary B. Poole
    Christian Canstein
    Akos Varga-Szemes
    Damiano Caruso
    Fabian Bamberg
    Konstantin Nikolaou
    Carlo N. De Cecco
    European Radiology, 2016, 26 : 3608 - 3616
  • [7] Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Wichmann, Julian L.
    Hardie, Andrew D.
    Schoepf, U. Joseph
    Felmly, Lloyd M.
    Perry, Jonathan D.
    Varga-Szemes, Akos
    Mangold, Stefanie
    Caruso, Damiano
    Canstein, Christian
    Vogl, Thomas J.
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2017, 27 (02) : 642 - 650
  • [8] Diagnostic Performance of Different Simulated Low-Dose Levels in Patients with Suspected Cervical Abscess Using a Third-Generation Dual-Source CT Scanner
    Winkelmann, Moritz T.
    Afat, Saif
    Walter, Sven S.
    Stock, Eva
    Schwarze, Vincent
    Brendlin, Andreas
    Kolb, Manuel
    Artzner, Christoph P.
    Othman, Ahmed E.
    DIAGNOSTICS, 2020, 10 (12)
  • [9] Head and neck single- and dual-energy CT: differences in radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Lenga, Lukas
    Lange, Marvin
    Martin, Simon S.
    Albrecht, Moritz H.
    Booz, Christian
    Yel, Ibrahim
    Arendt, Christophe T.
    Vogl, Thomas J.
    Leithner, Doris
    BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1122)
  • [10] Radiation exposure and establishment of diagnostic reference levels of whole-body low-dose CT for the assessment of multiple myeloma with second- and third-generation dual-source CT
    Zensen, Sebastian
    Bos, Denise
    Opitz, Marcel
    Haubold, Johannes
    Forsting, Michael
    Guberina, Nika
    Wetter, Axel
    ACTA RADIOLOGICA, 2022, 63 (04) : 527 - 535