Sinogram-based motion correction of PET images using optical motion tracking system and list-mode data acquisition

被引:29
作者
Woo, SK [1 ]
Watabe, H
Choi, Y
Kim, KM
Park, CC
Bloomfield, PM
Iida, H
机构
[1] Univ Konkuk, Comp & Informat Commun Engn Dept, Seoul 135710, South Korea
[2] Natl Cardiovasc Ctr, Inst Res, Dept Investigat Radiol, Osaka 5658565, Japan
[3] Sungkyunkwan Univ, Sch Med, Dept Nucl Med, Samsung Med Ctr, Seoul 135710, South Korea
[4] Konkuk Univ, Comp & Informat Commun Engn Dept, Seoul 143701, South Korea
[5] Ctr Addict & Mental Hlth, Vivian M Rakoff PET Ctr, Toronto, ON M5T 1R8, Canada
关键词
motion correction; optical tracking system; PET; sinogram;
D O I
10.1109/TNS.2004.829786
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Motion of the head during brain positron emission tomography (PET) acquisitions has been identified as a source of artifact in the reconstructed image. A number of techniques have been proposed to correct for this motion artifact, but they are unable to correct for a motion during an acquisition. The aim of this study was to develop a sinogram-based motion correction (SBMC) technique to correct directly the head motion during a PET scan using a motion tracking system and list-mode data acquisition. This method uses a rebinning procedure whereby the lines of response (LOR) are geometrically transformed according to the current values of six-dimensional motion data. A -Michelogram was recomposed using the rebinned LOR, and the motion-corrected sinogram was generated. In the motion corrected image, the blurring artifact due to the motion was reduced by the SBMC technique. This technique was applied to actual PET data acquired in the list-mode, and demonstrated the potential for real-time motion correction of head movements during a PET acquisition.
引用
收藏
页码:782 / 788
页数:7
相关论文
共 15 条
[1]  
Bendriem B., 1998, THEORY PRACTICE 3D P, DOI [10.1007/978-94-017-3475-2, DOI 10.1007/978-94-017-3475-2]
[2]   The design and implementation of a motion correction scheme for neurological PET [J].
Bloomfield, PM ;
Spinks, TJ ;
Reed, J ;
Schnorr, L ;
Westrip, AM ;
Livieratos, L ;
Fulton, R ;
Jones, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (08) :959-978
[3]  
Casey M. E., 1995, Proceedings of the 1995 International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, P67
[4]   Exact and approximate rebinning algorithms for 3-D PET data [J].
Defrise, M ;
Kinahan, PE ;
Townsend, DW ;
Michel, C ;
Sibomana, M ;
Newport, DF .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :145-158
[5]   Spatial registration and normalization of images [J].
Friston, KJ ;
Ashburner, J ;
Frith, CD ;
Poline, JB ;
Heather, JD ;
Frackowiak, RSJ .
HUMAN BRAIN MAPPING, 1995, 3 (03) :165-189
[6]   Correction for head movements in positron emission tomography using an optical motion-tracking system [J].
Fulton, RR ;
Meikle, SR ;
Eberl, S ;
Pfeiffer, J ;
Constable, CJ ;
Fulham, MJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) :116-123
[7]   A head motion measurement system suitable for emission computed tomography [J].
Goldstein, SR ;
DaubeWitherspoon, ME ;
Green, MV ;
Eidsath, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (01) :17-27
[8]  
GREEN MV, 1994, J NUCL MED, V35, P1538
[9]   3-D PHANTOM TO SIMULATE CEREBRAL BLOOD-FLOW AND METABOLIC IMAGES FOR PET [J].
HOFFMAN, EJ ;
CUTLER, PD ;
DIGBY, WM ;
MAZZIOTTA, JC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :616-620
[10]   Implementation and performance of an optical motion tracking system for high resolution brain PET imaging [J].
Lopresti, BJ ;
Russo, A ;
Jones, WF ;
Fisher, T ;
Crouch, DG ;
Altenburger, DE ;
Townsend, DW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (06) :2059-2067