Optical design and evaluation of a 4 mm cost-effective ultra-high-definition arthroscope

被引:16
作者
Cheng, Dewen [1 ]
Wang, Yongtian [1 ]
Yu, Lu [1 ]
Liu, Xiaohua [1 ]
机构
[1] Beijing Inst Technol, Sch Optoelect, Minist Educ China, Key Lab Photoelect Imaging Technol & Syst, Beijing 100081, Peoples R China
来源
BIOMEDICAL OPTICS EXPRESS | 2014年 / 5卷 / 08期
基金
美国国家科学基金会;
关键词
CONFOCAL REFLECTANCE MICROSCOPE; LAPAROSCOPIC SURGERY; RELAY SYSTEMS; WHITE-LIGHT; COLONOSCOPY; ENDOSCOPY; BIOPSY;
D O I
10.1364/BOE.5.002697
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
High definition and magnification rigid endoscope plays an important role in modern minimally invasive medical surgery and diagnosis. In this paper, we present the design and evaluation methods of a high definition rigid endoscope, specifically an arthroscope, with a large depth of field (DOF). The incident heights and exit angles of the sampled rays on the relay lens are controlled during the optimization process to ensure an effective field view (70 degrees) and a normal ray path within the limited lens diameter of 2.7 mm. The lens is set up as a multi-configuration system with two extreme and one middle object distances to cover a large DOF. As a result, an entrance pupil of 0.3 mm is achieved for the first time, to bring the theoretical resolution to 23.1 lps/mm in the object space at a working distance of 20 mm, with the wavelength of 0.532 um. The modulation transfer function (MTF) curves approach diffraction limit, and the values are all higher than 0.3 at 160 line pairs/mm (lps/mm) in the image space. Meanwhile, stray light caused by total internal reflection on the inner wall of the rod lenses and the objective lens is eliminated. The measured resolution in the object space at a 20 mm working distance is 22.3 lps/mm, and test results show that other performance characteristics also fulfill design requirements. The relay lenses are designed with only one type of the spacer and two types of lenses to greatly reduce the fabrication and assembly cost. The design method has important research and application values for lens systems used in modern minimally invasive medical surgery and industrial non-destructive testing area. (C) 2014 Optical Society of America
引用
收藏
页码:2697 / 2714
页数:18
相关论文
共 31 条
[1]  
[Anonymous], 2007, Infectious Disease Epidemiology: Theory and Practice
[2]  
[Anonymous], 2010, Using the glass expert macro, Patent No. 6038079
[3]  
[Anonymous], 2008, 00681 YY
[4]  
[Anonymous], 2014, CODE 5 REF MAN VERS
[5]  
[Anonymous], 2013, LIGHTTOOLS REF MAN V
[6]  
[Anonymous], 2008, Chinese National Standard, YY 0068.1, Patent No. 7724430
[7]  
Bozzini P., 1806, J PRAK HEILK, V24, P107
[8]  
Breault R.P., 1995, Handbook of optics
[9]   Design, assembly, and optical bench testing of a high-numerical-aperture miniature injection-molded objective for fiber-optic confocal reflectance microscopy [J].
Chidley, MD ;
Carlson, KD ;
Richards-Kortum, RR ;
Descour, MR .
APPLIED OPTICS, 2006, 45 (11) :2545-2554
[10]   THE PRIMARY ABERRATION CHARACTERISTICS OF THIN-LENS MODELS OF COMMON RELAY SYSTEMS [J].
DOBSON, SJ ;
RIBEIRO, JACL .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (01) :32-36